Effect of keratoconus grades on repeatability of keratometry readings: Comparison of 5 devices

被引:77
作者
Hashemi, Hassan [1 ]
Yekta, Abbasali [3 ]
Khabazkhoob, Mehdi [2 ]
机构
[1] Shahid Beheshti Univ Med Sci, Noor Eye Hosp, Noor Ophthalmol Res Ctr, Tehran, Iran
[2] Shahid Beheshti Univ Med Sci, Fac Publ Hlth, Dept Epidemiol, Tehran, Iran
[3] Mashhad Univ Med Sci, Sch Paramed Sci, Dept Optometry, Mashhad, Iran
关键词
OPTICAL COHERENCE TOMOGRAPHY; CORNEAL RING SEGMENTS; FOLLOW-UP; TOPOGRAPHY; PENTACAM; REPRODUCIBILITY; RELIABILITY; INSTRUMENTS; CURVATURE; SURFACES;
D O I
10.1016/j.jcrs.2014.08.043
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To determine the repeatability of keratometry (K) measurements of a Scheimpflug pachymeter (Pentacam), Placido topographer (Eyesys), scanning-slit corneal topographer (Orbscan), partial coherence interferometry (PCI) device (IOLMaster), and Javal manual keratometer with different grades of keratoconus. SETTING: Noor Eye Hospital, Tehran, Iran. DESIGN: Retrospective case series. METHODS: Keratometry was performed first with Scheimpflug pachymetry followed, in order, by Placido topography, scanning-slit corneal topography, PCI, and manual keratometry. Repeatability was examined in groups with a maximum K of less than 50.0 diopters (D), 50.0 to 55.0 D, and more than 55.0 D. RESULTS: Seventy-eight eyes of 45 keratoconus patients were assessed. In Group 1, repeatability was highest with Scheimpflug pachymetry and lowest with scanning-slit corneal topography (0.36 to 1.24). In Group 2, the intraclass correlation coefficients (ICCs) for maximum K ranged from 0.823 with the scanning-slit corneal topography to 0.974 with Scheimpflug pachymetry. The repeatability index for minimum K (0.53 to 2.11) and maximum K (0.60 to 1.92) in this group was highest with,Scheimpflug pachymetry and with for scanning-slit corneal topography. In Group 3, the ICCs for minimum K and maximum K ranged from 0.890 to 0.990, and the repeatability index for minimum K varied between 1.66 with Scheimpflug pachymetry to 2.98 with Placido topography; for maximum K, the index was from 2.15 with PCI to 2.81 with the manual keratometer. CONCLUSIONS: In mild keratoconus, the 5 devices had acceptable repeatability in K readings. In cases with a maximum K reading greater than 55.0 D, all devices had reduced repeatability as a result of measurement errors; thus, measurements might not be so reliable. (C) 2015 ASCRS and ESCRS
引用
收藏
页码:1065 / 1072
页数:8
相关论文
共 25 条
  • [21] Corneal power measurement with Fourier-domain optical coherence tomography
    Tang, Maolong
    Chen, Alex
    Li, Yan
    Huang, David
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (12) : 2115 - 2122
  • [22] COMPARISON OF 3 KERATOMETRY INSTRUMENTS
    TENNEN, DG
    KEATES, RH
    MONTOYA, C
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (04) : 407 - 408
  • [23] Long-term follow-up of intrastromal corneal ring segments in keratoconus
    Torquetti, Leonardo
    Berbel, Rodrigo Fabri
    Ferrara, Paulo
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (10) : 1768 - 1773
  • [24] Outcomes of intrastromal corneal ring segments for treatment of keratoconus: Five-year follow-up analysis
    Vega-Estrada, Alfredo
    Alio, Jorge L.
    Brenner, Luis F.
    Burguera, Neus
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (08) : 1234 - 1240
  • [25] Wilhelmus KR, 2004, J CATARACT REFR SURG, V30, P2425, DOI 10.1016/j.jcrs.2004.02.068