A retained guidewire fractured with subsequent pericardial tamponade two years after endovascular neurointervention

被引:4
作者
Han, Zongli [1 ]
Du, Yanli [2 ]
Chen, Jinhui [1 ]
Qi, Hui [1 ]
机构
[1] Peking Univ, Shenzhen Hosp, Dept Neurosurg, Shenzhen, Peoples R China
[2] Shenzhen Polytech, Sch Med Technol & Nursing, Shenzhen, Peoples R China
关键词
Retained wire; wire fracture; interventional neurosurgery; complication; pericardial tamponade;
D O I
10.1177/1591019918801538
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Entrapment of aneurysm embolization hardware is an extremely rare complication of endovascular neurointerventional procedures. We describe a case of a retained guidewire in a 42-year-old male during an aneurysm embolization. After unsuccessful attempts at removal via interventional methods, we decided to leave the guidewire within the vessel. A guidewire fracture resulted in several fragments in the carotid artery and aorta with subsequent cardiac tamponade, pseudoaneurysm and aortojejunal fistula two years later. The fragments in the aorta were removed via interventional and surgical methods. We advocate early surgical management of the retained guidewires after unsuccessful retractions via interventional methods. Meticulous and gentle maneuvering is necessary to prevent such serious complications.
引用
收藏
页码:117 / 120
页数:4
相关论文
共 6 条
[1]   Endovascular Rescue Method for Undesirably Stretched Coil [J].
Cho, Jae Hoon .
JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2014, 56 (04) :364-367
[2]   Consequences and treatment of guidewire entrapment and fracture during percutaneous coronary intervention [J].
Danek, Barbara A. ;
Karatasakis, Aris ;
Brilakis, Emmanouil S. .
CARDIOVASCULAR REVASCULARIZATION MEDICINE, 2016, 17 (02) :129-133
[3]  
Konstas A A, 2007, Br J Radiol, V80, pe290, DOI 10.1259/bjr/52196291
[4]   Fracture and migration of a retained wire into the thoracic cavity after endovascular neurointervention: report of 2 cases [J].
Koo, Hae-Won ;
Park, Wonhyoung ;
Yang, Kuhyun ;
Park, Jung Cheol ;
Ahn, Jae Sung ;
Kwon, Sun Uck ;
Hwang, Changmo ;
Lee, Deok Hee .
JOURNAL OF NEUROSURGERY, 2017, 126 (02) :354-367
[5]   Surgical management for retained distal embolic protection device and fractured guidewire after carotid artery stenting [J].
Li, Tongxun ;
Zha, Yuanting ;
Bo, Liyang ;
Wirthlin, Douglas J. ;
Zhang, Qinyi .
JOURNAL OF SURGICAL CASE REPORTS, 2016, (06)
[6]  