Computing the spatial QRS-T angle using reduced electrocardiographic lead sets

被引:1
作者
Guldenring, Daniel [1 ]
Finlay, Dewar D. [1 ]
Bond, Raymond R. [1 ]
Kennedy, Alan [1 ]
McLaughlin, James [1 ]
Galeotti, Loriano [2 ]
Strauss, David G. [2 ]
机构
[1] Univ Ulster, Jordanstown Campus,Shore Rd, Newtownabbey, Antrim, North Ireland
[2] US FDA, Off Sci & Engn Labs, CDRH, Silver Spring, MD USA
关键词
Linear lead transformations; Estimation of the Frank VCG; Derivation of the Frank VCG; Spatial QRS-T angle; Mason-Likar; 12-lead; ECG; Monitoring; 12-LEAD ELECTROCARDIOGRAM; VENTRICULAR GRADIENT; VECTORCARDIOGRAM; MATRIX; SYSTEM; ECG;
D O I
10.1016/j.jelectrocard.2016.07.015
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The 'spatial QRS-T angle' (SA) is frequently determined using linear lead transformation matrices that require the entire 12-lead electrocardiogram (ECG). While this approach is adequate when using 12-lead ECG data that is recorded in the resting supine position, it is not optimal in monitoring applications. This is because maintaining a good quality recording of the complete 12-lead ECG in monitoring applications is difficult. In this research, we assessed the differences between the 'gold standard' SA as determined using the Frank VGG and the SA as determined using different reduced lead systems (RLSs). The random error component (span of the Bland-Altman 95% limits of agreement) of the differences between the 'gold standard' SA and the SA values based upon the different RLSs was quantified. This was performed for all 62 RLSs that can be constructed from Mason-Likar (ML) limb leads I, II and all possible precordial lead subsets that contain between one and five of the precordial leads VI to V6. The RLS with the smallest lead set size that produced SA estimates of a quality similar to what is achieved using the ML 12-lead ECG was based upon ML limb leads I, II and precordial leads V1, V3 and V6. The random error component (mean [95% confidence interval]) associated with this RLS and the ML 12-lead ECG were found to be 40.74 [35.56 degrees-49.29 degrees] and 39.57 degrees [33.78 degrees-45.70 degrees], respectively. Our findings suggest that a RLS that is based upon the ML limb leads I and II and the three best precordial leads can yield SA estimates of a quality similar to what is achieved when using the complete ML 12-lead ECG. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:794 / 799
页数:6
相关论文
共 22 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Predicting Ventricular Arrhythmias in Patients With Ischemic Heart Disease Clinical Application of the ECG-Derived QRS-T Angle [J].
Borleffs, C. Jan Willem ;
Scherptong, Roderick W. C. ;
Man, Sum-Che ;
van Welsenes, Guido H. ;
Bax, Jeroen J. ;
van Erven, Lieselot ;
Swenne, Cees A. ;
Schalij, Martin J. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2009, 2 (05) :548-554
[3]  
BURGER HC, 1946, BRIT HEART J, V8, P157
[4]   When deriving the spatial QRS-T angle from the 12-lead electrocardiogram, which transform is more Frank: regression or inverse Dower? [J].
Cortez, Daniel L. ;
Schlegel, Todd T. .
JOURNAL OF ELECTROCARDIOLOGY, 2010, 43 (04) :302-309
[5]   Incremental prognostic value of an abnormal baseline spatial QRS-T angle in chronic dialysis patients [J].
de Bie, Mihaly K. ;
Koopman, Marion G. ;
Gaasbeek, Andre ;
Dekker, Friedo W. ;
Maan, Arie C. ;
Swenne, Cees A. ;
Scherptong, Roderick W. ;
van Dessel, Pascal F. ;
Wilde, Arthur A. ;
Schalij, Martin J. ;
Rabelink, Ton J. ;
Jukema, J. Wouter .
EUROPACE, 2013, 15 (02) :290-296
[6]   Elucidation of the spatial ventricular gradient and its link with dispersion of repolarization [J].
Draisma, Harmen H. M. ;
Schalij, Martin J. ;
van der Wall, Ernst E. ;
Swenne, Cees A. .
HEART RHYTHM, 2006, 3 (09) :1092-1099
[7]   DERIVED 12-LEAD ECG - COMPARISON WITH THE STANDARD ECG DURING MYOCARDIAL-ISCHEMIA AND ITS POTENTIAL APPLICATION FOR CONTINUOUS ST-SEGMENT MONITORING [J].
DREW, BJ ;
KOOPS, RR ;
ADAMS, MG ;
DOWER, GE .
JOURNAL OF ELECTROCARDIOLOGY, 1994, 27 :249-255
[8]   VECTORCARDIOGRAM SYNTHESIZED FROM A 12-LEAD ECG - SUPERIORITY OF THE INVERSE DOWER MATRIX [J].
EDENBRANDT, L ;
PAHLM, O .
JOURNAL OF ELECTROCARDIOLOGY, 1988, 21 (04) :361-367
[9]   Effects of electrode placement errors in the EASI-derived 12-lead electrocardiogram [J].
Finlay, Dewar D. ;
Nugent, Chris D. ;
Nelwan, Stefan P. ;
Bond, Raymond R. ;
Donnelly, Mark P. ;
Guldenring, Daniel .
JOURNAL OF ELECTROCARDIOLOGY, 2010, 43 (06) :606-611
[10]   GENERAL THEORY OF HEART-VECTOR PROJECTION [J].
FRANK, E .
CIRCULATION RESEARCH, 1954, 2 (03) :258-270