Effective dose in paediatric computed tomography

被引:64
作者
Chapple, CL [1 ]
Willis, S [1 ]
Frame, J [1 ]
机构
[1] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
关键词
D O I
10.1088/0031-9155/47/1/308
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is limited data currently available for making dose and risk assessments for paediatric patients undergoing computed tomographic examination. A method has been developed to correlate the risk-related quantity, effective dose, to the more simply derived quantity dose-length product. This involved scanning a series of paediatric anthropomorphic phantoms containing thermoluminescent dosimeters to measure effective dose for scans of various anatomic regions. The quantity effective dose per dose-length product was calculated and plotted as a function of patient size. This showed a simple exponential relationship. and equations of fit were derived to enable the calculation of effective dose for a patient of any size. Measurements carried out on a second scanner and for alternative scan volumes indicated that the method could be generally utilized.
引用
收藏
页码:107 / 115
页数:9
相关论文
共 19 条
[1]  
Almen A., 1996, Journal of Radiological Protection, V16, P81, DOI 10.1088/0952-4746/16/2/004
[2]   COMPUTED TOMOGRAPHIC SCANNING IN CHILDREN .2. AN UPDATED COMPARISON OF RADIATION-DOSE AND RESOLVING POWER OF COMMERCIAL SCANNERS [J].
BRASCH, RC ;
CANN, CE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 138 (01) :127-133
[3]  
Broadhead D. A., 1997, Journal of Radiological Protection, V17, P17, DOI 10.1088/0952-4746/17/1/003
[4]   ACTIVE BONE-MARROW DISTRIBUTION AS A FUNCTION OF AGE IN HUMANS [J].
CRISTY, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :389-400
[5]  
CRISTY M, 1980, NUREGICR1159
[6]  
European guidelines on quality criteria for computed tomography, 1999, 16262 EUR
[7]   PEDIATRIC-PATIENT EXPOSURES FROM CT EXAMINATIONS - GE CT/T 9800 SCANNER [J].
FEARON, T ;
VUCICH, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1985, 144 (04) :805-809
[8]   NORMALIZED PEDIATRIC ORGAN-ABSORBED DOSES FROM CT EXAMINATIONS [J].
FEARON, T ;
VUCICH, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1987, 148 (01) :171-174
[9]   Effective doses to patients undergoing thoracic computed tomography examinations [J].
Huda, W ;
Scalzetti, EM ;
Roskopf, M .
MEDICAL PHYSICS, 2000, 27 (05) :838-844
[10]  
HUDA W, 1997, RADIOLOGY, V203, P4117