Evaluation of commercial immunoassays for the detection of estrogens in water by comparison with high-performance liquid chromatography tandem mass spectrometry HPLC-MS/MS (QqQ)

被引:35
作者
Farre, Marinella
Brix, Rikke
Kuster, Marina
Rubio, Fernando
Goda, Yasuhiro
de Alda, Mara J. Lopez
Barcelo, Damia
机构
[1] CSIC, IIQAB, Dept Environm Chem, ES-08034 Barcelona, Spain
[2] UAB, Dept Analyt Chem, Sensors & Biosensors Grp, Bellaterra 08193, Cerdanyola Vall, Spain
[3] Abraxis, Warminster, PA 18974 USA
[4] Japan EnviroChem Ltd, Yodogawa Ku, Osaka 5320024, Japan
关键词
immunoassay; ELISA; estrogens; LC-MS/MS;
D O I
10.1007/s00216-006-0562-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this work four different commercially available enzyme-linked immunosorbent assays (ELISA) (from Japan EnviroChemicals, Ltd., Tokyo, Japan) were evaluated in terms of performance for the rapid screening of estrogens in different water matrices, including natural and spiked samples from urban wastewater, river water and ground water. All four test kits are based on monoclonal antibodies. The compounds detected by these immunoassays are (1) 17-beta-estradiol, (2) estrone, (3) 17-alpha-ethynyl estradiol and (4) estrogens in general, with high recognition properties for 17-beta-estradiol, estrone and estriol. Standards were prepared in water containing 10% (v/v) methanol. The IC (50) (corresponding to the 50% of the effective concentration) values, the dynamic ranges, and the limits of detection of the ELISA kits were 0.060-0.304 mu g/L, 0.05-5 mu g/L and 0.05 mu g/L, respectively. All samples were extracted by solid-phase extraction (SPE) beforehand, and the evaluation was carried out by comparing the results obtained by ELISA with those obtained by HPLC-MS/MS using a triple quadrupole (QqQ) instrument. In addition, two different solid-phase extraction procedures were carried out and compared. Except for moderate overestimation in the results observed with the ELISA kits in the analysis of complex wastewater samples, the results obtained using all of the tested techniques were generally in very good agreement.
引用
收藏
页码:1001 / 1011
页数:11
相关论文
共 23 条
  • [1] High frequency of sub-optimal semen quality in an unselected population of young men
    Andersen, AG
    Jensen, TK
    Carlsen, E
    Jorgensen, N
    Andersson, AM
    Krarup, T
    Keiding, N
    Skakkebæk, NE
    [J]. HUMAN REPRODUCTION, 2000, 15 (02) : 366 - 372
  • [2] Fate of estrogens in a municipal sewage treatment plant
    Andersen, H
    Siegrist, H
    Halling-Sorensen, B
    Ternes, TA
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2003, 37 (18) : 4021 - 4026
  • [3] Monitoring natural and synthetic estrogens at activated sludge sewage treatment plants and in a receiving river water
    Baronti, C
    Curini, R
    D'Ascenzo, G
    Di Corcia, A
    Gentili, A
    Samperi, R
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2000, 34 (24) : 5059 - 5066
  • [4] Environmental influence on reproductive health
    Bhatt, RV
    [J]. INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2000, 70 (01) : 69 - 75
  • [5] Enterolactone as a risk factor for breast cancer: A review of the published evidence
    Boccardo, F
    Puntoni, M
    Guglielmini, P
    Rubagotti, A
    [J]. CLINICA CHIMICA ACTA, 2006, 365 (1-2) : 58 - 67
  • [6] CRUZ MSD, 2003, J MASS SPECTROM, V38, P917
  • [7] Davis D.L., 1997, SCI MED, V4, P56
  • [8] Determination of steroid sex hormones and related synthetic compounds considered as endocrine disrupters in water by fully automated on-line solid-phase extraction-liquid chromatography-diode array detection
    de Alda, MJL
    Barceló, D
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2001, 911 (02) : 203 - 210
  • [9] Determination of steroid sex hormones and related synthetic compounds considered as endocrine disrupters in water by liquid chromatography-diode array detection-mass spectrometry
    de Alda, MJL
    Barceló, D
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2000, 892 (1-2) : 391 - 406
  • [10] *EC HLTH CONS PROT, 2005, SANCO104392005