Low-dose single-energy material decomposition in radiography using a sparse-view computed tomography scan

被引:1
作者
Kim, G. [1 ]
Lim, Y. [1 ]
Park, J. [1 ]
Cho, H. [1 ]
Park, C. [1 ]
Kim, K. [1 ]
Kang, S. [1 ]
Lee, D. [1 ]
Park, S. [1 ]
Lim, H. [1 ]
Lee, H. [1 ]
Jeon, D. [1 ]
Kim, W. [1 ]
Seo, C. [1 ]
Lee, M. [2 ]
机构
[1] Yonsei Univ, Dept Radiat Convergence Engn, Wonju, South Korea
[2] Asan Med Ctr, Dept Radiat Oncol, Seoul, South Korea
基金
新加坡国家研究基金会;
关键词
Computed tomography; dictionary-learning; dual-energy material decomposition; low-dose single-energy material decomposition; SPECTRAL CT; IMAGE; REPRESENTATIONS; APPARATUS;
D O I
10.1080/10739149.2018.1556685
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Dual-energy material decomposition (DEMD) is a well-established theoretical x-ray technique that uses low- and high-kilovoltage radiographs to separate soft tissue and bone in radiography and computed tomography (CT). However, it requires double exposures that result in increased patient radiation doses, causes increases in the execution time, and generates errors due to misregistration attributed to the patient motion between two scans. In this study, we investigated a low-dose, single-energy material decomposition (LSEMD) method in radiography using a sparse-view CT scan where the attenuation length in the object was estimated from the CT image. We performed a systematic simulation and an experiment to demonstrate the feasibility of use of the LSEMD method in radiography. Only 60 projections, far fewer than those required by the Nyquist sampling theory, were acquired at an x-ray tube voltage of 80 kV(p), and were used to reconstruct a sparse-view CT image with a state-of-the-art dictionary-learning (DL) algorithm. We investigated the image performance of the LSEMD and compared the elicited results with those obtained with the use of DEMD (80 kV(p) and 120 kV(p) were used). Our results indicate that the DL algorithm produced high-quality sparse-view CT images. Accordingly, the LSEMD method yielded material decomposition results that were very similar to the results elicited by the conventional DEMD method in radiography.
引用
收藏
页码:325 / 340
页数:16
相关论文
共 21 条
  • [1] Comparison of dual energy detector system performance
    Alvarez, RE
    Seibert, JA
    Thompson, SK
    [J]. MEDICAL PHYSICS, 2004, 31 (03) : 556 - 565
  • [2] ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY
    ALVAREZ, RE
    MACOVSKI, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) : 733 - 744
  • [3] [Anonymous], 1995, NIST XRAY ATTENUATIO
  • [4] A METHOD FOR SELECTIVE TISSUE AND BONE VISUALIZATION USING DUAL ENERGY SCANNED PROJECTION RADIOGRAPHY
    BRODY, WR
    BUTT, G
    HALL, A
    MACOVSKI, A
    [J]. MEDICAL PHYSICS, 1981, 8 (03) : 353 - 357
  • [5] Image denoising via sparse and redundant representations over learned dictionaries
    Elad, Michael
    Aharon, Michal
    [J]. IEEE TRANSACTIONS ON IMAGE PROCESSING, 2006, 15 (12) : 3736 - 3745
  • [6] QUANTITATIVE BONE-MINERAL ANALYSIS USING DUAL ENERGY COMPUTED TOMOGRAPHY
    GENANT, HK
    BOYD, D
    [J]. INVESTIGATIVE RADIOLOGY, 1977, 12 (06) : 545 - 551
  • [7] Material differentiation by dual energy CT: initial experience
    Johnson, Thorsten R. C.
    Krauss, Bernhard
    Sedlmair, Martin
    Grasruck, Michael
    Bruder, Herbert
    Morhard, Dominik
    Fink, Christian
    Weckbach, Sabine
    Lenhard, Miriam
    Schmidt, Bernhard
    Flohr, Thomas
    Reiser, Maximilian F.
    Becker, Christoph R.
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (06) : 1510 - 1517
  • [8] Johnson TRC, 2011, MED RADIOL DIAGN IMA, P1, DOI 10.1007/978-3-642-01740-7
  • [9] EVALUATION OF A PROTOTYPE DUAL-ENERGY COMPUTED TOMOGRAPHIC APPARATUS .1. PHANTOM STUDIES
    KALENDER, WA
    PERMAN, WH
    VETTER, JR
    KLOTZ, E
    [J]. MEDICAL PHYSICS, 1986, 13 (03) : 334 - 339
  • [10] Kim G., J KOREAN PHYS SOC