Computational fluid dynamics analysis to establish the design process of a centrifugal blood pump: Second report

被引:43
作者
Miyazoe, Y
Sawairi, T
Ito, K
Konishi, Y
Yamane, T
Nishida, M
Asztalos, B
Masuzawa, T
Tsukiya, T
Endo, S
Taenaka, Y
机构
[1] Nikkiso Co Ltd, Shizuoka Plant, Ctr Res & Dev, Haibara, Shizuoka 4210496, Japan
[2] AIST, Mech Engn Lab, Tsukuba, Ibaraki 305, Japan
[3] Ibaraki Univ, Ibaraki, Osaka, Japan
[4] Natl Cardiovasc Ctr, Osaka, Japan
关键词
centrifugal blood pump; computational fluid dynamics analysis; hemolysis test; flow visualization tests; shear stress; washout holes;
D O I
10.1046/j.1525-1594.1999.06418.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To establish an efficient design process for centrifugal blood pumps, the results of computational fluid dynamics (CFD) analysis were compared to the results of flow visualization tests and hemolysis tests, using the Nikkiso centrifugal blood pump. CFD analysis revealed that the radial gap greatly affected the shear stress in the outlet diffuser. The hemolysis study also indicated a similar tendency. To see the flow behind the impeller, we conducted a comparative study between models with and without washout holes using the CFD technique. CFD analysis indicated that flow and pressure distributions behind the impeller were different between both models, and a particle was observed to remain longer behind the impeller in the model without washout holes. In the future, CFD analysis could be a useful tool for developing blood pumps in comparison to flow visualization tests and hemolysis tests.
引用
收藏
页码:762 / 768
页数:7
相关论文
共 10 条
[1]   Development of design methods for a centrifugal blood pump with a fluid dynamic approach: Results in hemolysis tests [J].
Masuzawa, T ;
Tsukiya, T ;
Endo, S ;
Tatsumi, E ;
Taenaka, Y ;
Takano, H ;
Yamane, T ;
Nishida, M ;
Asztalos, B ;
Miyazoe, Y ;
Ito, K ;
Sawairi, T ;
Konishi, Y .
ARTIFICIAL ORGANS, 1999, 23 (08) :757-761
[2]   Computational fluid dynamic analyses to establish design process of centrifugal blood pumps [J].
Miyazoe, Y ;
Sawairi, T ;
Ito, K ;
Konishi, Y ;
Yamane, T ;
Nishida, M ;
Masuzawa, T ;
Takiura, K ;
Taenaka, Y .
ARTIFICIAL ORGANS, 1998, 22 (05) :381-385
[3]   Comparative hemolysis study of clinically available centrifugal pumps [J].
Naito, K ;
Suenaga, E ;
Cao, ZL ;
Suda, H ;
Ueno, T ;
Natsuaki, M ;
Itoh, T .
ARTIFICIAL ORGANS, 1996, 20 (06) :560-563
[4]   Flow visualization study to improve hemocompatibility of a centrifugal blood pump [J].
Nishida, M ;
Asztalos, B ;
Yamane, T ;
Masuzawa, T ;
Tsukiya, T ;
Endo, S ;
Taenaka, Y ;
Miyazoe, Y ;
Ito, K ;
Konishi, Y .
ARTIFICIAL ORGANS, 1999, 23 (08) :697-703
[5]   INITIAL CLINICAL-EXPERIENCE WITH THE BAYLOR-NIKKISO CENTRIFUGAL PUMP [J].
OHTSUBO, S ;
NAITO, K ;
MATSUURA, M ;
KAWAHITO, K ;
SHIMONO, T ;
MAKINOUCHI, K ;
TASAI, K ;
OHARA, Y ;
DAMM, G ;
GLUECK, J ;
RASKIN, S ;
TAKATANI, S ;
BENKOWSKI, R ;
SHORT, DH ;
SCHINEN, SA ;
NOON, GP ;
NOSE, Y .
ARTIFICIAL ORGANS, 1995, 19 (07) :769-773
[6]   CLINICAL-EXPERIENCE WITH NIKKISO CENTRIFUGAL PUMPS FOR EXTRACORPOREAL-CIRCULATION [J].
ONODA, K ;
KONDO, C ;
MIZUMOTO, T ;
KUSAGAWA, H ;
KATAYAMA, Y ;
HAYASHI, T ;
KOMADA, T ;
HIRANO, R ;
MIYAMURA, T ;
TANAKA, J ;
YASUDA, F ;
TANAKA, K ;
SHIMPO, H ;
YADA, I ;
YUASA, H ;
AIZAWA, T ;
KUSAGAWA, M .
ARTIFICIAL ORGANS, 1994, 18 (09) :706-710
[7]   CARDIOPULMONARY BYPASS WITH NIKKISO AND BIOMEDICUS CENTRIFUGAL PUMPS [J].
ORIME, Y ;
TAKATANI, S ;
SASAKI, T ;
AIZAWA, T ;
OHARA, Y ;
NAITO, K ;
GLUECK, J ;
NOON, GP ;
NOSE, Y ;
DEBAKEY, ME .
ARTIFICIAL ORGANS, 1994, 18 (01) :11-16
[8]   TOTAL ERYTHROCYTE DESTRUCTION TIME - THE NEW INDEX FOR THE HEMOLYTIC PERFORMANCE OF ROTARY BLOOD PUMPS [J].
SHIMONO, T ;
MAKINOUCHI, K ;
NOSE, Y .
ARTIFICIAL ORGANS, 1995, 19 (07) :571-575
[9]   Development of design methods of a centrifugal blood pump with in vitro tests, flow visualization, and computational fluid dynamics: Results in hemolysis tests [J].
Takiura, K ;
Masuzawa, T ;
Endo, S ;
Wakisaka, Y ;
Tatsumi, E ;
Taenaka, Y ;
Takano, H ;
Yamane, T ;
Nishida, M ;
Asztalos, B ;
Konishi, Y ;
Miyazoe, Y ;
Ito, K .
ARTIFICIAL ORGANS, 1998, 22 (05) :393-398
[10]   Flow visualization as a complementary tool to hemolysis testing in the development of centrifugal blood pumps [J].
Yamane, T ;
Asztalos, B ;
Nishida, M ;
Masuzawa, T ;
Takiura, K ;
Taenaka, Y ;
Konishi, Y ;
Miyazoe, Y ;
Ito, K .
ARTIFICIAL ORGANS, 1998, 22 (05) :375-380