Geometrical Enhancements for Abdominal Aortic Stent-Grafts

被引:35
作者
Molony, David S. [1 ]
Callanan, Anthony [1 ]
Morris, Liam G. [2 ]
Doyle, Barry J. [1 ]
Walsh, Michael T. [1 ]
McGloughlin, Tim M. [1 ]
机构
[1] Univ Limerick, CABER, Mat & Surface Sci Inst, Limerick, Ireland
[2] Galway Med Technol Ctr, Galway Mayo Inst Technol, Galway, Ireland
关键词
endovascular aneurysm repair; stent-graft; tapered graft; computational fluid dynamics; bifurcation; secondary flow; wall shear stress; out-of-plane curvature;
D O I
10.1583/08-2388.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To compare the function of 2 stent-graft designs for endovascular abdominal aortic aneurysm repair. Methods: Computational fluid dynamics was used to investigate the performance of a conventional stent-graft versus one with a novel tapered configuration (equal area ratios at the inlet and bifurcation). Idealized geometries (uniplanar) were formed first for both devices. To mimic the clinical setting with pulsatile blood flow, a realistic model (multiplanar) was created for the conventional stent-graft based on computed tomography scans from 3 patients with different aortic geometries. A similar model was created for the tapered stent-graft by mimicking the deployment of the conventional stent-graft through its centerline. Results: The tapered stent-graft model demonstrated reduced secondary flow vortices and wall shear stresses in the iliac limbs compared to the conventional graft in the idealized scenario. The drag forces in the idealized models were similar for both designs, though the tapered stent-graft showed a 4% reduction. Flow was split more evenly between the tapered stent-graft limbs in the realistic scenario. Conclusion: The novel tapered design reduced flow velocities and secondary flows due to its smooth trunk-to-limb transition, while also splitting the flow between the iliac limbs more evenly. In multiplanar models, the out-of-plane curvature was the greatest cause of skewed flow, which reduced the benefits of the tapered stent-graft.
引用
收藏
页码:518 / 529
页数:12
相关论文
共 27 条
[21]   Evidence suggests rigid aortic grafts increase systolic blood pressure: Results of a preliminary study [J].
O'Brien, T. ;
Morris, L. ;
McGloughlin, T. .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (01) :109-115
[22]  
Parodi J C, 1991, Ann Vasc Surg, V5, P491, DOI 10.1007/BF02015271
[23]   Abdominal aortic aneurysm [J].
Sakalihasan, N ;
Limet, R ;
Defawe, OD .
LANCET, 2005, 365 (9470) :1577-1589
[24]   Compliant biomechanics of abdominal aortic aneurysms: A fluid-structure interaction study [J].
Scotti, Christine M. ;
Finol, Ender A. .
COMPUTERS & STRUCTURES, 2007, 85 (11-14) :1097-1113
[25]   Numerical study on the effect of steady axial flow development in the human aorta on local shear stresses in abdominal aortic branches [J].
Shipkowitz, T ;
Rodgers, VGJ ;
Frazin, LJ ;
Chandran, KB .
JOURNAL OF BIOMECHANICS, 1998, 31 (11) :995-1007
[26]   Effect of improved endograft design on outcome of endovascular aneurysm repair [J].
Torella, F .
JOURNAL OF VASCULAR SURGERY, 2004, 40 (02) :216-221
[27]   Outflow boundary conditions for three-dimensional finite element modeling of blood flow and pressure in arteries [J].
Vignon-Clementel, Irene E. ;
Figueroa, C. Alberto ;
Jansen, Kenneth E. ;
Taylor, Charles A. .