Development of a wireless intra-ocular pressure monitoring system for incorporation into a therapeutic glaucoma drainage implant

被引:0
作者
Kakaday, Tarun [1 ]
Plunkett, Malcolm [3 ]
McInnes, Steven [2 ]
Li, Jim S. Jimmy [1 ]
Voelcker, Nicolas H. [2 ]
Craig, Jamie E. [4 ]
机构
[1] Flinders Univ S Australia, Sch Comp Sci Engn & Math, Adelaide, SA 5001, Australia
[2] Flinders Univ S Australia, Sch Chem Phys & Earth Sci, Adelaide, SA 5001, Australia
[3] Ellex Med Lasers R&D, Adelaide, SA, Australia
[4] Flinders Med Ctr, Dept Ophthalmol, Adelaide, SA, Australia
来源
BIOMEDICAL APPLICATIONS OF MICRO- AND NANOENGINEERING IV AND COMPLEX SYSTEMS | 2008年 / 7270卷
基金
澳大利亚研究理事会;
关键词
Glaucoma; intraocular pressure (IOP); glaucoma drainage implant (GDI); micro-electro-mechanical systems (MEMS); capacitive pressure sensors; reader antenna; printed circuit boards (PCB's);
D O I
10.1117/12.810647
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Glaucoma is a common cause of blindness. Wireless, continuous monitoring of intraocular pressure (IOP) is an important, unsolved goal in managing glaucoma. An IOP monitoring system incorporated into a glaucoma drainage implant (GDI) overcomes the design complexity associated with incorporating a similar system in a more confined space within the eye. The device consists of a micro-electro-mechanical systems (MEMS) based capacitive pressure sensor integrated with an inductor printed directly onto a polyimide printed circuit board (PCB). The device is designed to be incorporated onto the external plate of a therapeutic GDI. The resonance frequency changes as a function of IOP, and is tracked remotely using a spectrum analyzer. A theoretical model for the reader antenna was developed to enable maximal inductive coupling with the IOP sensor implant. Pressure chamber tests indicate that the sensor implant has adequate sensitivity in the IOP range with excellent reproducibility over time. Additionally, we show that sensor sensitivity does not change significantly after encapsulation with polydimethylsiloxane (PDMS) to protect the device from fluid environment. In vitro experiments showed that the signal measured wirelessly through sheep corneal and scleral tissue was adequate indicating potential for using the system in human subjects.
引用
收藏
页数:11
相关论文
共 11 条
[1]   A wireless batch sealed absolute capacitive pressure sensor [J].
Akar, O ;
Akin, T ;
Najafi, K .
SENSORS AND ACTUATORS A-PHYSICAL, 2001, 95 (01) :29-38
[2]  
EGGERS T, 2000, 1 ANN INT IEEE EMBS
[3]   Central corneal thickness, radius of the corneal curvature and intraocular pressure in normal subjects using non-contact techniques:: Reykjavik Eye Study [J].
Eysteinsson, T ;
Jonasson, F ;
Sasaki, H ;
Arnarsson, A ;
Sverrisson, T ;
Sasaki, K ;
Stefánsson, E .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2002, 80 (01) :11-15
[4]   First steps toward noninvasive intraocular pressure monitoring with a sensing contact lens [J].
Leonardi, M ;
Leuenberger, P ;
Bertrand, D ;
Bertsch, A ;
Renaud, P .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (09) :3113-3117
[5]   Understanding trabecular meshwork physiology: A key to the control of intraocular pressure? [J].
Llobet, A ;
Gasull, X ;
Gual, A .
NEWS IN PHYSIOLOGICAL SCIENCES, 2003, 18 :205-209
[6]   Design and application of a wireless, passive, resonant-circuit environmental monitoring sensor [J].
Ong, KG ;
Grimes, CA ;
Robbins, CL ;
Singh, RS .
SENSORS AND ACTUATORS A-PHYSICAL, 2001, 93 (01) :33-43
[7]  
Reinhold C., 2007, J COMMUNICATIONS, V2, P14, DOI DOI 10.4304/jcm.2.6.14-23
[8]   Passive telemetric readout system [J].
Sajeeda ;
Kaiser, Todd J. .
IEEE SENSORS JOURNAL, 2006, 6 (05) :1340-1345
[9]   Initial investigations on systems for measuring intraocular pressure [J].
Schnakenberg, U ;
Walter, P ;
Bögel, GV ;
Krüger, C ;
Lüdtke-Handjery, HC ;
Richter, HA ;
Specht, W ;
Ruokonen, P ;
Mokwa, W .
SENSORS AND ACTUATORS A-PHYSICAL, 2000, 85 (1-3) :287-291
[10]   Development of a completely encapsulated intraocular pressure sensor [J].
Walter, P ;
Schnakenberg, U ;
vom Bögel, G ;
Ruokonen, P ;
Krüger, C ;
Dinslage, S ;
Handjery, HCL ;
Richter, H ;
Mokwa, W ;
Diestelhorst, M ;
Krieglstein, GK .
OPHTHALMIC RESEARCH, 2000, 32 (06) :278-284