Magnetically driven middle ear ossicles with laser vibrometry as a new diagnostic tool to quantify ossicular fixation

被引:11
|
作者
Peacock, John [1 ]
Dirckx, Joris [1 ]
von Unge, Magnus [2 ,3 ]
机构
[1] Univ Antwerp, Lab Biomed Phys, B-2020 Antwerp, Belgium
[2] Akershus Univ Hosp, Dept Otorhinolaryngol, Oslo, Norway
[3] Univ Oslo, Oslo, Norway
关键词
Otitis media; ear surgery; incus fixation; umbo long process; CHAIN;
D O I
10.3109/00016489.2013.841990
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Conclusion: Information on the degree of incus fixation can be gathered by measuring the ratio of incus to umbo long process velocity through the ear canal. Objectives: To test a new method of quantifying partial ossicular fixation in an ear with an elevated tympanic membrane. Methods: Measurements were made on four fresh-frozen human temporal bones. After elevating the tympanic membrane a small magnet was attached to the manubrium and an electromagnetic excitation coil was used to vibrate the ossicles. The vibration response of the tip of the incus long process and the umbo were measured before and after artificially fixating the incus to the lateral attic wall. Results: Partial incus fixation resulted in a decrease in both the incus and umbo velocities, with the incus velocity being more severely reduced. The decreased ratio of their vibrations is a clear indicator of the degree of incus fixation.
引用
收藏
页码:352 / 357
页数:6
相关论文
共 8 条
  • [1] Laser Doppler Velocimeter Measurement of Acoustically and Magnetically Driven Middle Ear Ossicles
    Peacock, John
    von Unge, Magnus
    Dirckx, Joris J. J.
    OPTICAL MEASUREMENT TECHNIQUES FOR STRUCTURES & SYSTEMS2 (OPTIMESS2012), 2013, : 311 - 317
  • [2] Measurement of the vibration of the middle ear ossicles with removed eardrum: A method for quantification of ossicular fixation
    Peacock, John
    von Unge, Magnus
    Dirckx, Joris
    MEDICAL ENGINEERING & PHYSICS, 2013, 35 (12) : 1786 - 1792
  • [3] Magnetically Driven Middle Ear Ossicles for Optical Measurement of Vibrations in a Ear with Opened Eardraum
    Peacock, John
    von Unge, Magnus
    Dirckx, Joris
    10TH INTERNATIONAL CONFERENCE ON VIBRATION MEASUREMENTS BY LASER AND NONCONTACT TECHNIQUES - AIVELA 2012, 2012, 1457 : 259 - 265
  • [4] Magnetically driven middle ear ossicles for optical measurement of vibrations in an ear with opened tympanic membrane
    Peacock, J.
    von Unge, M.
    Dirckx, J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (12):
  • [5] Towards quantitative diagnosis of ossicular fixation: Measurement of stapes fixations using magnetically driven ossicles in human temporal bones
    Peacock, John
    Dirckx, Joris
    Von Unge, Magnus
    ACTA OTO-LARYNGOLOGICA, 2015, 135 (09) : 880 - 885
  • [6] EVALUATION OF LASER VIBROMETRY AS DIAGNOSTIC UTILITY BY MEANS OF A SIMULATION MODEL OF THE MIDDLE EAR
    Bornitz, Matthias
    Lasurashvili, Nikoloz
    Hardtke, Hans-Juergen
    Zahnert, Thomas
    MIDDLE EAR MECHANICS IN RESEARCH AND OTOLOGY, 2007, : 222 - +
  • [7] Minimally invasive laser vibrometry (MIVIB) with a floating mass transducer - A new method for objective evaluation of the middle ear demonstrated on stapes fixation
    Wales, Jeremy
    Gladine, Kilian
    Van de Heyning, Paul
    Topsakal, Vedat
    von Unge, Magnus
    Dirckx, Joris
    HEARING RESEARCH, 2018, 357 : 46 - 53
  • [8] A new tool for testing ossicular mobility during middle ear surgery: Preliminary report of four cases
    Hato, Naohito
    Kohno, Hisashi
    Okada, Masahiro
    Hakuba, Nobuhiro
    Gyo, Kiyofumi
    Iwakura, Takashi
    Tateno, Makoto
    OTOLOGY & NEUROTOLOGY, 2006, 27 (05) : 592 - 595