T1 Mapping MOLLI 5(3)3 Acquisition Scheme Yields High Accuracy in 1.5 T Cardiac Magnetic Resonance

被引:3
作者
Krumm, Patrick [1 ]
Martirosian, Petros [2 ]
Brendel, Alexander [1 ]
Kuebler, Jens M. [1 ]
Brendel, Jan M. [1 ]
Gassenmaier, Sebastian [1 ]
Estler, Arne [1 ]
Gawaz, Meinrad [3 ]
Nikolaou, Konstantin [1 ]
Greulich, Simon [3 ]
机构
[1] Univ Tubingen, Dept Radiol Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
[2] Univ Tubingen, Dept Radiol, Sect Expt Radiol, D-72076 Tubingen, Germany
[3] Univ Tubingen, Dept Internal Med Cardiol & Angiol 3, D-72076 Tubingen, Germany
关键词
T1; mapping; cardiac magnetic resonance imaging; MOLLI; accuracy;
D O I
10.3390/diagnostics12112729
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To systematically compare two modified Look-Locker inversion recovery (MOLLI) T1 mapping sequences and their impact on (1) myocardial T1 values native, (2) post-contrast and (3) extracellular volume (ECV). Methods: 200 patients were prospectively included for 1.5 T CMR for work-up of ischemic or non-ischemic cardiomyopathies. To determine native and post-contrast T1 for ECV calculation, two different T1 mapping MOLLI acquisition schemes, 5(3)3 (designed for native scans with long T1) and 4(1)3(1)2 (designed for post-contrast scans with short T1), were acquired in identical mid-ventricular short-axis slices. Both schemes were acquired in native and post-contrast scans. Results: Datasets from 163 patients were evaluated (age 55 +/- 17 years; 38% female). Myocardial T1 native for 5(3)3 was 1017 +/- 42 ms vs. 956 +/- 40 ms for 4(1)3(1)2, with mean intraindividual difference -61 ms (p < 0.0001). Post-contrast myocardial T1 in patients was similar for both acquisition schemes, with 494 +/- 48 ms for 5(3)3 and 490 +/- 45 ms for 4(1)3(1)2 and mean intraindividual difference -4 ms. Myocardial ECV for 5(3)3 was 27.6 +/- 4% vs. 27 +/- 4% for 4(1)3(1)2, with mean difference -0.6 percentage points (p < 0.0001). Conclusions: The T1 MOLLI 5(3)3 acquisition scheme provides a reliable estimation of myocardial T1 for the clinically relevant range of long and short T1 values native and post-contrast. In contrast, the T1 MOLLI 4(1)3(1)2 acquisition scheme may only be used for post-contrast scans according to its designed purpose.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Global and Regional Test-Retest Reproducibility of Native T1 and T2 Mapping in Cardiac Magnetic Resonance Imaging
    Bottcher, Benjamin
    Lorbeer, Roberto
    Stocklein, Sophia
    Beller, Ebba
    Lang, Cajetan I.
    Weber, Marc-Andre
    Meinel, Felix G.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2021, 54 (06) : 1763 - 1772
  • [32] Assessment of cardiac involvement in myotonic muscular dystrophy by T1 mapping on magnetic resonance imaging
    Turkbey, Evrim B.
    Gai, Neville
    Lima, Joao A. C.
    van der Geest, Rob J.
    Wagner, Kathryn R.
    Tomaselli, Gordon F.
    Bluemke, David A.
    Nazarian, Saman
    HEART RHYTHM, 2012, 9 (10) : 1691 - 1697
  • [33] Parallel simulations for QUAntifying RElaxation magnetic resonance constants (SQUAREMR): an example towards accurate MOLLI T1 measurements
    Christos G. Xanthis
    Sebastian Bidhult
    George Kantasis
    Einar Heiberg
    Håkan Arheden
    Anthony H. Aletras
    Journal of Cardiovascular Magnetic Resonance, 17
  • [34] A medical device-grade T1 and ECV phantom for global T1 mapping quality assurance-the T1 Mapping and ECV Standardization in cardiovascular magnetic resonance (T1MES) program
    Captur, Gabriella
    Gatehouse, Peter
    Keenan, Kathryn E.
    Heslinga, Friso G.
    Bruehl, Ruediger
    Prothmann, Marcel
    Graves, Martin J.
    Eames, Richard J.
    Torlasco, Camilla
    Benedetti, Giulia
    Donovan, Jacqueline
    Ittermann, Bernd
    Boubertakh, Redha
    Bathgate, Andrew
    Royet, Celine
    Pang, Wenjie
    Nezafat, Reza
    Salerno, Michael
    Kellman, Peter
    Moon, James C.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2016, 18 : 1 - 20
  • [35] T1 Mapping for Myocardial Fibrosis by Cardiac Magnetic Resonance Relaxometry - A Comprehensive Technical Review
    Hamilton-Craig, Christian R.
    Strudwick, Mark W.
    Galloway, Graham J.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2017, 3
  • [36] Cardiac magnetic resonance T1 mapping. Part 2: Diagnostic potential and applications
    Reiter, Ursula
    Reiter, Clemens
    Kraeuter, Corina
    Fuchsjaeger, Michael
    Reiter, Gert
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 109 : 235 - 247
  • [37] Detection of cardiac iron overload with native magnetic resonance T1 and T2 mapping in patients with thalassemia
    Krittayaphong, Rungroj
    Zhang, Shuo
    Saiviroonporn, Pairash
    Viprakasit, Vip
    Tanapibunpon, Prajak
    Komoltri, Chulaluk
    Wangworatrakul, Wipaporn
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2017, 248 : 421 - 426
  • [38] Parallel simulations for QUAntifying RElaxation magnetic resonance constants (SQUAREMR): an example towards accurate MOLLI T1 measurements
    Xanthis, Christos G.
    Bidhult, Sebastian
    Kantasis, George
    Heiberg, Einar
    Arheden, Hakan
    Aletras, Anthony H.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2015, 17
  • [39] Cardiac metastasis of neuroendocrine tumor of the thymus diagnosed by cardiac magnetic resonance T1 mapping: a case report
    Zhang, Jiamin
    Zeng, Mu
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2020, 48 (06)
  • [40] 3D myocardial T1 mapping using saturation recovery
    Nordio, Giovanna
    Henningsson, Markus
    Chiribiri, Amedeo
    Villa, Adriana D. M.
    Schneider, Torben
    Botnar, Rene M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2017, 46 (01) : 218 - 227