Four dimensional computed tomography (4DCT) scans are routinely used in radiation therapy to determine the internal treatment volume for targets that are moving (e.g. lung tumors). The use of these studies has allowed clinicians to create target volumes based upon the motion of the tumor during the imaging study. The purpose of this work is to determine if a target volume based on a single 4DCT scan at simulation is sufficient to capture thoracic motion. Phantom studies were performed to determine expected differences between volumes contoured on 4DCT scans and those on the evaluation CT scans (slow scans). Evaluation CT scans acquired during treatment of 11 patients were compared to the 4DCT scans used for treatment planning. The images were assessed to determine if the target remained within the target volume determined during the first 4DCT scan. A total of 55 slow scans were compared to the 11 planning 4DCT scans. Small differences were observed in phantom between the 4DCT volumes and the slow scan volumes, with a maximum of 2.9%, that can be attributed to minor differences in contouring and the ability of the 4DCT scan to adequately capture motion at the apex and base of the motion trajectory. Larger differences were observed in the patients studied, up to a maximum volume difference of 33.4%. These results demonstrate that a single 4DCT scan is not adequate to capture all thoracic motion throughout treatment.
引用
收藏
页数:7
相关论文
共 23 条
  • [11] Motion-aware temporal regularization for improved 4D cone-beam computed tomography
    Mory, Cyril
    Janssens, Guillaume
    Rit, Simon
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (18) : 6856 - 6877
  • [12] Treatment planning and 4D robust evaluation strategy for proton therapy of lung tumors with large motion amplitude
    Taasti, Vicki Trier
    Hattu, Djoya
    Vaassen, Femke
    Canters, Richard
    Velders, Marije
    Mannens, Jolein
    van Loon, Judith
    Rinaldi, Ilaria
    Unipan, Mirko
    van Elmpt, Wouter
    MEDICAL PHYSICS, 2021, 48 (08) : 4425 - 4437
  • [13] Utility of single photon emission computed tomography perfusion scans in radiation treatment planning of locally advanced lung cancers
    Agrawal, Sushma
    Raj, M. Karthick
    Kheruka, Subhash Chand
    Das, K. J. Maria
    Gambhir, Sanjay
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2012, 27 (01): : 10 - 15
  • [14] 4D-CT PULMONARY VENTILATION IMAGING IN FUNCTIONAL TREATMENT PLANNING FOR THORACIC CANCER RADIOTHERAPY
    Huang, Tzung-Chi
    Hsiao, Chien-Yi
    Chien, Chun-Ru
    Liang, Ji-An
    Shih, Tzu-Ching
    Chi, Kwan-Hwa
    Zhang, Geoffrey
    JOURNAL OF THORACIC ONCOLOGY, 2012, 7 (11) : S478 - S479
  • [15] IMRT treatment plans and functional planning with functional lung imaging from 4D-CT for thoracic cancer patients
    Huang, Tzung-Chi
    Hsiao, Chien-Yi
    Chien, Chun-Ru
    Liang, Ji-An
    Shih, Tzu-Ching
    Zhang, Geoffrey G.
    RADIATION ONCOLOGY, 2013, 8
  • [16] Physics-Based Modeling of Aortic Wall Motion from ECG-Gated 4D Computed Tomography
    Xiong, Guanglei
    Taylor, Charles A.
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2010, PT I, 2010, 6361 : 426 - +
  • [17] Differences in respiratory-induced pancreatic tumor motion between 4D treatment planning CT and daily cone beam CT, measured using intratumoral fiducials
    Lens, Eelco
    van der Horst, Astrid
    Kroon, Petra S.
    van Hooft, Jeanin E.
    Fajardo, Raquel Davila
    Fockens, Paul
    van Tienhoven, Geertjan
    Bel, Arjan
    ACTA ONCOLOGICA, 2014, 53 (09) : 1257 - 1264
  • [18] Noise reduction and motion elimination in low-dose 4D myocardial computed tomography perfusion (CTP): preliminary clinical evaluation of the ASTRA4D algorithm
    Lukas, Steffen
    Feger, Sarah
    Rief, Matthias
    Zimmermann, Eike
    Dewey, Marc
    EUROPEAN RADIOLOGY, 2019, 29 (09) : 4572 - 4582
  • [19] IMRT treatment plans and functional planning with functional lung imaging from 4D-CT for thoracic cancer patients
    Tzung-Chi Huang
    Chien-Yi Hsiao
    Chun-Ru Chien
    Ji-An Liang
    Tzu-Ching Shih
    Geoffrey G Zhang
    Radiation Oncology, 8
  • [20] Four-dimensional computed tomography scan analysis of tumor and organ motion at varying levels of abdominal compression during stereotactic treatment of lung and liver
    Heinzerling, John H.
    Anderson, John F.
    Papiez, Lech
    Boike, Thomas
    Chien, Stanley
    Zhang, Geoffrey
    Abdulrahman, Ramzi
    Timmerman, Robert
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (05): : 1571 - 1578