Preclinical evaluation of an MRI-compatible pneumatic robot for angulated needle placement in transperineal prostate interventions

被引:39
作者
Tokuda, Junichi [1 ]
Song, Sang-Eun [1 ]
Fischer, Gregory S. [2 ]
Iordachita, Iulian I. [3 ]
Seifabadi, Reza [5 ]
Cho, Nathan B. [4 ]
Tuncali, Kemal [1 ]
Fichtinger, Gabor [5 ]
Tempany, Clare M. [1 ]
Hata, Nobuhiko [1 ]
机构
[1] Harvard Univ, Brigham & Womens Hosp, Sch Med, Dept Radiol, Boston, MA 02115 USA
[2] Worcester Polytech Inst, Worcester, MA 01609 USA
[3] Johns Hopkins Univ, Lab Computat Sensing & Robot, Baltimore, MD 21218 USA
[4] Johns Hopkins Univ, Dept Comp Sci, Baltimore, MD 21218 USA
[5] Queens Univ, Sch Comp, Kingston, ON K7L 3N6, Canada
关键词
MRI-guided therapy; Medical robotics; Prostate cancer; Biopsy; Brachytherapy; SYSTEM; BIOPSY; BRACHYTHERAPY; CANCER; INTEGRATION; TRACKING; GUIDANCE; ANGLE;
D O I
10.1007/s11548-012-0750-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose To evaluate the targeting accuracy of a small profile MRI-compatible pneumatic robot for needle placement that can angulate a needle insertion path into a large accessible target volume. Methods We extended our MRI-compatible pneumatic robot for needle placement to utilize its four degrees-of-freedom (4-DOF) mechanism with two parallel triangular structures and support transperineal prostate biopsies in a closed-bore magnetic resonance imaging (MRI) scanner. The robot is designed to guide a needle toward a lesion so that a radiologist can manually insert it in the bore. The robot is integrated with navigation software that allows an operator to plan angulated needle insertion by selecting a target and an entry point. The targeting error was evaluated while the angle between the needle insertion path and the static magnetic field was between -5.7 degrees and 5.7 degrees horizontally and between -5.7 degrees and 4.3 degrees vertically in the MRI scanner after sterilizing and draping the device. Results The robot positioned the needle for angulated insertion as specified on the navigation software with overall targeting error of 0.8 +/- 0.5mm along the horizontal axis and 0.8 +/- 0.8mm along the vertical axis. The two-dimensional root-mean-square targeting error on the axial slices as containing the targets was 1.4mm. Conclusions Our preclinical evaluation demonstrated that the MRI-compatible pneumatic robot for needle placement with the capability to angulate the needle insertion path provides targeting accuracy feasible for clinical MRI-guided prostate interventions. The clinical feasibility has to be established in a clinical study.
引用
收藏
页码:949 / 957
页数:9
相关论文
共 43 条
  • [1] Transperineal prostate biopsy under magnetic resonance image guidance: A needle placement accuracy study
    Blumenfeld, Philip
    Hata, Nobuhiko
    DiMaio, Simon
    Zou, Kelly
    Haker, Steven
    Fichtinger, Gabor
    Tempany, Clare M. C.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (03) : 688 - 694
  • [2] Chinzei K, 2000, LECT NOTES COMPUT SC, V1935, P921
  • [3] Feasibility of transperineal prostate biopsy under interventional magnetic resonance guidance
    Cormack, RA
    D'Amico, AV
    Hata, N
    Silverman, S
    Weinstein, M
    Tempany, CM
    [J]. UROLOGY, 2000, 56 (04) : 663 - 664
  • [4] Transperineal magnetic resonance image guided prostate biopsy
    D'Amico, AV
    Tempany, CM
    Cormack, R
    Hata, N
    Jinzaki, M
    Tuncali, K
    Weinstein, M
    Richie, JP
    [J]. JOURNAL OF UROLOGY, 2000, 164 (02) : 385 - 387
  • [5] Robot-assisted needle placement in open MRI: System architecture, integration and validation
    Dimaio, S. P.
    Pieper, S.
    Chinzei, K.
    Hata, N.
    Haker, S. J.
    Kacher, D. F.
    Fichtinger, G.
    Tempany, C. M.
    Kikins, R.
    [J]. COMPUTER AIDED SURGERY, 2007, 12 (01) : 15 - 24
  • [6] DiMaio S, 2007, LECT NOTES COMPUT SC, V4792, P50
  • [7] Elhawary H, 2007, LECT NOTES COMPUT SC, V4792, P604
  • [8] Elhawary H, 2006, LECT NOTES COMPUT SC, V4190, P519
  • [9] Prostate biopsy in the supine position in a standard 1.5-T scanner under real time MR-imaging control using a MR-compatible endorectal biopsy device
    Engelhard, K
    Hollenbach, HP
    Kiefer, B
    Winkel, A
    Goeb, K
    Engehausen, D
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (06) : 1237 - 1243
  • [10] MRI-compatible pneumatic robot for transperineal prostate needle placement
    Fischer, Gregory S.
    Iordachita, Iulian
    Csoma, Csaba
    Tokuda, Junichi
    DiMaio, Simon P.
    Tempany, Clare M.
    Hata, Nobuhiko
    Fichtinger, Gabor
    [J]. IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2008, 13 (03) : 295 - 305