Validation of an ultra-performance liquid chromatography-tandem mass spectrometry method for the determination of flecainide in human plasma and its clinical application

被引:6
作者
Mano, Yuji [1 ]
Asakawa, Yoshiki [2 ]
Kita, Kenji [2 ]
Ishii, Takuho [2 ]
Hotta, Koichiro [1 ]
Kusano, Kazutomi [1 ]
机构
[1] Eisai & Co Ltd, Biopharmaceut Assessment Core Funct Unit, Drug Metab & Pharmacokinet, Tsukuba, Ibaraki 3002635, Japan
[2] Sunplanet Co Ltd, Tsukuba Div, Anal Grp, Tsukuba, Ibaraki 3002635, Japan
关键词
UPLC-MS; MS; flecainide; validation; ISR; hemolysis; SUPRAVENTRICULAR TACHYCARDIA; SERUM FLECAINIDE; CHILDREN; THERAPY;
D O I
10.1002/bmc.3437
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple and reproducible bioanalytical method for the determination of flecainide in human plasma was developed and validated using an ultra-performance liquid chromatography with tandem mass spectrometry (UPLC-MS/MS) to obtain higher sensitivity than the current available methods. After simple protein precipitation, flecainide and a stable isotope-labeled internal standard (IS) were chromatographed on an Acquity UPLC BEH C-18 column (2.1 x 100 mm, 1.7 mu m) with isocratic elution of mobile phase consisting of 45% methanol containing 0.1% formic acid at a flow rate 0.25 mL/min. Detection was performed in positive electrospray ionization by monitoring the selected ion transitions at m/z 415.4/301.1 for flecainide and m/z 419.4/305.1 for the IS. The method was validated according to current bioanalytical method validation guidelines. The calibration standard curve was linear from 2.5 to 1000 ng/mL using 0.1 mL of plasma. No significant interferences were detected in blank human plasma. Accuracy and precision in the intra- and inter-batch reproducibility study were within acceptance criteria. Neither hemolysis effects nor matrix effects were observed. The UPLC-MS/MS method developed was successfully applied to determine plasma flecainide concentrations to support clinical studies and incurred sample reanalysis also ensured the reproducibility of the method. Copyright (c) 2015 John Wiley & Sons, Ltd.
引用
收藏
页码:1399 / 1405
页数:7
相关论文
共 11 条
[1]   Therapeutic drug monitoring of flecainide in serum using high-performance liquid chromatography and electrospray mass spectrometry [J].
Breindahl, T .
JOURNAL OF CHROMATOGRAPHY B, 2000, 746 (02) :249-254
[2]   Simultaneous determination of serum flecainide and its metabolites by using high performance liquid chromatography [J].
Doki, K ;
Homma, M ;
Kuga, K ;
Watanabe, S ;
Yamaguchi, I ;
Kohda, Y .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 35 (05) :1307-1312
[3]   FLECAINIDE AND AMIODARONE - COMBINED THERAPY FOR REFRACTORY TACHYARRHYTHMIAS IN INFANCY [J].
FENRICH, AL ;
PERRY, JC ;
FRIEDMAN, RA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (05) :1195-1198
[4]   GAS-CHROMATOGRAPHY MASS-SPECTROMETRY VALIDATION OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ANALYSIS OF FLECAINIDE ENANTIOMERS IN SERUM [J].
FISCHER, C ;
BUHL, K .
THERAPEUTIC DRUG MONITORING, 1992, 14 (05) :433-435
[5]   High performance liquid chromatography for routine monitoring of serum flecainide [J].
Nakagawa, R ;
Homma, M ;
Kuga, K ;
Yamaguchi, I ;
Kobayashi, D ;
Morimoto, Y ;
Kohda, Y .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 30 (01) :171-174
[6]   Flecainide and sotalol:: A new combination therapy for refractory supraventricular tachycardia in children <1 year of age [J].
Price, JF ;
Kertesz, NJ ;
Snyder, CS ;
Friedman, RA ;
Fenrich, AL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (03) :517-520
[7]   Antiarrhythmic Drug Therapy for Maintaining Sinus Rhythm Early after Pulmonary Vein Ablation in Patients with Symptomatic Atrial Fibrillation [J].
Sohns, Christian ;
von Gruben, Valerie ;
Sossalla, Samuel ;
Bergau, Leonard ;
Seegers, Joachim ;
Luethje, Lars ;
Vollmann, Dirk ;
Zabel, Markus .
CARDIOVASCULAR THERAPEUTICS, 2014, 32 (01) :7-12
[8]   Appropriate dosing of antiarrhythmic drugs in Japan requires therapeutic drug monitoring [J].
Takada, M ;
Goto, T ;
Kotake, T ;
Saito, M ;
Kawato, N ;
Nakai, M ;
Gunji, T ;
Shibakawa, M .
JOURNAL OF CLINICAL PHARMACY AND THERAPEUTICS, 2005, 30 (01) :5-12
[9]   USE OF FLECAINIDE ACETATE FOR REFRACTORY JUNCTIONAL TACHYCARDIAS IN CHILDREN WITH THE WOLFF-PARKINSON-WHITE SYNDROME [J].
WARD, DE ;
JONES, S ;
SHINEBOURNE, EA .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (10) :787-790
[10]  
WREN C, 1987, BRIT HEART J, V57, P171