Design of a dispersion interferometer combined with a polarimeter to increase the electron density measurement reliability on ITER

被引:19
作者
Akiyama, T. [1 ]
Sirinelli, A. [2 ]
Watts, C. [2 ]
Shigin, P. [3 ]
Vayakis, G. [2 ]
Walsh, M. [2 ]
机构
[1] Natl Inst Nat Sci, Natl Inst Fus Sci, 322-6 Oroshi Cho, Toki, Gifu 5095292, Japan
[2] ITER Org, Route Vinon Sur Verdon,CS 90 046, F-13067 St Paul Les Durance, France
[3] NRNU MEPhI, 31 Kashirskoe Sh, Moscow 115409, Russia
关键词
Vibrations; (mechanical); -; Dispersions; Interferometers; Polarimeters;
D O I
10.1063/1.4962050
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A dispersion interferometer is a reliable density measurement system and is being designed as a complementary density diagnostic on ITER. The dispersion interferometer is inherently insensitive to mechanical vibrations, and a combined polarimeter with the same line of sight can correct fringe jump errors. A proof of the principle of the CO2 laser dispersion interferometer combined with the PEM polarimeter was recently conducted, where the phase shift and the polarization angle were successfully measured simultaneously. Standard deviations of the line-average density and the polarization angle measurements over 1 s are 9 x 10(16) m(-2) and 0.19 degrees, respectively, with a time constant of 100 mu s. Drifts of the zero point, which determine the resolution in steady-state operation, correspond to 0.25% and 1% of the phase shift and the Faraday rotation angle expected on ITER. Published by AIP Publishing.
引用
收藏
页数:4
相关论文
共 15 条
[1]   Conceptual design of high resolution and reliable density measurement system on helical reactor FFHR-d1 and demonstration on LHD [J].
Akiyama, T. ;
Yasuhara, R. ;
Tanaka, T. ;
Isobe, M. ;
Sakamoto, R. ;
Goto, T. ;
Kawahata, K. ;
Nakayama, K. ;
Okajima, S. ;
Sagara, A. .
NUCLEAR FUSION, 2015, 55 (09)
[2]   Dispersion interferometer using modulation amplitudes on LHD [J].
Akiyama, T. ;
Yasuhara, R. ;
Kawahata, K. ;
Okajima, S. ;
Nakayama, K. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (11)
[3]   Conceptual design of electron density measurement system for DEMO-relevant helical plasmas [J].
Akiyama, Tsuyoshi ;
Yasuhara, Ryo ;
Tokuzawa, Tokihiko ;
Isobe, Mitsutaka ;
Kawahata, Kazuo ;
Nakayama, Kazuya ;
Okajima, Shigeki .
Plasma and Fusion Research, 2012, 7 (SPL.ISS.1)
[4]  
Akiyama Tsuyoshi, 2010, Plasma and Fusion Research, V5, DOI 10.1585/pfr.5.S1041
[5]   CO2 laser polarimeter for electron density profile measurement on the Large Helical Device [J].
Akiyama, T ;
Tsuji-Iio, S ;
Shimada, R ;
Nakayama, K ;
Okajima, S ;
Takahashi, M ;
Terai, K ;
Tanaka, K ;
Tokuzawa, T ;
Kawahata, K .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (05) :2695-2703
[6]   Short wavelength far infrared laser polarimeter with silicon photoelastic modulators [J].
Akiyama, T. ;
Kawahata, K. ;
Tanaka, K. ;
Okajima, S. ;
Nakayama, K. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2008, 79 (10)
[7]  
Akiyama T., REV SCI INSTRU UNPUB
[8]  
[Anonymous], 2015, ITER INTERNAL DOCUME
[9]   CO2 laser-based dispersion interferometer utilizing orientation-patterned gallium arsenide for plasma density measurements [J].
Bamford, D. J. ;
Cummings, E. A. ;
Panasenko, D. ;
Fenner, D. B. ;
Hensley, J. M. ;
Boivin, R. L. ;
Carlstrom, T. N. ;
Van Zeeland, M. A. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (09)
[10]   DISPERSION INTERFEROMETER FOR CONTROLLED FUSION DEVICES [J].
DRACHEV, VP ;
KRASNIKOV, YI ;
BAGRYANSKY, PA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (04) :1010-1013