Tissue mimicking materials for the detection of prostate cancer using shear wave elastography: A validation study

被引:32
作者
Cao, Rui [1 ]
Huang, Zhihong [1 ]
Varghese, Tomy [2 ]
Nabi, Ghulam [3 ,4 ]
机构
[1] Univ Dundee, Sch Engn Phys & Math, Dundee DD1 4HN, Scotland
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
[3] Sch Med, Acad Sect Urol, Dundee DD1 9SY, Scotland
[4] Ninewells Hosp, Dundee DD1 9SY, Scotland
关键词
ultrasound; elastography; prostate cancer; strain; shear wave elastography; ACOUSTIC RADIATION FORCE; INTENSITY-FOCUSED-ULTRASOUND; IN-VIVO DEMONSTRATION; SOFT-TISSUE; PHANTOMS; ELASTICITY; BREAST; FUSION;
D O I
10.1118/1.4773315
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Quantification of stiffness changes may provide important diagnostic information and aid in the early detection of cancers. Shear wave elastography is an imaging technique that assesses tissue stiffness using acoustic radiation force as an alternate to manual palpation reported previously with quasistatic elastography. In this study, the elastic properties of tissue mimicking materials, including agar, polyacrylamide (PAA), and silicone, are evaluated with an objective to determine material characteristics which resemble normal and cancerous prostate tissue. Methods: Acoustic properties and stiffness of tissue mimicking phantoms were measured using compressional mechanical testing and shear wave elastography using supersonic shear imaging. The latter is based on the principles of shear waves generated using acoustic radiation force. The evaluation included tissue mimicking materials (TMMs) within the prostate at different positions and sizes that could mimic cancerous and normal prostate tissue. Patient data on normal and prostate cancer tissues quantified using biopsy histopathology were used to validate the findings. Pathologist reports on histopathology were blinded to mechanical testing and elastographic findings. Results: Young's modulus values of 86.2 +/- 4.5 and 271.5 +/- 25.7 kPa were obtained for PAA mixed with 2% Al2O3 particles and silicone, respectively. Young's modulus of TMMs from mechanical compression testing showed a clear trend of increasing stiffness with an increasing percentage of agar. The silicone material had higher stiffness values when compared with PAA with Al2O3. The mean Young's modulus value in cancerous tissue was 90.5 +/- 4.5 kPa as compared to 93.8 +/- 4.4 and 86.2 +/- 4.5 kPa obtained with PAA with 2% Al2O3 phantom at a depth of 52.4 and 36.6 mm, respectively. Conclusions: PAA mixed with Al2O3 provides the most suitable tissue mimicking material for prostate cancer tumor material, while agar could form the surrounding background to simulate normal prostate tissue. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4773315]
引用
收藏
页数:9
相关论文
共 50 条
[41]   Evaluation of the Effect of Tissue Compression on the Results of Shear Wave Elastography Measurements [J].
Vachutka, Jaromir ;
Sedlackova, Zuzana ;
Furst, Tomas ;
Herman, Miroslav ;
Herman, Jan ;
Salzman, Richard ;
Dolezal, Ladislav .
ULTRASONIC IMAGING, 2018, 40 (06) :380-393
[42]   Characterisation of Prostate Lesions Using Transrectal Shear Wave Elastography (SWE) Ultrasound Imaging: A Systematic Review [J].
Anbarasan, Thineskrishna ;
Wei, Cheng ;
Bamber, Jeffrey C. ;
Barr, Richard G. ;
Nabi, Ghulam .
CANCERS, 2021, 13 (01) :1-15
[43]   Prostate Cancer: Diagnostic Performance of Real-time Shear-Wave Elastography [J].
Correas, Jean-Michel ;
Tissier, Anne-Marie ;
Khairoune, Ahmed ;
Vassiliu, Viorel ;
Mejean, Arnaud ;
Helenon, Olivier ;
Memo, Richard ;
Barr, Richard G. .
RADIOLOGY, 2015, 275 (01) :280-289
[44]   PLANTAR SOFT TISSUE CHARACTERIZATION USING REVERBERANT SHEAR WAVE ELASTOGRAPHY: A PROOF-OF-CONCEPT STUDY [J].
Romero, Stefano E. ;
Naemi, Roozbeh ;
Flores, Gilmer ;
Allan, David ;
Ormachea, Juvenal ;
Gutierrez, Evelyn ;
Casado, Fanny L. ;
Castaneda, Benjamin .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2022, 48 (01) :35-46
[45]   VISCOELASTICITY MEASUREMENT IN RAT LIVERS USING SHEAR-WAVE US ELASTOGRAPHY [J].
Sugimoto, Katsutoshi ;
Moriyasu, Fuminori ;
Oshiro, Hisashi ;
Takeuchi, Hirohito ;
Yoshimasu, Yu ;
Kasai, Yoshitaka ;
Furuichi, Yoshihiro ;
Itoi, Takao .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (09) :2018-2024
[46]   Feasibility of monitoring HIFU prostate cancer therapy using elastography [J].
Souchon, R ;
Chapelon, JY ;
Bertrand, M ;
Kallel, F ;
Ophir, J .
MEDICAL IMAGING 2001: ULTRASONIC IMAGING AND SIGNAL PROCESSING, 2001, 4325 :392-399
[47]   Elastic modulus contrast enhancement in shear wave imaging using mechanical nonlinearity: In vitro tissue mimicking phantom study [J].
Park, Dae Woo ;
Kim, Kang .
2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, :2335-2338
[48]   Shear wave vibrometry evaluation in transverse isotropic tissue mimicking phantoms and skeletal muscle [J].
Aristizabal, Sara ;
Amador, Carolina ;
Qiang, Bo ;
Kinnick, Randall R. ;
Nenadic, Ivan Z. ;
Greenleaf, James F. ;
Urban, Matthew W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (24) :7735-7752
[49]   Viscoelastic tissue mimicking phantom validation study with shear wave elasticity imaging and viscoelastic spectroscopy [J].
Amador, Carolina ;
Kinnick, Randall R. ;
Urban, Matthew W. ;
Fatemi, Mostafa ;
Greenleaf, James F. .
2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
[50]   Kelvin-Voigt Parameters Reconstruction of Cervical Tissue-Mimicking Phantoms Using Torsional Wave Elastography [J].
Callejas, Antonio ;
Gomez, Antonio ;
Faris, Inas H. ;
Melchor, Juan ;
Rus, Guillermo .
SENSORS, 2019, 19 (15)