MR spectroscopy using static higher order shimming with dynamic linear terms (HOS-DLT) for improved water suppression, interleaved MRS-fMRI, and navigator-based motion correction at 7T

被引:13
作者
Boer, Vincent O. [1 ]
Andersen, Mads [2 ]
Lind, Anna [1 ]
Lee, Nam Gyun [1 ,3 ]
Marsman, Anouk [1 ]
Petersen, Esben T. [1 ,4 ]
机构
[1] Copenhagen Univ Hosp Hvidovre, Danish Res Ctr Magnet Resonance, Ctr Funct & Diagnost Imaging & Res, Hvidovre, Denmark
[2] Philips Healthcare, Copenhagen, Denmark
[3] Univ Southern Calif, Dept Biomed Engn, Los Angeles, CA 90007 USA
[4] Tech Univ Denmark, Ctr Magnet Resonance, Dept Hlth Technol, Lyngby, Denmark
关键词
H-1 MR spectroscopy; B-0; shimming; EPI distortions; prospective motion correction; ultra-high field; water suppression; REAL-TIME MOTION; VIVO H-1-NMR SPECTROSCOPY; HUMAN BRAIN; SHORT TE; REGION; B0;
D O I
10.1002/mrm.28202
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To interleave global and local higher order shimming for single voxel MRS. Single voxel MR spectroscopy requires optimization of the B-0 field homogeneity in the region of the voxel to obtain a narrow linewidth and provide high data quality. However, the optimization of local higher order fields on a localized MRS voxel typically leads to large field offsets outside that volume. This compromises interleaved MR sequence elements that benefit from global field homogeneity such as water suppression, interleaved MRS-fMRI, and MR motion correction. Methods: A shimming algorithm was developed to optimize the MRS voxel homogeneity and the whole brain homogeneity for interleaved sequence elements, using static higher order shims and dynamic linear terms (HOS-DLT). Shimming performance was evaluated using 6 brain regions and 10 subjects. Furthermore, the benefits of HOS-DLT was demonstrated for water suppression, MRS-fMRI, and motion corrected MRS using fat-navigators. Results: The HOS-DLT algorithm was shown to improve the whole brain homogeneity compared to an MRS voxel-based shim, without compromising the MRS voxel homogeneity. Improved water suppression over the brain, reduced image distortions in MRS-fMRI, and improved quality of motion navigators were demonstrated using the HOS-DLT method. Conclusion: HOS-DLT shimming allowed for both local and global field homogeneity, providing excellent MR spectroscopy data quality, as well as good field homogeneity for interleaved sequence elements, even without the need for dynamic higher order shimming capabilities.
引用
收藏
页码:1101 / 1112
页数:12
相关论文
共 41 条
[1]  
Andersen M, 2019, ISMRM
[2]   Improvement in diagnostic quality of structural and angiographic MRI of the brain using motion correction with interleaved, volumetric navigators [J].
Andersen, Mads ;
Bjorkman-Burtscher, Isabella M. ;
Marsman, Anouk ;
Petersen, Esben Thade ;
Boer, Vincent Oltman .
PLOS ONE, 2019, 14 (05)
[3]   How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging [J].
Andersson, JLR ;
Skare, S ;
Ashburner, J .
NEUROIMAGE, 2003, 20 (02) :870-888
[4]   Image-Based Method to Measure and Characterize Shim-Induced Eddy Current Fields [J].
Bhogal, Alex ;
Versluis, Maarten ;
Koonen, Jos ;
Siero, Jeroen C. W. ;
Boer, Vincent O. ;
Klomp, Dennis ;
Luijten, Peter R. ;
Hoogduin, Hans .
CONCEPTS IN MAGNETIC RESONANCE PART A, 2013, 42 (06) :245-260
[5]   7-T 1H MRS with adiabatic refocusing at short TE using radiofrequency focusing with a dual-channel volume transmit coil [J].
Boer, V. O. ;
van Lier, A. L. H. M. W. ;
Hoogduin, J. M. ;
Wijnen, J. P. ;
Luijten, P. R. ;
Klomp, D. W. J. .
NMR IN BIOMEDICINE, 2011, 24 (09) :1038-1046
[6]   High-field MRS of the human brain at short TE and TR [J].
Boer, Vincent O. ;
Siero, Jeroen C. W. ;
Hoogduin, Hans ;
van Gorp, Jetse S. ;
Luijten, Peter R. ;
Klomp, Dennis W. J. .
NMR IN BIOMEDICINE, 2011, 24 (09) :1081-1088
[7]   Multislice 1H MRSI of the human brain at 7 T using dynamic B0 and B1 shimming [J].
Boer, Vincent O. ;
Klomp, Dennis W. J. ;
Juchem, Christoph ;
Luijten, Peter R. ;
de Graaf, Robin A. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (03) :662-670
[8]   Real-time motion- and Bo-correction for LASER-localized spiral-accelerated 3D-MRSI of the brain at 3 T [J].
Bogner, Wolfgang ;
Hess, Aaron T. ;
Gagoski, Borjan ;
Tisdall, M. Dylan ;
van der Kouwe, Andre J. W. ;
Trattnig, Siegfried ;
Rosen, Bruce ;
Andronesi, Ovidiu C. .
NEUROIMAGE, 2014, 88 :22-31
[9]   Degraded water suppression in small volume 1H MRS due to localised shimming [J].
Carlsson, Asa ;
Ljungberg, Maria ;
Starck, Goran ;
Forssell-Aronsson, Eva .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2011, 24 (02) :97-107
[10]   An interior trust region approach for nonlinear minimization subject to bounds [J].
Coleman, TF ;
Li, YY .
SIAM JOURNAL ON OPTIMIZATION, 1996, 6 (02) :418-445