Comparison of blood pressure measurements between an automated oscillometric device and a Hawksley random-zero sphygmomanometer in the northern Sweden MONICA study

被引:10
作者
Eriksson, Marie [1 ]
Carlberg, Bo [2 ]
Jansson, Jan-Hakan [2 ,3 ]
机构
[1] Umea Univ, Umea Sch Business & Econ, Dept Stat, SE-90187 Umea, Sweden
[2] Umea Univ, Dept Publ Hlth & Clin Med, SE-90187 Umea, Sweden
[3] Skelleftea Hosp, Dept Med, Skelleftea, Sweden
基金
瑞典研究理事会;
关键词
blood pressure; blood pressure measurement; blood pressure monitoring; Omron; oscillometric; random-zero; HYPERTENSION; MERCURY; VALIDATION; AGREEMENT; PROTOCOL; SOCIETY;
D O I
10.1097/MBP.0b013e328356ef58
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The Hawksley random-zero sphygmomanometer (random-zero) has been used widely in epidemiological observation studies. This study compares blood pressure measurements using the random-zero with measurements using an automated oscillometric device and suggests a correction of the automated oscillometric measurements to enable comparisons of blood pressure levels over time. Methods The northern Sweden MONICA population survey 2009 included 1729 participants, 853 men and 876 women, 25-74 years old. Blood pressure was measured using both random-zero and an automated oscillometric device in all participants. The Omron M7 digital blood pressure monitor was used for automated oscillometric measurements. A linear mixed model was used to derive a formula to adjust the automated oscillometric readings. Results Automated oscillometric measurements of systolic blood pressure were generally lower than random-zero measurements in women [oscillometric mean 122.1 mmHg (95% confidence interval: 121.0-123.2) versus random-zero mean 124.4 mmHg (123.5-125.5)], whereas automated oscillometric measurements of systolic blood pressure were generally higher than random-zero measurements in men [oscillometric 131.1 mmHg mmHg (130.0-132.2) versus random-zero 129.0 mmHg (127.9-130.1)]. For diastolic blood pressure, automated oscillometric measurements were higher in both women [oscillometric 79.9 mmHg (79.2-80.5) versus random-zero 76.7 mmHg (76.0-77.4)] and men [oscillometric 83.1 mmHg (82.4-83.8) vs. random-zero 81.2 mmHg (80.6-81.9)]. The difference also varied with age and order of measurement. Adjustment of the automated oscillometric measurements using mixed model regression coefficients produced estimates of blood pressure that were close to the random-zero measurements. Conclusion Blood pressure measurements using an automated oscillometric device differ from those with random-zero, but the oscillometric measurements can be adjusted, on the basis of sex, age and measurement order, to be similar to the random-zero measurements. Blood Press Monit 17:164-170 (C) 2012 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:164 / 170
页数:7
相关论文
共 22 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Difference in blood pressure readings with mercury and automated devices: impact on hypertension prevalence estimates in Dar es Salaam, Tanzania [J].
Chiolero, Arnaud ;
Gervasoni, Jean-Pierre ;
Rwebogora, Anne ;
Balampama, Marianna ;
Paccaud, Fred ;
Bovet, Pascal .
EUROPEAN JOURNAL OF EPIDEMIOLOGY, 2006, 21 (06) :427-433
[3]  
Cnaan A, 1997, STAT MED, V16, P2349, DOI 10.1002/(SICI)1097-0258(19971030)16:20<2349::AID-SIM667>3.0.CO
[4]  
2-E
[5]   Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol [J].
Coleman, Andrew ;
Steel, Stephen ;
Freeman, Paul ;
de Greeff, Annemarie ;
Shennan, Andrew .
BLOOD PRESSURE MONITORING, 2008, 13 (01) :49-54
[6]   MEASUREMENT ERROR IN THE HAWKSLEY RANDOM ZERO SPHYGMOMANOMETER - WHAT DAMAGE HAS BEEN DONE AND WHAT CAN WE LEARN [J].
CONROY, RM ;
OBRIEN, E ;
OMALLEY, K ;
ATKINS, N .
BRITISH MEDICAL JOURNAL, 1993, 306 (6888) :1319-1322
[7]   Determinants of white-coat hypertension [J].
Dolan, E ;
Stanton, A ;
Atkins, N ;
Den Hond, E ;
Thijs, L ;
McCormack, P ;
Staessen, J ;
O'Brien, E .
BLOOD PRESSURE MONITORING, 2004, 9 (06) :307-309
[8]   Large improvements in major cardiovascular risk factors in the population of northern Sweden: the MONICA study 1986-2009 [J].
Eriksson, M. ;
Holmgren, L. ;
Janlert, U. ;
Jansson, J. -H. ;
Lundblad, D. ;
Stegmayr, B. ;
Soderberg, S. ;
Eliasson, M. .
JOURNAL OF INTERNAL MEDICINE, 2011, 269 (02) :219-231
[10]   THE HAWKSLEY RANDOM ZERO SPHYGMOMANOMETER - SHOULD BE ABANDONED [J].
LAWSON, M ;
JOHNSTON, A .
BRITISH MEDICAL JOURNAL, 1993, 307 (6896) :123-123