Quality Assurance of a Two-Dimensional CCD Detector System Applied in Dosimetry

被引:8
|
作者
Sampaio, Francisco G. A. [1 ]
Del Lama, Lucas S. [1 ]
Sato, Rodrigo [2 ]
de Oliveira, David M. M. [1 ]
Czelusniak, Caroline [1 ]
de Oliveira, Lucas N. [3 ]
de Almeida, Adelaide [1 ]
机构
[1] Univ Sao Paulo, Dept Fis FFCLRP, BR-14015400 Ribeirao Preto, SP, Brazil
[2] Univ Tsukuba, Grad Sch Pure & Appl Sci, Tsukuba, Ibaraki 3058577, Japan
[3] Inst Fed Educ Ciencia & Tecnol Goias, BR-75400000 Inhumas, Go, Brazil
关键词
Charge-coupled device detector system; Fricke Xylenol Gel; modulation transfer function; radiotherapy dosimetry; FRICKE XYLENOL GEL; MODULATION TRANSFER-FUNCTION; CHARGE-COUPLED-DEVICES;
D O I
10.1109/TNS.2013.2240464
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Nowadays, charge-coupled devices (CCDs) are applied for many technological purposes, including areas as astronomy, medical diagnosis and the ones related to light detecting sensors. One of the main advantages of a digital detector is its ability to store and rapidly transfer information to digital equipments, where data can be computed (post-processing). In fact, this is a relevant feature for medical applications, specially for imaging fields, in which the radiation exposure time is highly controlled when different image quality adjustments are necessary. The main goal of this article is to demonstrate a CCD application for optical density-based dosimeters, where high accuracy and spatial resolution are required. An optical prototype was developed considering a visible light source and a CCD camera, allied with a computational processing routine. The system resolution was inferred through two different methodologies, which support a spatial resolution better than 0.5 mm: 1) a radiological test system (line pair gauge) and 2) the modulation transfer function (MTF). Results also indicate that, due to CCD suitable resolution, one and two dimension responses can be acquired. Considering the easiness and fastness of the process, the developed system exhibits potential applicability for radiotherapy dosimetry.
引用
收藏
页码:810 / 816
页数:7
相关论文
共 50 条
  • [1] Two-dimensional thermoluminescence dosimetry system for proton beam quality assurance
    Gajewski, J.
    Klosowski, M.
    Olko, P.
    RADIATION MEASUREMENTS, 2016, 90 : 224 - 227
  • [2] Application of Two-Dimensional Detector Arrays to the Delivery Quality Assurance in Breast IMRT
    Cho, Shin-Haeng
    Jeong, Jae-Uk
    Yoon, Mee Sun
    Nam, Taek-Keun
    Ahn, Sung-Ja
    Song, Ju-Young
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2020, 76 (05) : 430 - 435
  • [3] Application of Two-Dimensional Detector Arrays to the Delivery Quality Assurance in Breast IMRT
    Shin-Haeng Cho
    Jae-Uk Jeong
    Mee Sun Yoon
    Taek-Keun Nam
    Sung-Ja Ahn
    Ju-Young Song
    Journal of the Korean Physical Society, 2020, 76 : 430 - 435
  • [4] The study of a auto ellipsometer system by using a two-dimensional CCD array detector
    You, HY
    Jia, JH
    Chen, JK
    Han, T
    Ni, WM
    Wang, SY
    Li, J
    Yang, YM
    Chen, LY
    JOURNAL OF INFRARED AND MILLIMETER WAVES, 2003, 22 (01) : 45 - 50
  • [5] Quality assurance in IMRT: A comparison between two-dimensional ionization chamber array and film dosimetry
    Seguro Fernandez, A.
    Urena Llinares, A.
    Ramos Caballero, L. J.
    Castro Ramirez, I.
    Quinones Rodriguez, L. A.
    Mora Melendez, R.
    Angulo Pain, E.
    Iborra Oquendo, M. A.
    EUROPEAN JOURNAL OF CANCER, 2013, 49 : S230 - S230
  • [6] Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniques
    Boon, SN
    van Luijk, P
    Böhringer, T
    Coray, A
    Lomax, A
    Pedroni, E
    Schaffner, B
    Schippers, JM
    MEDICAL PHYSICS, 2000, 27 (10) : 2198 - 2208
  • [7] In situ two-dimensional micro-imaging XAFS with CCD detector
    Tanida, H.
    Yamashige, H.
    Orikasa, Y.
    Fujimoto, T.
    Oishi, M.
    Murayama, H.
    Arai, H.
    Katayama, M.
    Inada, Y.
    Ohta, T.
    Uchimoto, Y.
    Ogumi, Z.
    15TH INTERNATIONAL CONFERENCE ON X-RAY ABSORPTION FINE STRUCTURE (XAFS15), 2013, 430
  • [8] Dosimetric characterization of a new two-dimensional diode detector array used for stereotactic radiosurgery quality assurance
    Yasui, K.
    Saito, Y.
    Ogawa, S.
    Hayashi, N.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (02): : 281 - 289
  • [9] Improvement of thermoluminescent sheet for a two-dimensional dosimetry system
    Konnai, Akiko
    Ozasa, Naoto
    Ishikawa, Yuh-zoh
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 998 - 1003
  • [10] Novel method for quality assurance of two-dimensional pattern fidelity
    Maeda, Shimon
    Ogawa, Ryuji
    Shibazaki, Seiji
    Nakajima, Tadashi
    DESIGN FOR MANUFACTURABILITY THROUGH DESIGN-PROCESS INTEGRATION, 2007, 6521