Personalized Superficial HDR Brachytherapy-Dosimetric Verification of Dose Distribution with Lead Shielding of Critical Organs in the Head and Neck Region

被引:7
作者
Zwierzchowski, Grzegorz [1 ,2 ]
Bieleda, Grzegorz [1 ,2 ]
Szymbor, Agata [3 ]
Boehlke, Marek [4 ]
机构
[1] Poznan Univ Med Sci, Electroradiol Dept, PL-61866 Poznan, Poland
[2] Greater Poland Canc Ctr, Med Phys Dept, PL-61866 Poznan, Poland
[3] Adam Mickiewicz Univ, Fac Phys, PL-61614 Poznan, Poland
[4] West Pomeranian Oncol Ctr, Med Phys Dept, PL-71730 Szczecin, Poland
来源
JOURNAL OF PERSONALIZED MEDICINE | 2022年 / 12卷 / 09期
关键词
brachytherapy; dosimetry; dose calculations; quality assurance; 3D printing; RECOMMENDATIONS; PERTURBATION; KILOVOLTAGE; ALGORITHM; 3D;
D O I
10.3390/jpm12091432
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Surface brachytherapy, usually characterized by a high dose gradient, allows the dose to be precisely deposited in the irradiated area while protecting critical organs. When the lesion is located in the nasal or ocular region, the organ of vision must be protected. The aim of this study was to verify the dose distributions near critical organs in the head and neck region during a brachytherapy procedure using lead shielding of the eye. Methods: Anthropomorphic phantom using 3D-printing technology was prepared. The doses deposited at a point in the lens of the eye and on the surface of the eyelid, directly under the lead shield were calculated and measured using EBT3 radiochromic films. Comparison of doses planned in the treatment planning system using the TG-43 formalism, TG-186 formalism, and measured were also performed. Results: Comparing the planned and calculated doses with TG186 formalism it can be assumed that the use of lead shields is a method for protecting the organ of vision from the adverse effects of ionizing radiation. Conclusions: The decision to use a lead shield during facial surface brachytherapy procedures should be made on a patient-by-patient basis and based on model-based calculation methods recommended by TG186.
引用
收藏
页数:25
相关论文
共 24 条
[1]   Collapsed cone dose calculations for heterogeneous tissues in brachytherapy using primary and scatter separation source data [J].
Ahnesjo, Anders ;
van Veelen, Bob ;
Tedgren, Asa Carlsson .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2017, 139 :17-29
[2]   Radiation-induced lens opacities: Epidemiological, clinical and experimental evidence, methodological issues, research gaps and strategy [J].
Ainsbury, Elizabeth A. ;
Dalke, Claudia ;
Hamada, Nobuyuki ;
Benadjaoud, Mohamed Amine ;
Chumak, Vadim ;
Ginjaume, Merce ;
Kok, Judith L. ;
Mancuso, Mariateresa ;
Sabatier, Laure ;
Struelens, Lara ;
Thariat, Juliette ;
Jourdain, Jean-Rene .
ENVIRONMENT INTERNATIONAL, 2021, 146
[3]   Gafchromic film dosimetry of a new HDR 192Ir brachytherapy source [J].
Ayoobian, Navid ;
Asl, Akbar Sarabi ;
Poorbaygi, Hosein ;
Javanshir, Mohammad Reza .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02) :194-205
[4]   Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation [J].
Beaulieu, Luc ;
Tedgren, Asa Carlsson ;
Carrier, Jean-Francois ;
Davis, Stephen D. ;
Mourtada, Firas ;
Rivard, Mark J. ;
Thomson, Rowan M. ;
Verhaegen, Frank ;
Wareing, Todd A. ;
Williamson, Jeffrey F. .
MEDICAL PHYSICS, 2012, 39 (10) :6208-6236
[5]   Improving the energy response of external beam therapy (EBT) GafChromic™ dosimetry films at low energies (≤100 keV) [J].
Bekerat, H. ;
Devic, S. ;
DeBlois, F. ;
Singh, K. ;
Sarfehnia, A. ;
Seuntjens, J. ;
Shih, Shelley ;
Yu, Xiang ;
Lewis, D. .
MEDICAL PHYSICS, 2014, 41 (02)
[6]  
Biele da Grzegorz, 2021, J Contemp Brachytherapy, V13, P549, DOI 10.5114/jcb.2021.110304
[7]   3D printing of individual skin brachytherapy applicator: design, manufacturing, and early clinical results [J].
Biele, Grzegorz ;
Chichel, Adam ;
Boehlke, Marek ;
Zwierzchowski, Grzegorz ;
Chyrek, Artur ;
Burchardt, Wojciech ;
Stefaniak, Patrycja ;
Wis'niewska, Natalia ;
Czereba, Kinga ;
Malicki, Julian .
JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (02) :205-214
[8]   Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy [J].
Bieleda, Grzegorz ;
Zwierzchowski, Grzegorz ;
Roslan, Katarzyna ;
Adamus, Agnieszka ;
Malicki, Julian .
JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2021, 13 (02) :188-194
[9]   Design and characterization of a new high-dose-rate brachytherapy Valencia applicator for larger skin lesions [J].
Candela-Juan, C. ;
Niatsetski, Y. ;
van der Laarse, R. ;
Granero, D. ;
Ballester, F. ;
Perez-Calatayud, J. ;
Vijande, J. .
MEDICAL PHYSICS, 2016, 43 (04) :1639-1648
[10]   Dosimetric perturbations of a lead shield for surface and interstitial high-dose-rate brachytherapy [J].
Candela-Juan, Cristian ;
Granero, Domingo ;
Vijande, Javier ;
Ballester, Facundo ;
Perez-Calatayud, Jose ;
Rivard, Mark J. .
JOURNAL OF RADIOLOGICAL PROTECTION, 2014, 34 (02) :297-311