Seesaw Balancing Radiation Dose and IV Contrast Dose: Evaluation of a New Abdominal CT Protocol for Reducing Age-Specific Risk

被引:6
作者
Falt, Tobias [1 ]
Soderberg, Marcus [2 ]
Horberg, Lisa [1 ]
Carlgren, Ingela [1 ]
Leander, Peter [1 ]
机构
[1] Lund Univ, Skane Univ Hosp, Dept Radiol, SE-20502 Malmo, Sweden
[2] Lund Univ, Skane Univ Hosp, Dept Clin Sci, SE-20502 Malmo, Sweden
关键词
contrast medium; CT; image quality; radiation dose; INDUCED NEPHROTOXICITY; IMAGE QUALITY; ENHANCEMENT; REDUCTION;
D O I
10.2214/AJR.12.8534
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to evaluate an abdominal CT protocol in which radiation dose was reduced and IV contrast dose increased for young patients and radiation dose was increased and IV medium dose decreased for elderly patients. The hypothesis was that these adjustments would result in constant image quality and a reduction in age-specific risk. MATERIALS AND METHODS. Patients were divided into four age groups of 25 patients each: group 1, 16-25 years; group 2, 26-50 years; group 3, 51-75 years; and group 4, older than 75 years. The quality reference tube load ranged from 100 to 300 mAs, and the IV contrast dose ranged from 600 to 350 mg I/kg. Group 3 was the reference group. Signal-to-noise and contrast-to-noise ratios for a hypothetical hypovascular liver metastatic lesion were calculated. Subjective image quality was evaluated by visual grading characteristic analysis in which four readers assessed the reproduction of seven image-quality criteria. RESULTS. Radiation dose was reduced 57% in the youngest group, and the IV contrast dose was reduced 18% in elderly patients. There were no statistically significant differences between the groups with respect to signal-to-noise and contrast-to-noise ratios. Subjective image quality was graded significantly lower for four criteria in group 1 compared with group 3. No significant difference was found in comparisons of groups 2 (except for one criterion) and 4 with group 3. CONCLUSION. It is possible to balance radiation dose and contrast dose against each other and maintain signal-to-noise and contrast-to-noise ratios. Subjective image quality was affected by increased noise level on the images but was judged acceptable in all groups except the one with the lowest radiation dose.
引用
收藏
页码:383 / 388
页数:6
相关论文
共 18 条
[1]  
[Anonymous], 2009, National Council on Radiation Protection and Measurement report No. 160: Ionizing radiation exposure of the population of the United States
[2]   Visual grading characteristics (VGC) analysis:: a non-parametric rank-invariant statistical method for image quality evaluation [J].
Bath, M. ;
Mansson, L. G. .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (951) :169-176
[3]  
BONGARTZ G, 1999, 16262 EUR EUR COMM
[4]   A software tool for increased efficiency in observer performance studies in radiology [J].
Börjesson, S ;
Håkansson, M ;
Båth, M ;
Kheddache, S ;
Svensson, S ;
Tingberg, A ;
Grahn, A ;
Ruschin, M ;
Hemdal, B ;
Mattsson, S ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :45-52
[5]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[6]   Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know [J].
Brenner, DJ ;
Doll, R ;
Goodhead, DT ;
Hall, EJ ;
Land, CE ;
Little, JB ;
Lubin, JH ;
Preston, DL ;
Preston, RJ ;
Puskin, JS ;
Ron, E ;
Sachs, RK ;
Samet, JM ;
Setlow, RB ;
Zaider, M .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (24) :13761-13766
[7]   Background Fluctuation of Kidney Function Versus Contrast-Induced Nephrotoxicity [J].
Bruce, Richard J. ;
Djamali, Aji ;
Shinki, Kazuhiko ;
Michel, Steven J. ;
Fine, Jason P. ;
Pozniak, Myron A. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (03) :711-718
[8]  
Heiken JP, 2006, COMPUTED BODY TOMOGR, V1, P829
[9]   Converting dose-length product to effective dose at CT [J].
Huda, Walter ;
Ogden, Kent M. ;
Khorasani, Mohammad R. .
RADIOLOGY, 2008, 248 (03) :995-1003
[10]   Sixteen-detector row CT of abdomen and pelvis: Study for optimization of Z-axis modulation technique performed in 153 patients [J].
Kalra, MK ;
Maher, MM ;
Kamath, RS ;
Horiuchi, T ;
Toth, TL ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 233 (01) :241-249