Precise determination of 235U/238U isotope ratio in soil samples by using thermal ionisation mass spectrometry

被引:19
作者
Sahoo, SK [1 ]
Yonehara, H [1 ]
Kurotaki, K [1 ]
Fujimoto, K [1 ]
Nakamura, Y [1 ]
机构
[1] Natl Inst Radiol Sci, Inage Ku, Chiba 2638555, Japan
关键词
D O I
10.1023/A:1015745717871
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A rapid method based on extraction chromatography was developed for the separation of uranium from soil samples for TIMS. The isotope ratios were measured on three standards using a VG 54-30 thermal ionisation mass spectrometer in static mode with Faraday cup and Daly ion counting system. The use of a WARP (Wide Aperture Retardation Potential) energy filter improves abundance sensitivity by an order of magnitude over a conventional VG Sector TIMS. An abundance sensitivity of 28 ppb at mass 237 with respect to U-238 can be achieved. By correcting mass bias as a function of ion beam intensity, precision of the U-235/U-238 ratio was enhanced with a sample load of around 300 ng U. The resulting reproducibility for standards and soil samples was better than 0.2% at two standard deviations. Uranium isotopic compositions have been determined in soil samples and compared between Chernobyl nuclear accident site and the criticality accident, which occurred in Japan at the JCO site.
引用
收藏
页码:241 / 245
页数:5
相关论文
共 23 条
[1]   DETERMINATION OF NATURAL ACTINIDES AND PLUTONIUM IN MARINE PARTICULATE MATERIAL [J].
ANDERSON, RF ;
FLEER, AP .
ANALYTICAL CHEMISTRY, 1982, 54 (07) :1142-1147
[2]   ISOTOPIC DETERMINATION OF URANIUM IN PICOMOLE AND SUBPICOMOLE QUANTITIES [J].
CHEN, JH ;
WASSERBURG, GJ .
ANALYTICAL CHEMISTRY, 1981, 53 (13) :2060-2067
[3]   HIGH-PRECISION URANIUM, THORIUM AND RADIUM ISOTOPE RATIO MEASUREMENTS BY HIGH DYNAMIC-RANGE THERMAL IONIZATION MASS-SPECTROMETRY [J].
COHEN, AS ;
BELSHAW, NS ;
ONIONS, RK .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1992, 116 (01) :71-81
[4]   Rapid procedure for plutonium and uranium determination in soils using a borate fusion followed by ion-exchange and extraction chromatography [J].
Croudace, I ;
Warwick, P ;
Taylor, RN ;
Dee, S .
ANALYTICA CHIMICA ACTA, 1998, 371 (2-3) :217-225
[5]   DISTRIBUTION OF URANIUM IN HUMAN ORGANS OF AN URBAN INDIAN POPULATION AND ITS RELATIONSHIP WITH CLEARANCE HALF-LIVES [J].
DANG, HS ;
PULLAT, VR ;
SHARMA, RC .
HEALTH PHYSICS, 1995, 68 (03) :328-331
[6]  
Friedlander G., 1981, Nuclear and radiochemistry
[7]   Rapid separation of uranium and plutonium by extraction chromatography for determination by thermal ionisation mass spectrometry [J].
Goodall, P ;
Lythgoe, C .
ANALYST, 1999, 124 (03) :263-269
[8]  
HALL RJB, 1990, PLASMA SOURCE MASS S
[9]   CONCENTRATION AND SEPARATION OF ACTINIDES FROM URINE USING A SUPPORTED BIFUNCTIONAL ORGANOPHOSPHORUS EXTRACTANT [J].
HORWITZ, EP ;
DIETZ, ML ;
NELSON, DM ;
LAROSA, JJ ;
FAIRMAN, WD .
ANALYTICA CHIMICA ACTA, 1990, 238 (02) :263-271
[10]  
IVANOVICH M, 1993, URANIUM SERIES DISEQ