摘要
An integrated platform comprising a PMMA (Polymethyl-Methacrylate) microfluidic array chip and a LED photometer system is proposed for the concentration detection of sulfur dioxide (SO2) and methanol (CH4O). In the proposed device, the sample and reagent are injected into six open chambers on the microfluidic chip and are mixed via a vortex stirring effect. Following the mixing process, a colorimetric reaction is induced between the sample and the reagent by means of a microhotplate positioned beneath the microfluidic chip array. The microfluidic chip is then transferred to a LED photometer system for detection purposes. The experimental results show that correlation coefficients of R-2 = 0.9915 and R-2 = 0.9941 are obtained when plotting the photometer output voltage against the SO2 and CH4O concentrations of control samples with known concentrations ranging from 50 to 500 ppm. The real-world applicability of the proposed experimental platform is demonstrated by measuring the SO2 concentration in two commercial Chinese herbs and the CH4O concentration in two commercial wines. It is shown that the concentration measurements obtained using the proposed system deviate by no more than 9% from those obtained using standard macroscale methods. Overall, the results presented in this study show that the proposed integrated microfluidic platform provides a compact and reliable tool for SO2 and CH4O concentration measurement purposes. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:421 / 427
页数:7
相关论文
共 54 条
[11]   A high-discernment microflow cytometer with microweir structure [J].
Fu, Lung-Ming ;
Tsai, Chien-Hsiung ;
Lin, Che-Hsin .
ELECTROPHORESIS, 2008, 29 (09) :1874-1880
[12]   A rapid DNA digestion system [J].
Fu, Lung-Ming ;
Lin, Che-Hsin .
BIOMEDICAL MICRODEVICES, 2007, 9 (02) :277-286
[13]   Chaotic vortex micromixer utilizing gas pressure driving force [J].
Fu, Lung-Ming ;
Ju, Wei-Jhong ;
Tsai, Chien-Hsiung ;
Hou, Hui-Hsiung ;
Yang, Ruey-Jen ;
Wang, Yao-Nan .
CHEMICAL ENGINEERING JOURNAL, 2013, 214 :1-7
[14]   Rapid prototyping of glass-based microfluidic chips utilizing two-pass defocused CO2 laser beam method [J].
Fu, Lung-Ming ;
Ju, Wei-Jhong ;
Yang, Ruey-Jen ;
Wang, Yao-Nan .
MICROFLUIDICS AND NANOFLUIDICS, 2013, 14 (3-4) :479-487
[15]   Optical microflow cytometer based on external total reflection [J].
Fu, Lung-Ming ;
Wang, Yao-Nan .
ELECTROPHORESIS, 2012, 33 (21) :3229-3235
[16]   Mixing performance assessment of a multi-channel mini heat exchanger reactor with arborescent distributor and collector [J].
Guo, Xiaofeng ;
Fan, Yilin ;
Luo, Lingai .
CHEMICAL ENGINEERING JOURNAL, 2013, 227 :116-127
[17]   Infrared imaging of temperature profiles in microreactors for fast and exothermic reactions [J].
Haber, J. ;
Kashid, M. N. ;
Borhani, N. ;
Thome, J. ;
Krtschil, U. ;
Renken, A. ;
Kiwi-Minsker, L. .
CHEMICAL ENGINEERING JOURNAL, 2013, 214 :97-105
[18]   Controlling lateral nanomixing and velocity profile of dilute ferrofluid capillary flows in uniform stationary, oscillating and rotating magnetic fields [J].
Hajiani, Pouya ;
Larachi, Faical .
CHEMICAL ENGINEERING JOURNAL, 2013, 223 :454-466
[19]   Rapid prototyping of PMMA microfluidic chips utilizing a CO2 laser [J].
Hong, Ting-Fu ;
Ju, Wei-Jhong ;
Wu, Ming-Chang ;
Tai, Chang-Hsien ;
Tsai, Chien-Hsiung ;
Fu, Lung-Ming .
MICROFLUIDICS AND NANOFLUIDICS, 2010, 9 (06) :1125-1133
[20]   Rapid glucose concentration detection utilizing disposable integrated microfluidic chip [J].
Hou, Hui-Hsiung ;
Wang, Yao-Nan ;
Chang, Chin-Lung ;
Yang, Ruey-Jen ;
Fu, Lung-Ming .
MICROFLUIDICS AND NANOFLUIDICS, 2011, 11 (04) :479-487