We describe the implementation of optimal controllers for damping the motion of cantilevers used in magnetic resonance force microscopy. We demonstrate that optimal control is achievable and that torsional magnetic coupling provides an effective actuation method. Cantilever Brownian vibrational amplitude was reduced from 2 to 0.16 Angstrom and resonant quality was reduced from 2000 to 5. Applied control fields were sufficiently small that they would not affect magnetic resonance phenomena. (C) 1996 American Institute of Physics.
引用
收藏
页码:1959 / 1964
页数:6
相关论文
共 14 条
[1]   ANHARMONIC MODULATION FOR NOISE-REDUCTION IN MAGNETIC-RESONANCE FORCE MICROSCOPY [J].
BRULAND, KJ ;
KRZYSTEK, J ;
GARBINI, JL ;
SIDLES, JA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (04) :2853-2856
[2]   Optimal control of force microscope cantilevers .1. Controller design [J].
Garbini, JL ;
Bruland, KJ ;
Dougherty, WM ;
Sidles, JA .
JOURNAL OF APPLIED PHYSICS, 1996, 80 (04) :1951-1958
[3]  
HOEN S, 1994, 1994 SOL STAT SENS A, P209
[4]   CALIBRATION OF ATOMIC-FORCE MICROSCOPE TIPS [J].
HUTTER, JL ;
BECHHOEFER, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (07) :1868-1873
[5]   REGULATION OF A MICROCANTILEVER RESPONSE BY FORCE FEEDBACK [J].
MERTZ, J ;
MARTI, O ;
MLYNEK, J .
APPLIED PHYSICS LETTERS, 1993, 62 (19) :2344-2346
[6]   MECHANICAL DETECTION OF MAGNETIC-RESONANCE [J].
RUGAR, D ;
YANNONI, CS ;
SIDLES, JA .
NATURE, 1992, 360 (6404) :563-566
[7]   FORCE DETECTION OF NUCLEAR-MAGNETIC-RESONANCE [J].
RUGAR, D ;
ZUGER, O ;
HOEN, S ;
YANNONI, CS ;
VIETH, HM ;
KENDRICK, RD .
SCIENCE, 1994, 264 (5165) :1560-1563
[8]   IMPROVED FIBER-OPTIC INTERFEROMETER FOR ATOMIC FORCE MICROSCOPY [J].
RUGAR, D ;
MAMIN, HJ ;
GUETHNER, P .
APPLIED PHYSICS LETTERS, 1989, 55 (25) :2588-2590
[9]   FORCE MICROSCOPE USING A FIBER-OPTIC DISPLACEMENT SENSOR [J].
RUGAR, D ;
MAMIN, HJ ;
ERLANDSSON, R ;
STERN, JE ;
TERRIS, BD .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1988, 59 (11) :2337-2340
[10]   SIGNAL-TO-NOISE RATIOS IN INDUCTIVE AND MECHANICAL DETECTION OF MAGNETIC-RESONANCE [J].
SIDLES, JA ;
RUGAR, D .
PHYSICAL REVIEW LETTERS, 1993, 70 (22) :3506-3509