Mixed reality ultrasound guidance system: a case study in system development and a cautionary tale

被引:23
作者
Ameri, Golafsoun [1 ]
Baxter, John S. H. [1 ]
Bainbridge, Daniel [2 ]
Peters, Terry M. [1 ]
Chen, Elvis C. S. [1 ]
机构
[1] Western Univ, Robarts Res Inst, London, ON, Canada
[2] London Hlth Sci Ctr, London, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Ultrasound guidance; Mixed reality; Augmented reality; Augmented virtuality; System development; Guidelines; Internal jugular vein catheterization; Central line; INTERNAL JUGULAR-VEIN; CENTRAL VENOUS CATHETERIZATION; MITRAL-VALVE REPAIR; AUGMENTED REALITY; 3D ULTRASOUND; ULTRASONOGRAPHY; NEUROSURGERY; ACCESS; ECHOCARDIOGRAPHY; INTERVENTIONS;
D O I
10.1007/s11548-017-1665-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Real-time ultrasound has become a crucial aspect of several image-guided interventions. One of the main constraints of such an approach is the difficulty in interpretability of the limited field of view of the image, a problem that has recently been addressed using mixed reality, such as augmented reality and augmented virtuality. The growing popularity and maturity of mixed reality has led to a series of informal guidelines to direct development of new systems and to facilitate regulatory approval. However, the goals of mixed reality image guidance systems and the guidelines for their development have not been thoroughly discussed. The purpose of this paper is to identify and critically examine development guidelines in the context of a mixed reality ultrasound guidance system through a case study. A mixed reality ultrasound guidance system tailored to central line insertions was developed in close collaboration with an expert user. This system outperformed ultrasound-only guidance in a novice user study and has obtained clearance for clinical use in humans. A phantom study with 25 experienced physicians was carried out to compare the performance of the mixed reality ultrasound system against conventional ultrasound-only guidance. Despite the previous promising results, there was no statistically significant difference between the two systems. Guidelines for developing mixed reality image guidance systems cannot be applied indiscriminately. Each design decision, no matter how well justified, should be the subject of scientific and technical investigation. Iterative and small-scale evaluation can readily unearth issues and previously unknown or implicit system requirements. We recommend a wary eye in development of mixed reality ultrasound image guidance systems emphasizing small-scale iterative evaluation alongside system development. Ultimately, we recommend that the image-guided intervention community furthers and deepens this discussion into best practices in developing image-guided interventions.
引用
收藏
页码:495 / 505
页数:11
相关论文
共 48 条
[1]   Training for Planning Tumour Resection: Augmented Reality and Human Factors [J].
Abhari, Kamyar ;
Baxter, John S. H. ;
Chen, Elvis C. S. ;
Khan, Ali R. ;
Peters, Terry M. ;
de Ribaupierre, Sandrine ;
Eagleson, Roy .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (06) :1466-1477
[2]   Real-time ultrasound guided placement of temporary internal jugular vein catheters: Assessment of technical success and complication rates in nephrology practice [J].
Akoglu, Hadim ;
Piskinpasa, Serhan ;
Yenigun, Ezgi C. ;
Ozturk, Ramazan ;
Dede, Fatih ;
Odabas, Ali R. .
NEPHROLOGY, 2012, 17 (07) :603-606
[3]   Augmented Reality Ultrasound Guidance for Central Line Procedures: Preliminary Results [J].
Ameri, Golafsoun ;
Baxter, John S. H. ;
McLeod, A. Jonathan ;
Peters, Terry M. ;
Chen, Elvis C. S. .
AUGMENTED ENVIRONMENTS FOR COMPUTER-ASSISTED INTERVENTIONS, AE-CAI 2015, 2015, 9365 :11-20
[4]  
[Anonymous], 2010, INT J VIRTUAL REALIT
[5]   Anatomic considerations for central venous cannulation [J].
Bannon, Michael P. ;
Heller, Stephanie F. ;
Rivera, Mariela .
RISK MANAGEMENT AND HEALTHCARE POLICY, 2011, 4 :27-39
[6]  
Baumann M, 2009, LECT NOTES COMPUT SC, V5761, P67, DOI 10.1007/978-3-642-04268-3_9
[7]   Mechanically assisted 3D ultrasound guided prostate biopsy system [J].
Bax, Jeffrey ;
Cool, Derek ;
Gardi, Lori ;
Knight, Kerry ;
Smith, David ;
Montreuil, Jacques ;
Sherebrin, Shi ;
Romagnoli, Cesare ;
Fenster, Aaron .
MEDICAL PHYSICS, 2008, 35 (12) :5397-5410
[8]  
Bichlmeier C., 2007, INT S MIX AUGM REAL, P129, DOI DOI 10.1109/ISMAR.2007.4538837
[9]   An unseen danger: Frequency of posterior vessel wall penetration by needles during attempts to place internal jugular vein central catheters using ultrasound guidance [J].
Blaivas, Michael ;
Adhikari, Srikar .
CRITICAL CARE MEDICINE, 2009, 37 (08) :2345-2349