Prevention of post-focal thermal damage by formation of bubbles at the focus during high intensity focused ultrasound therapy

被引:30
作者
Zderic, Vesna [2 ]
Foley, Jessica [1 ]
Luo, Wenbo [1 ]
Vaezy, Shahram [1 ]
机构
[1] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[2] George Washington Univ, Dept Elect & Comp Engn, Washington, DC 20052 USA
基金
美国国家卫生研究院;
关键词
HIFU; focused ultrasound; cavitation; boiling; bubbles; thermal damage; reflection; scattering; tissue-air interface; safety considerations;
D O I
10.1118/1.2975149
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Safety concerns exist for potential thermal damage at tissue-air or tissue-bone interfaces located in the post-focal region during high intensity focused ultrasound (HIFU) treatments. We tested the feasibility of reducing thermal energy deposited at the post-focal tissue-air interfaces by producing bubbles (due to acoustic cavitation and/or boiling) at the HIFU focus. HIFU (in-situ intensities of 460-3500 W/cm(2), frequencies of 3.2-5.5 MHz) was applied for 30 s to produce lesions (in turkey breast in-vitro (n=37), and rabbit liver (n=4) and thigh muscle in-vivo (n=11)). Tissue temperature was measured at the tissue-air interface using a thermal (infrared) camera. Ultrasound imaging was used to detect bubbles at the HIFU focus, appearing as a hyperechoic region. In-vitro results showed that when no bubbles were present at the focus (at lower intensities of 460-850 W/cm(2)), the temperature at the interface increased continuously, up to 7.3 +/- 4.0 degrees C above the baseline by the end of treatment. When bubbles formed immediately after the start of HIFU treatment (at the high intensity of 3360 W/cm(2)), the temperature increased briefly for 3.5 s to 7.4 +/- 3.6 degrees C above the baseline temperature and then decreased to 4.0 +/- 1.4 degrees C above the baseline by the end of treatment. Similar results were obtained in in-vivo experiments with the temperature increases (above the baseline temperature) at the muscle-air and liver-air interfaces at the end of the high intensity treatment lower by 7.1 degrees C and 6.0 degrees C, respectively, as compared to the low intensity treatment. Thermal effects of HIFU at post-focal tissue-air interfaces, such as in bowels, could result in clinically significant increases in temperature. Bubble formation at the HIFU focus may provide a method for shielding the post-focal region from potential thermal damage. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:4292 / 4299
页数:8
相关论文
共 46 条
[1]  
[Anonymous], 1988, ULTRASOUND ITS CHEM
[2]   Use of overpressure to assess the role of bubbles in focused ultrasound lesion shape in vitro [J].
Bailey, MR ;
Couret, LN ;
Sapozhnikov, OA ;
Khokhlova, VA ;
Ter Haar, G ;
Vaezy, S ;
Shi, XG ;
Martin, R ;
Crum, LA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (05) :695-708
[3]   Mechanisms of phrenic nerve injury during radiofrequency ablation at the pulmonary vein orifice [J].
Bunch, TJ ;
Bruce, GK ;
Mahapatra, S ;
Johnson, SB ;
Miller, DV ;
Sarabanda, AV ;
Milton, MA ;
Packer, DL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (12) :1318-1325
[4]  
Christensen D., 1988, ULTRASONIC BIOINSTRU
[5]   Temperature rise recorded during lesion formation by high-intensity focused ultrasound [J].
Clarke, RL ;
terHaar, GR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (02) :299-306
[6]   GROWTH OF AIR BUBBLES IN TISSUE BY RECTIFIED DIFFUSION [J].
CRUM, LA ;
HANSEN, GM .
PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (03) :413-417
[7]   Experimental evaluation of lesion prediction modelling in the presence of cavitation bubbles: intended for high-intensity focused ultrasound prostate treatment [J].
Curiel, L ;
Chavrier, F ;
Gignoux, B ;
Pichardo, S ;
Chesnais, S ;
Chapelon, JY .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2004, 42 (01) :44-54
[8]   MRI monitoring of the effect of tissue interfaces in the penetration of high intensity focused ultrasound in kidney in vivo [J].
Damianou, C .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (09) :1209-1215
[9]   MRI-guided focused ultrasound surgery of uterine leiomyomas [J].
Fennessy, FM ;
Tempany, CM .
ACADEMIC RADIOLOGY, 2005, 12 (09) :1158-1166
[10]  
Harris GR, 2006, AIP CONF PROC, V829, P595, DOI 10.1063/1.2205543