Physical evaluation of a system for direct digital intra-oral radiography based on a charge-coupled device

被引:15
作者
Yoshiura, K [1 ]
Stamatakis, HC [1 ]
Welander, U [1 ]
McDavid, WD [1 ]
Shi, XQ [1 ]
Ban, S [1 ]
Kawazu, T [1 ]
Tatsumi, M [1 ]
Kanda, S [1 ]
机构
[1] Kyushu Univ, Fac Dent, Dept Oral & Maxillofacial Radiol, Higashi Ku, Fukuoka 8128582, Japan
关键词
radiography; dental; digital radiography; technology; radiologic;
D O I
10.1038/sj.dmfr.4600457
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To determine technical properties of a direct digital intra-oral radiographic system, the Dixel(R)(J Morita Corporation, Kyoto, Japan). Methods: A dose response function and the dark current were calculated from two series of exposures to a homogeneous radiation field. The line spread function (LSF) and the modulation transfer function (MTF) were determined from radiographs of an edge. The noise power spectrum (NPS) was determined at three exposures from radiographs exposed to homogeneous radiation fields. Noise equivalent quanta (NEQ) were calculated from the one-dimensional NPS and the MTF. The detective quantum efficiency (DQE) was determined from the NEQ and a representative value of the photon fluence. Signal-to-noise ratios (SNR) were calculated from the NEQs and different signal contrasts. Results: The dose response function demonstrated a slight curvature. There was no effect of the dark current. NPS ranged from 10(-5)-10(-7) mm(2) depending on exposure and frequency. At a peak of about 2 cycles/mm the DQE is on an average about 30 per cent. SNRs are favorable. Conclusion: The technical properties found in this study indicate that the Dixel(R)system is suitable for intra-oral dental radiography.
引用
收藏
页码:277 / 283
页数:7
相关论文
共 24 条
[1]  
ARAI Y, 1994, DENT RADIOL, V34, P229
[2]  
BAN S, 1995, QUINTESSENCE, V14, P1716
[3]  
BARTEN PGJ, 1992, SPIE P, V1666, P57
[4]   The imaging performance of a storage phosphor system for dental radiography [J].
Brettle, DS ;
Workman, A ;
Ellwood, RP ;
Launders, JH ;
Horner, K ;
Davies, RM .
BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (819) :256-261
[5]  
DOI K, 1982, HLTH HUMAN SERVICES
[6]  
Farman A G, 1995, Quintessence Int, V26, P399
[7]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[8]  
ICRU, 1996, 54 ICRU
[9]  
JOHNS HE, 1983, PHYSICS RADIOLOGY
[10]   Physical imaging performance of a compact computed radiography acquisition device [J].
Kengyelics, SM ;
Launders, JH ;
Cowen, AR .
MEDICAL PHYSICS, 1998, 25 (03) :354-360