Tensile testing as a novel method for quantitatively evaluating bioabsorbable material degradation

被引:31
作者
Bowen, Patrick K. [1 ]
Gelbaugh, Jesse A. [1 ]
Mercier, Phillip J. [2 ]
Goldman, Jeremy [2 ]
Drelich, Jaroslaw [1 ]
机构
[1] Michigan Technol Univ, Dept Mat Sci & Engn, Houghton, MI 49931 USA
[2] Michigan Technol Univ, Dept Biomed Engn, Houghton, MI 49931 USA
关键词
bioabsorbable stent; iron; magnesium; corrosion; tensile testing; BIOMEDICAL MAGNESIUM ALLOY; IN-VITRO CORROSION; BEHAVIOR; BIOCOMPATIBILITY; STENTS;
D O I
10.1002/jbm.b.32775
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Bioabsorbable metallic materials have become a topic of interest in the field of interventional cardiology because of their potential application in stents. A well-defined, quantitative method for evaluating the degradation rate of candidate materials is currently needed in this area. In this study, biodegradation of 0.25-mm iron and magnesium wires was simulated in vitro through immersion in cell-culture medium with and without a fibrin coating (meant to imitate the neointima). The immersed samples were corroded under physiological conditions (37 degrees C, 5% CO2). Iron degraded in a highly localized manner, producing voluminous corrosion product while magnesium degraded more uniformly. To estimate the degradation rate in a quantitative manner, both raw and corroded samples underwent tensile testing using a protocol similar to that used on polymeric nanofibers. The effective ultimate tensile stress (tensile stress holding constant cross-sectional area) was determined to be the mechanical metric that exhibited the smallest amount of variability. When the effective tensile stress data were aggregated, a statistically significant downward, linear trend in strength was observed in both materials (Fe and Mg) with and without the fibrin coating. It was also demonstrated that tensile testing is able to distinguish between the higher degradation rate of the bare wires and the lower degradation rate of the fibrin-coated wires with confidence.(c) 2012 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2012.
引用
收藏
页码:2101 / 2113
页数:13
相关论文
共 30 条
  • [1] Bates R.C., 1984, FRACTURE INTERACTION
  • [2] The Tensile Properties of Electrospun Nylon 6 Single Nanofibers
    Bazbouz, Mohamed Basel
    Stylios, George K.
    [J]. JOURNAL OF POLYMER SCIENCE PART B-POLYMER PHYSICS, 2010, 48 (15) : 1719 - 1731
  • [3] Bertha RM, 1995, STAT METHODS ENG SCI
  • [4] Chawla K.K., 2003, Ceramic Matrix Composites, DOI [10.1007/978-1-4615-1029-15, DOI 10.1007/978-1-4615-1029-15]
  • [5] In vitro corrosion and biocompatibility of binary magnesium alloys
    Gu, Xuenan
    Zheng, Yufeng
    Cheng, Yan
    Zhong, Shengping
    Xi, Tingfei
    [J]. BIOMATERIALS, 2009, 30 (04) : 484 - 498
  • [6] Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology?
    Heublein, B
    Rohde, R
    Kaese, V
    Niemeyer, M
    Hartung, W
    Haverich, A
    [J]. HEART, 2003, 89 (06) : 651 - 656
  • [7] Kainer K.U., 2003, MAGNESIUM ALLOYS TEC
  • [8] MAKAR GL, 1993, INT MATER REV, V38, P138, DOI 10.1179/095066093790326320
  • [9] McCrum NG., 1997, Principles of Polymer engineering, V2nd
  • [10] Biodegradable Metals for Cardiovascular Stent Application: Interests and New Opportunities
    Moravej, Maryam
    Mantovani, Diego
    [J]. INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2011, 12 (07) : 4250 - 4270