Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis

被引:45
作者
Chang, Hao-Sheng [1 ,2 ]
Chen, Yi-Chin [3 ]
Hsieh, Yao-Dung [1 ,4 ]
Hsu, Ming-Lun [5 ]
机构
[1] Kaohsiung Vet Gen Hosp, Dept Dent, Kaohsiung, Taiwan
[2] Shu Zen Coll Med & Management, Kaohsiung, Taiwan
[3] Met Ind Res & Dev Ctr, Med Device & Optoelect Equipment Dept, Kaohsiung, Taiwan
[4] Natl Def Natl Med Ctr, Sch Dent, Taipei, Taiwan
[5] Natl Yang Ming Univ, Dept Dent, Taipei 112, Taiwan
关键词
dental implant; finite element analysis; platform switching; PLATFORM-SWITCHING TECHNIQUE; ORAL IMPLANTS; COMPACT-BONE; OCCLUSAL OVERLOAD; CORTICAL BONE; DESIGN; PROSTHESES; TISSUE; COMPLICATIONS; STABILITY;
D O I
10.1016/j.jds.2012.04.006
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background/purpose: This study investigates the stress distributions in an implant, abutment, and crown restoration with different implant systems, in various bone qualities, and with different loading protocols using a three-dimensional finite element model. Materials and methods: Eight three-dimensional finite element models with 16 test conditions containing four types of dental implants embedded in two different bone qualities (types II and IV) under 100-N axial and 30 oblique loading forces were applied to analyze the stress distribution in the crown restoration, abutment, abutment screw, implant, and supporting bone. Results: The highest maximum von Mises stress was noted in the abutment of a tissue-level implant with the Straumann system (1203.04 MPa) under a 30 degrees oblique loading force. With axial load application, stresses in the screw and abutment of the NobelBiocare system were greater in the tissue-level implant (MK III) than in the bone-level implant (Active). The von Mises stresses in the cortical bone were mostly greater in the tissue-level implant (MK III) than in the bone-level implant (Active) of the NobelBiocare system. However, von Mises stresses in cancellous bone were mostly greater in the bone-level implant (Active) than in the tissue-level implant (MK III) of the NobelBiocare system. Conclusion: Within the limitations of the present study, the Straumann system produced greater stresses than the NobelBiocare system in type IV cortical bone, but they were almost equal in type II bone. By contrast, the NobelBiocare system produced greater stresses than the Straumann system in cancellous bone, regardless of the type of loading angle or bone quality. Copyright (C) 2013, Association for Dental Sciences of the Republic of China. Published by Elsevier Taiwan LLC. All rights reserved.
引用
收藏
页码:261 / 271
页数:11
相关论文
共 42 条
[1]   Finite element analysis of non-axial versus axial loading of oral implants in the mandible of the dog [J].
Barbier, L ;
Vander Sloten, J ;
Krzesinski, G ;
Schepers, E ;
Van Der Perre, G .
JOURNAL OF ORAL REHABILITATION, 1998, 25 (11) :847-858
[2]  
Barbier L, 1997, INT J ORAL MAX IMPL, V12, P215
[3]   Influence of platform switching on crestal bone changes at non-submerged titanium implants: a histomorphometrical study in dogs [J].
Becker, Juergen ;
Ferrari, Daniel ;
Herten, Monika ;
Kirsch, Axel ;
Schaer, Alex ;
Schwarz, Frank .
JOURNAL OF CLINICAL PERIODONTOLOGY, 2007, 34 (12) :1089-1096
[4]   Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis [J].
Bozkaya, D ;
Muftu, S ;
Muftu, A .
JOURNAL OF PROSTHETIC DENTISTRY, 2004, 92 (06) :523-530
[5]   Bone microdamage acid skeletal fragility in osteoporotic and stress fractures [J].
Burr, DB ;
Forwood, MR ;
Fyhrie, DP ;
Martin, B ;
Schaffler, MB ;
Turner, CH .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (01) :6-15
[6]   Platform switching and marginal bone-level alterations: the results of a randomized-controlled trial [J].
Canullo, Luigi ;
Fedele, Giampiero Rossi ;
Iannello, Giuliano ;
Jepsen, Soren .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (01) :115-121
[7]   Role of mechanical environment and implant design on bone tissue differentiation:: current knowledge and future contexts [J].
Çehreli, M ;
Sahin, S ;
Akça, K .
JOURNAL OF DENTISTRY, 2004, 32 (02) :123-132
[8]   Failures in implant dentistry [J].
Chee, W. ;
Jivraj, S. .
BRITISH DENTAL JOURNAL, 2007, 202 (03) :123-129
[9]   Bone Response to Loaded Implants With Non-Matching Implant-Abutment Diameters in the Canine Mandible [J].
Cochran, David L. ;
Bosshardt, Dieter D. ;
Grize, Leticia ;
Higginbottom, Frank L. ;
Jones, Archie A. ;
Jung, Ronald E. ;
Wieland, Marco ;
Dard, Michel .
JOURNAL OF PERIODONTOLOGY, 2009, 80 (04) :609-617
[10]  
Craig RG, 1980, RESTORATIVE DENT MAT, P286