Comparison of the Polar S810i monitor and the ECG for the analysis of heart rate variability in the time and frequency domains

被引:176
作者
Vanderlei, L. C. M. [1 ]
Silva, R. A. [2 ]
Pastre, C. M. [1 ]
Azevedo, F. M. [3 ]
Godoy, M. F. [4 ]
机构
[1] UNESP, FCT, Dept Fisioterapia, BR-19060900 Presidente Prudente, SP, Brazil
[2] UNESP, FCT, Curso Especializacao Fisioterapia, BR-19060900 Presidente Prudente, SP, Brazil
[3] UNESP, FCT, Lab Fisioterapia Aplicada Movimento Humano, BR-19060900 Presidente Prudente, SP, Brazil
[4] Fac Med Sao Jose do Rio Preto, Dept Cardiol & Cirurgia Cardiovasc, Sao Jose Do Rio Preto, SP, Brazil
关键词
Heart rate variability; Autonomic nervous system; Exercise; Spectral analysis; ECG; Polar S810i monitor;
D O I
10.1590/S0100-879X2008005000039
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The aim of the present study was to compare heart rate variability (HRV) at rest and during exercise using a temporal series obtained with the Polar S810i monitor and a signal from a LYNX (R) signal conditioner (BIO EMG 1000 model) with a channel configured for the acquisition of ECG signals. Fifteen healthy subjects aged 20.9 +/- 1.4 years were analyzed. The subjects remained at rest for 20 min and performed exercise for another 20 min with the workload selected to achieve 60% of submaximal heart rate. RR series were obtained for each individual with a Polar S810i instrument and with an ECG analyzed with a biological signal conditioner. The HRV indices (rMSSD, pNN50, LFnu, HFnu, and LF/HF) were calculated after signal processing and analysis. The unpaired Student t-test and intraclass correlation coefficient were used for data analysis. No statistically significant differences were observed when comparing the values analyzed by means of the two devices for HRV at rest and during exercise. The intraclass correlation coefficient demonstrated satisfactory correlation between the values obtained by the devices at rest (pNN50 = 0.994; rMSSD = 0.995; LFnu = 0.978; HFnu = 0.978; LF/HF = 0.982) and during exercise (pNN50 = 0.869; rMSSD = 0.929; LFnu = 0.973; HFnu = 0.973; LF/HF = 0.942). The calculation of HRV values by means of temporal series obtained from the Polar S810i instrument appears to be as reliable as those obtained by processing the ECG signal captured with a signal conditioner.
引用
收藏
页码:854 / 859
页数:6
相关论文
共 50 条
  • [1] Comparison of Polar® RS800G3 heart rate monitor with Polar® S810i and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest
    da Costa de Rezende Barbosa, Marianne Penachini
    da Silva, Natalia Turri
    de Azevedo, Fabio Micolis
    Pastre, Carlos Marcelo
    Marques Vanderlei, Luiz Carlos
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2016, 36 (02) : 112 - 117
  • [2] Comparison of three mobile devices for measuring R-R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system
    Weippert, Matthias
    Kumar, Mohit
    Kreuzfeld, Steffi
    Arndt, Dagmar
    Rieger, Annika
    Stoll, Regina
    EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2010, 109 (04) : 779 - 786
  • [3] Comparison of three mobile devices for measuring R–R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system
    Matthias Weippert
    Mohit Kumar
    Steffi Kreuzfeld
    Dagmar Arndt
    Annika Rieger
    Regina Stoll
    European Journal of Applied Physiology, 2010, 109 : 779 - 786
  • [4] Comparison of Omega Wave System and Polar S810i to Detect R-R Intervals at Rest
    Parrado, E.
    Garcia, M. A.
    Ramos, J.
    Cervantes, J. C.
    Rodas, G.
    Capdevila, L.
    INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 2010, 31 (05) : 336 - 341
  • [5] Accuracy Assessment of Oura Ring Nocturnal Heart Rate and Heart Rate Variability in Comparison With Electrocardiography in Time and Frequency Domains: Comprehensive Analysis
    Cao, Rui
    Azimi, Iman
    Sarhaddi, Fatemeh
    Niela-Vilen, Hannakaisa
    Axelin, Anna
    Liljeberg, Pasi
    Rahmani, Amir M.
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2022, 24 (01)
  • [6] Validity and Reliability of Short-Term Heart-Rate Variability from the Polar S810
    Nunan, David
    Donovan, Gay
    Jakovljevic, Djordje G.
    Hodges, Lynette D.
    Sandercock, Gavin R. H.
    Brodie, David A.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2009, 41 (01) : 243 - 250
  • [7] TIME AND FREQUENCY-DOMAIN METHODS FOR HEART-RATE-VARIABILITY ANALYSIS - A METHODOLOGICAL COMPARISON
    LITVACK, DA
    OBERLANDER, TF
    CARNEY, LH
    SAUL, JP
    PSYCHOPHYSIOLOGY, 1995, 32 (05) : 492 - 504
  • [8] Time and Frequency Analysis of Heart Rate Variability Data in Heart Failure Patients
    Georgieva-Tsaneva, Galya N.
    INTERNATIONAL JOURNAL OF ADVANCED COMPUTER SCIENCE AND APPLICATIONS, 2019, 10 (11) : 456 - 462
  • [9] Post-exercise heart rate variability recovery: a time-frequency analysis
    Pecanha, Tiago
    de Paula-Ribeiro, Marcelle
    Nasario-Junior, Olivasse
    Perrout de Lima, Jorge Roberto
    ACTA CARDIOLOGICA, 2013, 68 (06) : 607 - 613
  • [10] Validity of the polar S810 heart rate monitor to measure R-R intervals at rest
    Gamelin, Franicois Xavier
    Berthoin, Serge
    Bosquet, Laurent
    MEDICINE & SCIENCE IN SPORTS & EXERCISE, 2006, 38 (05) : 887 - 893