Estimates of the Image Quality and the Radiation Dose for Head and Abdomen Phantom Image Acquisition by Using Dual-energy CT

被引:0
作者
Kim, Dae-Hong [1 ,2 ]
Kim, Hee-Joung [1 ,2 ]
Lee, Chang-Lae [1 ,2 ]
Jeon, Pil-Hyun [1 ,2 ,3 ]
Lee, Won-Hyung [3 ]
Jeon, Sung-Soo [3 ]
机构
[1] Yonsei Univ, Coll Hlth Sci, Dept Radiol Sci, Wonju 220710, South Korea
[2] Yonsei Univ, Coll Hlth Sci, Inst Hlth Sci, Wonju 220710, South Korea
[3] Yonsei Univ, Wonju Christian Hosp, Dept Radiol, Wonju 220710, South Korea
基金
新加坡国家研究基金会;
关键词
Linearly-mixed image; CTDIw; Contrast-to-noise ratios (CNR); Dual-energy CT; CARLO-BASED METHOD; TISSUE SEGMENTATION; MULTIDETECTOR CT;
D O I
10.3938/jkps.60.1150
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Using dual-energy computed tomography (CT) scans, we obtained images from a raw data set by using low-and high-energy scans (usually 80 and 140 kV, respectively). Generally, the head and the abdomen examinations were performed using single-energy (120 kV) scanning with a routine exposure protocol, and the images were used for diagnostic interpretation. The dual-energy CT scans can also be used for diagnosis by using the linearly-mixed method with low-and high-energy images. In the current study, we evaluated mixed images in dual-energy and single-energy scans for image quality and radiation dose. The CT scan protocol for single energy was adopted from IEC protocols, and the recommended dose from the EC. The dual-energy scan protocol was based on the Siemens dual-energy CT scan protocol. The CT scan protocols were extended further in the mA range for both scans to estimate the image quality corresponding to these dose alternations. The results demonstrate that the Contrast-to-noise ratio (CNR) value of a dual-energy mixed image is higher than that of a single-energy image in similar dose regions for both the head and the abdomen phantoms. We observed that the dual-energy images could reduce the dose compared to single-energy images. In the abdomen phantom study, the CNR of dual-energy images was even higher than that of single-energy images with half the radiation dose of the single-energy scan. Therefore, the dual-energy CT scan can accomplish a remarkable dose reduction while preserving image quality for head and abdomen imaging.
引用
收藏
页码:1150 / 1156
页数:7
相关论文
共 14 条
[1]   Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculations [J].
Bazalova, Magdalena ;
Carrier, Jean-Francois ;
Beaulieu, Luc ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2439-2456
[2]   Tissue segmentation in Monte Carlo treatment planning: A simulation study using dual-energy CT images [J].
Bazatova, Magdatena ;
Carrier, Jean-Francois ;
Beautieu, Luc ;
Verhaegen, Frank .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :93-98
[3]   Calcified vascular plaque specimens: Assessment with cardiac dual-energy multidetector CT in anthropomorphically moving heart phantom [J].
Boll, Daniel T. ;
Merkle, Elmar M. ;
Paulson, Erik K. ;
Mirza, Rizvan A. ;
Fleiter, Thorsten R. .
RADIOLOGY, 2008, 249 (01) :119-126
[4]   Estimating Effective Dose for CT Using Dose-Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy Scanning [J].
Christner, Jodie A. ;
Kofler, James M. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (04) :881-889
[5]   A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms [J].
DeMarco, JJ ;
Cagnon, CH ;
Cody, DD ;
Stevens, DM ;
McCollough, CH ;
O'Daniel, J ;
McNitt-Gray, MF .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :3989-4004
[6]  
European Commission, 1999, 16262 EUR EN EUR COM
[7]   Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner [J].
Goodsitt, Mitchell M. ;
Christodoulou, Emmanuel G. ;
Larson, Sandra C. .
MEDICAL PHYSICS, 2011, 38 (04) :2222-2232
[8]   A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models [J].
Jarry, G ;
DeMarco, JJ ;
Beifuss, U ;
Cagnon, CH ;
McNitt-Gray, MF .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (16) :2645-2663
[9]   Radiation dose and image quality in pediatric CT: Effect of technical factors and phantom size and shape [J].
Siegel, MJ ;
Schmidt, B ;
Bradley, D ;
Suess, C ;
Hildebolt, C .
RADIOLOGY, 2004, 233 (02) :515-522
[10]   Electron density measurement with dual-energy x-ray CT using synchrotron radiation [J].
Torikoshi, M ;
Tsunoo, T ;
Sasaki, M ;
Endo, M ;
Noda, Y ;
Ohno, Y ;
Kohno, T ;
Hyodo, K ;
Uesugi, K ;
Yagi, N .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (05) :673-685