A vertical flow paper-microarray assay with isothermal DNA amplification for detection of Neisseria meningitidis

被引:34
作者
Rivas, Lourdes [1 ]
Reutersward, Philippa [1 ]
Rasti, Reza [2 ,3 ]
Herrmann, Bjorn [4 ]
Martensson, Andreas [5 ]
Alfven, Tobias [2 ,3 ]
Gantelius, Jesper [1 ,2 ]
Andersson-Svahn, Helene [1 ]
机构
[1] KTH Royal Inst Technol, Sci Life Lab, Div Prote & Nanobiotechnol, Stockholm, Sweden
[2] Karolinska Inst, Dept Publ Hlth Sci, Stockholm, Sweden
[3] South Gen Hosp, Sachs Children & Youth Hosp, Stockholm, Sweden
[4] Uppsala Univ, Dept Med Sci, Sect Clin Bacteriol, Uppsala, Sweden
[5] Uppsala Univ, IMCH, Dept Womens & Childrens Hlth, Uppsala, Sweden
基金
欧洲研究理事会; 瑞典研究理事会;
关键词
Microarray; Paper-based biosensor; Point-of-care; Gold nanoparticles; Neisseria meningitidis; RECOMBINASE POLYMERASE AMPLIFICATION; HAEMOPHILUS-INFLUENZAE; BACTERIAL-MENINGITIS; CLINICAL VALIDATION; NUCLEIC-ACIDS; TEST STRIPS; DESIGN; POINT; OPTIMIZATION; SENSITIVITY;
D O I
10.1016/j.talanta.2018.02.070
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Paper-based biosensors offer a promising technology to be used at the point of care, enabled by good performance, convenience and low-cost. In this article, we describe a colorimetric vertical-flow DNA microarray (DNAVFM) that takes advantage of the screening capability of DNA microarrays in a paper format together with isothermal amplification by means of Recombinase Polymerase Amplification (RPA). Different assay parameters such as hybridization buffer, flow rate, printing buffer and capture probe concentration were optimized. A limit of detection (LOD) of 4.4 nM was achieved as determined by tabletop scanning. The DNA-VFM was applied as a proof of concept for detection of Neisseria meningitidis, a primary cause of bacterial meningitis. The LOD was determined to be between 38 and 2.1 x 10(6) copies/VFMassay, depending on the choice of DNA capture probes. The presented approach provides multiplex capabilities of DNA microarrays in a paper-based format for future point-of-care applications.
引用
收藏
页码:192 / 200
页数:9
相关论文
共 63 条
  • [1] Multiplex quantitative PCR for detection of lower respiratory tract infection and meningitis caused by Streptococcus pneumoniae, Haemophilus influenzae and Neisseria meningitidis
    Abdeldaim, Guma M. K.
    Stralin, Kristoffer
    Korsgaard, Jens
    Blomberg, Jonas
    Welinder-Olsson, Christina
    Herrmann, Bjorn
    [J]. BMC MICROBIOLOGY, 2010, 10
  • [2] A universal nucleic acid sequence biosensor with nanomolar detection limits
    Baeumner, AJ
    Pretz, J
    Fang, S
    [J]. ANALYTICAL CHEMISTRY, 2004, 76 (04) : 888 - 894
  • [3] Evaluating mixtures of 14 hygroscopic additives to improve antibody microarray performance
    Bergeron, Sebastien
    Laforte, Veronique
    Lo, Pik-Shan
    Li, Huiyan
    Juncker, David
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (28) : 8451 - 8462
  • [4] Diagnostic accuracy of loop-mediated isothermal amplification as a near-patient test for meningococcal disease in children: an observational cohort study
    Bourke, Thomas W.
    McKenna, James P.
    Coyle, Peter V.
    Shields, Michael D.
    Fairley, Derek J.
    [J]. LANCET INFECTIOUS DISEASES, 2015, 15 (05) : 552 - 558
  • [5] Application of DNA Aptamers and Quantum Dots to Lateral Flow Test Strips for Detection of Foodborne Pathogens with Improved Sensitivity versus Colloidal Gold
    Bruno, John G.
    [J]. PATHOGENS, 2014, 3 (02) : 341 - 355
  • [6] Bumgarner Roger, 2013, Curr Protoc Mol Biol, VChapter 22, DOI 10.1002/0471142727.mb2201s101
  • [7] Lateral flow microarrays: a novel platform for rapid nucleic acid detection based on miniaturized lateral flow chromatography
    Carter, Darren J.
    Cary, R. Bruce
    [J]. NUCLEIC ACIDS RESEARCH, 2007, 35 (10)
  • [8] Point-of-Care Vertical Flow Allergen Microarray Assay: Proof of Concept
    Chinnasamy, Thiruppathiraja
    Segerink, Loes I.
    Nystrand, Mats
    Gantelius, Jesper
    Svahn, Helene Andersson
    [J]. CLINICAL CHEMISTRY, 2014, 60 (09) : 1209 - 1216
  • [9] A lateral flow paper microarray for rapid allergy point of care diagnostics
    Chinnasamy, Thiruppathiraja
    Segerink, Loes I.
    Nystrand, Mats
    Gantelius, Jesper
    Svahn, Helene Andersson
    [J]. ANALYST, 2014, 139 (10) : 2348 - 2354
  • [10] Sensitive biomolecule detection in lateral flow assay with a portable temperature-humidity control device
    Choi, Jane Ru
    Hu, Jie
    Feng, Shangsheng
    Abas, Wan Abu Bakar Wan
    Pingguan-Murphy, Belinda
    Xu, Feng
    [J]. BIOSENSORS & BIOELECTRONICS, 2016, 79 : 98 - 107