Device Cleaning and Infection Control in Aerosol Therapy

被引:57
作者
O'Malley, Catherine A. [1 ]
机构
[1] Ann & Robert H Lurie Childrens Hosp Chicago, Chicago, IL 60611 USA
关键词
aerosol drug therapy; medical devices; nebulizers; inhalers; valved holding chambers; inhalation spacers; equipment contamination; infection prevention; infection control; respiratory therapy; CYSTIC-FIBROSIS; BURKHOLDERIA-CEPACIA; BACTERIAL-CONTAMINATION; NEBULIZERS; OUTBREAK; DELIVERY; RECOMMENDATIONS; MEDICATIONS; PREVENTION; GUIDELINE;
D O I
10.4187/respcare.03513
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aerosol delivery equipment used to administer inhaled medications includes the nebulizer, positive expiratory pressure devices added to the nebulizer, and valved holding chambers (spacers). These devices are semi-critical medical devices, and as such, infection prevention and control (IPC) guidelines recommend that they be cleaned, disinfected, rinsed with sterile water, and air-dried. There is confusion surrounding the care of aerosol devices because of inconsistencies in the various published IPC guidelines, lack of a standard of practice among institutions and respiratory therapists (RTs), and manufacturer's instructions for use of these devices are not always compatible with guidelines or practice. Challenges lie in awareness of IPC guidelines and establishing a standard for the care of aerosol delivery devices among all stakeholders/manufacturers, governments, vendors, and users. The latest IPC guideline from the Cystic Fibrosis Foundation, reviewed and endorsed by the Society for Healthcare Epidemiology of America and the Association for Professionals in Infection Control, has a recommendation for disposable nebulizers and a recommendation for reusable nebulizers. Reusable nebulizers should be cleaned, disinfected, rinsed with sterile water (if using a cold disinfectant), and air-dried between uses. The mouthpiece/mask of disposable nebulizers should be wiped with an alcohol pad, the residual volume should be rinsed out with sterile water after use, and the nebulizer should be replaced every 24 h. The RT plays a significant and responsible role in providing and teaching aerosol therapy to patients. The RT and all stakeholders need to work together to provide a standard of care for the safe use of aerosol delivery devices.
引用
收藏
页码:917 / 927
页数:11
相关论文
共 50 条
  • [31] A schlieren optical study of the human cough with and without wearing masks for aerosol infection control
    Tang, Julian W.
    Liebner, Thomas J.
    Craven, Brent A.
    Settles, Gary S.
    JOURNAL OF THE ROYAL SOCIETY INTERFACE, 2009, 6 : S727 - S736
  • [32] Extracorporeal membrane oxygenation (ECMO): an infection control perspective
    Delpiano, Luis
    Herve, Beatrice
    Jemenao, Maria, I
    Jofre, Leonor
    Medel, Myriam
    Tinoco, Javier
    Diaz, Rodrigo
    REVISTA CHILENA DE INFECTOLOGIA, 2021, 38 (05): : 622 - 633
  • [33] Improving implementation of infection control guidelines to reduce nosocomial infection rates: pioneering the report card
    Yinnon, A. M.
    Wiener-Well, Y.
    Jerassy, Z.
    Dor, M.
    Freund, R.
    Mazouz, B.
    Lupyan, T.
    Shapira, S.
    Attias, D.
    Assous, M. V.
    Kopuit, P.
    Block, C.
    Raveh, D.
    Freier-Dror, Y.
    Moses, A. E.
    Benenson, S.
    JOURNAL OF HOSPITAL INFECTION, 2012, 81 (03) : 169 - 176
  • [34] Profile of anesthetic infection control in Taiwan: a questionnaire report
    Or, Richard C. H.
    Hsieh, Teng K.
    Lan, Kuo M.
    Kang, Fu C.
    Chen, Ying H.
    So, Edmund C.
    JOURNAL OF CLINICAL ANESTHESIA, 2009, 21 (01) : 13 - 18
  • [35] Safety and Effectiveness of Medical Device Therapy
    Zusterzeel, Robbert
    SEX-SPECIFIC ANALYSIS OF CARDIOVASCULAR FUNCTION, 2018, 1065 : 107 - 121
  • [36] Evaluation of Phage Therapy for Pulmonary Infection of Mouse by Liquid Aerosol-Exposure Pseudomonas aeruginosa
    Zhang, Yajun
    Meng, Biao
    Wei, Xiao
    Li, Yan
    Wang, Xiaohui
    Zheng, Yan
    Wang, Changjun
    Cui, Lihong
    Zhao, Xiangna
    INFECTION AND DRUG RESISTANCE, 2021, 14 : 4457 - 4469
  • [37] Effects of flow pattern, device and formulation on particle size distribution of nebulized aerosol
    Hu, Junhua
    Zhang, Rui
    Beng, Huimin
    Deng, Liangjun
    Ke, Qingjin
    Tan, Wen
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2019, 560 : 35 - 46
  • [38] Infection Control in Anesthesia
    Bonsignore, Marzia
    Nachtigall, Irit
    ANASTHESIOLOGIE INTENSIVMEDIZIN NOTFALLMEDIZIN SCHMERZTHERAPIE, 2021, 56 (07/08): : 474 - 484
  • [39] Informatics in Infection Control
    Lin, Michael Y.
    Trick, William E.
    INFECTIOUS DISEASE CLINICS OF NORTH AMERICA, 2016, 30 (03) : 759 - +
  • [40] Infection Control in the Era of COVID-19: A Narrative Review
    Shbaklo, Nour
    Lupia, Tommaso
    De Rosa, Francesco G.
    Corcione, Silvia
    ANTIBIOTICS-BASEL, 2021, 10 (10):