Bone drilling; temperature; elevation; FEA; osteonecrosis; drill bit; IMPLANT SITE PREPARATION; HEAT-GENERATION; TEMPERATURE RISE; CORTICAL BONE; SINGLE; PARAMETERS; EFFICIENCY; NECROSIS; FAILURE;
D O I
10.1177/0954411918819113
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Bone-drilling operation necessitates an accurate and efficient surgical drill bit to minimize thermal damage to the bone. This article provides a methodology for predicting the bone temperature elevation during surgical bone drilling and to gain a better understanding on the influences of the point angle, helix angle and web thickness of the drill bit. The proposed approach utilized the normalized Cockroft-Latham damage criterion to predict material cracking in the drilling process. Drilling simulation software DEFORM-3D is used to approximate the bone temperature elevation corresponding to different drill bit geometries. To validate the simulation results, bone temperature elevations were evaluated by comparison with ex vivo bone-drilling process using bovine femurs. The computational results fit well with the ex vivo experiments with respect to different drill geometries. All the investigated drill bit geometries significantly affect bone temperature rise. It is discovered that the thermal osteonecrosis risk regions could be reduced with a point angle of 110 degrees to 140 degrees, a helix angle of 5 degrees to 30 degrees and a web thickness of 5% to 40%. The drilling simulation could accurately estimate the maximum bone temperature elevation for various surgical drill bit point angles, web thickness and helix angles. Looking into the future, this work will lead to the research and redesign of the optimum surgical drill bit to minimize thermal insult during bone-drilling surgeries.
引用
收藏
页码:207 / 218
页数:12
相关论文
共 53 条
  • [11] Bertollo N., 2011, BIOMECHANICS APPL, DOI [DOI 10.5772/1424, 10.5772/1424]
  • [12] Effects of a drill diameter on the temperature rise in a bone during implant site preparation under clinical conditions
    Bogovic, Valerija
    Svete, Andrej
    Bajsic, Ivan
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (10) : 907 - 917
  • [13] BONFIELD W, 1968, Journal of Biomechanics, V1, P323, DOI 10.1016/0021-9290(68)90026-2
  • [14] Assessment of thermal necrosis risk regions for different bone qualities as a function of drilling parameters
    Chen, Yung-Chuan
    Tu, Yuan-Kun
    Tsai, Yi-Jung
    Tsai, Yi-Shan
    Yen, Cheng-Yo
    Yang, Shih-Chieh
    Hsiao, Chih-Kun
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2018, 162 : 253 - 261
  • [15] The effects of cutting parameters on cutting forces and heat generation when drilling animal bone and biomechanical test materials
    Cseke, Akos
    Heinemann, Robert
    [J]. MEDICAL ENGINEERING & PHYSICS, 2018, 51 : 24 - 30
  • [16] Drilling in bone: Modeling heat generation and temperature distribution
    Davidson, SRH
    James, DF
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2003, 125 (03): : 305 - 314
  • [17] Slow drilling speeds for single-drill implant bed preparation. Experimental in vitro study
    Delgado-Ruiz, R. A.
    Velasco Ortega, E.
    Romanos, G. E.
    Gerhke, S.
    Newen, I.
    Calvo-Guirado, J. L.
    [J]. CLINICAL ORAL INVESTIGATIONS, 2018, 22 (01) : 349 - 359
  • [18] TEMPERATURE THRESHOLD LEVELS FOR HEAT-INDUCED BONE TISSUE-INJURY - A VITAL-MICROSCOPIC STUDY IN THE RABBIT
    ERIKSSON, AR
    ALBREKTSSON, T
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (01) : 101 - 107
  • [19] Farnworth G. H., 1975, P 15 INT MACH TOOL D, P227
  • [20] Experimental determination of the emissivity of bone
    Feldmann, Arne
    Zysset, Philippe