A phantom for diffusion-weighted imaging of acute stroke

被引:61
作者
Laubach, HJ [1 ]
Jakob, PM [1 ]
Loevblad, KO [1 ]
Baird, AE [1 ]
Bovo, MP [1 ]
Edelman, RR [1 ]
Warach, S [1 ]
机构
[1] Beth Israel Deaconess Med Ctr, Dept Neurol, Boston, MA 02215 USA
关键词
diffusion magnetic resonance imaging; stroke; phantom volumetry;
D O I
10.1002/jmri.1880080627
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A tissue phantom for diffusion-weighted imaging was developed, basing its contrast between two compartments on different apparent diffusion coefficients, without contrast due to T2 relaxation and proton density. These contrast properties of the phantom simulate the situation found in normal gray matter and areas of acute ischemia, A possible application of the phantom was demonstrated for the investigation of the accuracy of volume measurements based on diffusion-weighted images.
引用
收藏
页码:1349 / 1354
页数:6
相关论文
共 25 条
[1]   TISSUE-MIMICKING GELATIN AGAR GELS FOR USE IN MAGNETIC-RESONANCE IMAGING PHANTOMS [J].
BLECHINGER, JC ;
MADSEN, EL ;
FRANK, GR .
MEDICAL PHYSICS, 1988, 15 (04) :629-636
[2]   T1 AND T2 MEASUREMENTS ON A 1.5-T COMMERCIAL MR IMAGER [J].
BREGER, RK ;
RIMM, AA ;
FISCHER, ME ;
PAPKE, RA ;
HAUGHTON, VM .
RADIOLOGY, 1989, 171 (01) :273-276
[3]  
BROSIO E, 1994, CELL MOL BIOL, V40, P569
[4]  
CHIEN D, 1992, AM J NEURORADIOL, V13, P1097
[5]   NICKEL-DOPED AGAROSE-GEL PHANTOMS IN MR IMAGING [J].
CHRISTOFFERSSON, JO ;
OLSSON, LE ;
SJOBERG, S .
ACTA RADIOLOGICA, 1991, 32 (05) :426-431
[6]   AN MRI TISSUE EQUIVALENT LESION PHANTOM USING A NOVEL POLYSACCHARIDE MATERIAL [J].
GROCH, MW ;
URBON, JA ;
ERWIN, WD ;
ALDOOHAN, S .
MAGNETIC RESONANCE IMAGING, 1991, 9 (03) :417-421
[7]   PROTON NMR RELAXATION STUDIES OF AQUEOUS POLYSACCHARIDE SYSTEMS [J].
HILLS, BP ;
CANO, C ;
BELTON, PS .
MACROMOLECULES, 1991, 24 (10) :2944-2950
[8]  
JAKOB PM, 1997, ADV MRI, V2, P177
[9]   MR IMAGING OF INTRAVOXEL INCOHERENT MOTIONS - APPLICATION TO DIFFUSION AND PERFUSION IN NEUROLOGIC DISORDERS [J].
LEBIHAN, D ;
BRETON, E ;
LALLEMAND, D ;
GRENIER, P ;
CABANIS, E ;
LAVALJEANTET, M .
RADIOLOGY, 1986, 161 (02) :401-407
[10]   Single-shot diffusion MRI of human brain on a conventional clinical instrument [J].
Liu, GY ;
vanGelderen, P ;
Duyn, J ;
Moonen, CTW .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :671-677