Improved SPECT quantitation using fully three-dimensional iterative spatially variant scatter response compensation

被引:69
作者
Beekman, FJ
Kamphuis, C
Viergever, MA
机构
[1] Department of Radiology and Nuclear Medicine, University Hospital Utrecht, 3584CX Utrecht
关键词
D O I
10.1109/42.511752
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The quality and quantitative accuracy of iteratively reconstructed SPECT images improves when better point spread function (PSF) models of the gamma camera are used during reconstruction, In this paper, inclusion in the PSF model of photon crosstalk between different slices caused by limited gamma camera resolution and scatter is examined, A three-dimensional (3-D) projector back-projector (proback) has been developed which models both the distance dependent detector point spread function and the object shape-dependent scatter point spread function of single photon emission computed tomography (SPECT), A table occupying only a few megabytes of memory is sufficient to represent this scatter model, The contents of this table are obtained by evaluating an analytical expression for object shape-dependent scatter, The proposed approach avoids the huge memory requirements of storing the full transition matrix needed for 3-D reconstruction including object shape-dependent scatter. In addition, the method avoids the need for lengthy Monte Carlo simulations to generate such a matrix, In order to assess the quantitative accuracy of the method, reconstructions of a water filled cylinder containing regions of different activity levels and of simulated 3-D brain projection data have been evaluated for technetium-99m. It is shown that fully 3-D reconstruction including complete detector response and object shape-dependent scatter modeling clearly outperforms simpler methods that lack a complete detector response and/or a complete scatter response model, Fully 3-D scatter correction yields the best quantitation of volumes of interest and the best contrast-to-noise curves.
引用
收藏
页码:491 / 499
页数:9
相关论文
共 35 条
[1]   FAST SPECT SIMULATION INCLUDING OBJECT SHAPE DEPENDENT SCATTER [J].
BEEKMAN, FJ ;
VIERGEVER, MA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (02) :271-282
[2]   OBJECT SHAPE DEPENDENT PSF MODEL FOR SPECT IMAGING [J].
BEEKMAN, FJ ;
EIJKMAN, EGJ ;
VIERGEVER, MA ;
BORM, GF ;
SLIJPEN, ETP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (01) :31-39
[3]   A NEW PHANTOM FOR FAST DETERMINATION OF THE SCATTER RESPONSE OF A GAMMA-CAMERA [J].
BEEKMAN, FJ ;
FREY, EC ;
KAMPHUIS, C ;
TSUI, BMW ;
VIERGEVER, MA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1481-1488
[4]  
BEEKMAN FJ, 1995, INFORMATION PROCESSI, V14, P53
[5]  
BEEKMAN FJ, 1992, 3DCV UTR U
[6]  
BOWSHER JE, 1991, J NUCL MED, V32, P1285
[7]  
FLOYD CE, 1986, J NUCL MED, V27, P1577
[8]  
FORMICONI AR, 1989, PHYS MED BIOL, V34, P68
[9]   A PRACTICAL METHOD FOR INCORPORATING SCATTER IN A PROJECTOR-BACKPROJECTOR FOR ACCURATE SCATTER COMPENSATION IN SPECT [J].
FREY, EC ;
TSUI, BMW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) :1107-1116
[10]   MODELING THE SCATTER RESPONSE FUNCTION IN INHOMOGENEOUS SCATTERING MEDIA FOR SPECT [J].
FREY, EC ;
TSUI, BMW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1585-1593