Effect of inhaled gas density on the pendelluft-induced lung injury

被引:3
作者
Alzahrany, Mohammed [1 ]
Banerjee, Arindam [1 ]
机构
[1] Lehigh Univ, Dept Mech Engn & Mech, 19 Mem Dr West, Bethlehem, PA 18015 USA
关键词
Pendelluft flow; High frequency ventilation; Lung injury; Helium; Sulfur hexafluoride; Loops; FREQUENCY OSCILLATORY VENTILATION; UPPER TRACHEOBRONCHIAL AIRWAYS; HELIUM-OXYGEN MIXTURES; AIR-FLOW; PARTICLE DEPOSITION; COMPUTATIONAL MODEL; AEROSOL DEPOSITION; FLUID-DYNAMICS; LARGE-EDDY; SIMULATION;
D O I
10.1016/j.jbiomech.2016.10.041
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Helium, sulfur hexafluoride-oxygen, and air were modeled to examine the role of the gas density on the pendelluft-induced lung injury (PILI) under high frequency oscillatory ventilation (HFOV). Large eddy simulation coupled with physiological resistance-compliance boundary conditions was applied to capture pendelluft-induced gas entrapment and mechanical stresses in an image-based human lung model. The flow characteristics were strongly dependent on the inspired gas density. The flow partitioning, globally between the left and right lung and locally between adjacent units branches, was significantly affected by the density of inhaled gas and was more balanced when inspiring lighter gas. The incomplete loops of flow-volume and volume-pressure curves were significantly influenced by the variations of the flow redistribution, resistance, and turbulence associated with the pendelluft mechanism. Inhaling light gas reduced the entrapped gas volume and mechanical stress surrounding carina ridges signifying the important role of inhaled gas properties on PILL In general, lung ventilation by HFOV with a gas mixture of large amounts of Helium is thought to mitigate ventilator complications. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:4039 / 4047
页数:9
相关论文
共 45 条
[1]   Role of carrier gases in enhancement of gas exchange and aerosol-drug delivery under invasive high frequency oscillatory ventilation [J].
Alzahrany, Mohammed ;
Banerjee, Arindam .
JOURNAL OF AEROSOL SCIENCE, 2015, 88 :1-18
[2]   A biomechanical model of pendelluft induced lung injury [J].
Alzahrany, Mohammed ;
Banerjee, Arindam .
JOURNAL OF BIOMECHANICS, 2015, 48 (10) :1804-1810
[3]   Aerosolized drug delivery in patient-specific lung model during invasive high frequency oscillatory ventilation [J].
Alzahrany, Mohammed ;
Banerjee, Arindam .
JOURNAL OF AEROSOL SCIENCE, 2015, 81 :1-20
[4]   The role of coupled resistance-compliance in upper tracheobronchial airways under high frequency oscillatory ventilation [J].
Alzahrany, Mohammed ;
Banerjee, Arindam ;
Salzman, Gary .
MEDICAL ENGINEERING & PHYSICS, 2014, 36 (12) :1593-1604
[5]   Flow transport and gas mixing during invasive high frequency oscillatory ventilation [J].
Alzahrany, Mohammed ;
Banerjee, Arindam ;
Salzman, Gary .
MEDICAL ENGINEERING & PHYSICS, 2014, 36 (06) :647-658
[6]  
Andrey V, 2012, PULM MED, V2012, P1
[7]  
[Anonymous], 2001, Measurement Science and Technology
[8]  
[Anonymous], 2005, LARGE EDDY SIMULATIO
[9]  
Blanch Lluis, 2005, Respir Care, V50, P110
[10]   MECHANISMS OF GAS-TRANSPORT DURING VENTILATION BY HIGH-FREQUENCY OSCILLATION [J].
CHANG, HK .
JOURNAL OF APPLIED PHYSIOLOGY, 1984, 56 (03) :553-563