Optimization of capillary liquid chromatography with electrochemical detection for determining femtogram levels of baicalin and baicalein on the basis of the FUMI theory

被引:10
作者
Kotani, Akira [1 ]
Kojima, Satoshi [1 ]
Hayashi, Yuzuru [2 ]
Matsuda, Rieko [2 ]
Kusu, Fumiyo [1 ]
机构
[1] Tokyo Univ Pharm & Life Sci, Sch Pharm, Tokyo 1920392, Japan
[2] Natl Inst Hlth Sci, Setagaya Ku, Tokyo 1588501, Japan
关键词
Capillary liquid chromatography; Electrochemical detection; FUMI theory; Optimization;
D O I
10.1016/j.jpba.2008.08.002
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Capillary liquid chromatography with electrochemical detection (CLC-ECD) was developed for determining the femtogram levels of baicalin and baicalein. The CLC-ECD system and the experimental conditions were optimized based on the precision (=information content) phi and efficiency (=information content/time) theta, which were calculated from a relative standard deviation (R.S.D.) by a chemometric tool called the FUMI theory. When CLC-ECD was established using a capillary column (Inertsil ODS-3, 150 mm x 0.2 mm i.d.), a sample injector fitted witha 0.2 mu Linjection loop, an applied potential of +650 mV vs. Ag/AgCl, and a flow rate at 1.8 mu L/min, baicalin and baicalein were determined at femtogram levels. Moreover, the present method was validated using a chemometric tool and a conventional method. Since the FUMI theory makes it possible to predict R.S.D. without repetitive measurements, the chemometric tool saves considerable amounts of chemicals and experimental time, and was found to be useful for the optimization of conditions and validation for determination by CLC-ECD. The present method was applied to the analysis of Japanese Pharmacopoeia Scutellaria Root and Scutellaria baicalensis Georgi for determining baicalin and baicalein.
引用
收藏
页码:780 / 787
页数:8
相关论文
共 46 条
[41]  
VANDERHEYDEN Y, 2000, HDB ANAL SEPARATIONS, V1, P163
[42]   Operational variables in high-performance liquid chromatography-electrospray ionization mass spectrometry of peptides and proteins using poly(styrene-divinylbenzene) monoliths [J].
Walcher, W ;
Toll, H ;
Ingendoh, A ;
Huber, CG .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1053 (1-2) :107-117
[43]   Interaction of baicalin and baicalein with antibiotics in the gastrointestinal tract [J].
Xing, J ;
Chen, XY ;
Sun, YM ;
Luan, Y ;
Zhong, DF .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 2005, 57 (06) :743-750
[44]   New analytical possibilities of amperometric detection in high-performance liquid chromatography [J].
Yashin, AY .
JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 55 (11) :1092-1095
[45]   Statistical optimization of a reversed-phase liquid chromatographic method for the analysis of amiloride and hydrochlorothiazide in tablets [J].
Zecevic, M ;
Zivanovic, LJ ;
Agatonovic-Kustrin, S ;
Ivanovic, D ;
Maksimovic, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 22 (01) :1-6
[46]   Quantitative determination of 14 major constituents in the herbal preparation Luan-Pao-Prescription using HPLC coupled with photodiode array detection [J].
Zhang, Jin-qiang ;
Wang, Xiao-ming ;
Lu, Zhi-qiang ;
Huang, Hui-lian ;
Chen, Guang-tong ;
Liu, Rong-xia ;
Bi, Kai-Shun ;
Guo, De-an .
CHROMATOGRAPHIA, 2007, 66 (3-4) :267-270