Comparison of glucose monitoring between Freestyle Libre Pro and iPro2 in patients with diabetes mellitus

被引:24
作者
Kumagai, Ryo [1 ]
Muramatsu, Aiko [1 ]
Fujii, Masanao [1 ]
Katakura, Yukino [1 ]
Ito, Kei [1 ]
Fujie, Keiko [2 ]
Nakata, Yoshio [2 ]
Hashimoto, Koichi [2 ]
Yagyu, Hiroaki [1 ]
机构
[1] Mito Kyodo Gen Hosp, Dept Endocrinol & Metab, Tsukuba Univ Hosp, Mito Clin Educ & Training Ctr, Mito, Ibaraki, Japan
[2] Univ Tsukuba, Fac Med, Tsukuba, Ibaraki, Japan
关键词
Nocturnal hypoglycemia; Postprandial hyperglycemia; Self-management; BLOOD-GLUCOSE; FLASH; SYSTEMS;
D O I
10.1111/jdi.12970
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims/Introduction Flash and continuous glucose monitoring systems are becoming prevalent in clinical practice. We directly compared a flash glucose monitoring system (FreeStyle Libre Pro [FSL-Pro]) with a continuous glucose monitoring system (iPro2) in patients with diabetes mellitus. Materials and Methods Glucose concentrations were simultaneously measured using the FSL-Pro, iPro2 and self-monitoring blood glucose in 10 patients with diabetes mellitus, and agreement among them was assessed. Results Parkes error grid analysis showed that the 92.9 and 7.1% of glucose values measured using the FSL-Pro fell into areas A and B, respectively, and that 96.3, 2.8 and 0.9% of those determined using iPro2 fell into areas A, B and C, respectively. The median absolute relative differences compared with self-monitoring blood glucose were 8.1% (3.9-12.7%) and 5.0% (2.6-9.1%) for the FSL-Pro and iPro2, respectively. Analysis of 5,555 paired values showed a close correlation between FSL-Pro and iPro2 glucose values (rho = 0.96, P < 0.01). Notably, 65.3% of all glucose values were lower for the FSL-Pro than the iPro2. Median glucose values also decreased by 3.3% for the FSL-Pro compared with the iPro2 (177.0 [133.0-228.0] vs 183.0 [145.0-230.0] mg/dL, P < 0.01). The difference in glucose values between the two systems was more pronounced in hypoglycemia. The median absolute relative difference between FSL-Pro and iPro2 during hypoglycemia was much larger than that during euglycemia and hyperglycemia. Conclusions Both the FSL-Pro and iPro2 systems are clinically acceptable, but glucose values tended to be lower when measured using the FSL-Pro than the iPro2. Agreement was not close between these systems during hypoglycemia.
引用
收藏
页码:851 / 856
页数:6
相关论文
共 20 条
[1]   Evaluation of subcutaneous glucose monitoring systems under routine environmental conditions in patients with type 1 diabetes [J].
Aberer, Felix ;
Hajnsek, Martin ;
Rumpler, Markus ;
Zenz, Sabine ;
Baumann, Petra M. ;
Elsayed, Hesham ;
Puffing, Adelheid ;
Treiber, Gerlies ;
Pieber, Thomas R. ;
Sourij, Harald ;
Mader, Julia K. .
DIABETES OBESITY & METABOLISM, 2017, 19 (07) :1051-1055
[2]   The Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System [J].
Bailey, Timothy ;
Bode, Bruce W. ;
Christiansen, Mark P. ;
Klaff, Leslie J. ;
Alva, Shridhara .
DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 (11) :787-794
[3]  
Blevins Thomas C, 2010, J Diabetes Sci Technol, V4, P440
[4]   Limitations of conventional methods of self-monitoring of blood glucose - Lessons learned from 3 days of continuous glucose sensing in pediatric patients with type 1 diabetes [J].
Boland, E ;
Monsod, T ;
Delucia, M ;
Brandt, CA ;
Fernando, S ;
Tamborlane, WV .
DIABETES CARE, 2001, 24 (11) :1858-1862
[5]   Head-to-head comparison between flash and continuous glucose monitoring systems in outpatients with type 1 diabetes [J].
Bonora, B. ;
Maran, A. ;
Ciciliot, S. ;
Avogaro, A. ;
Fadini, G. P. .
JOURNAL OF ENDOCRINOLOGICAL INVESTIGATION, 2016, 39 (12) :1391-1399
[6]   FreeStyle Libre and Dexcom G4 Platinum sensors: Accuracy comparisons during two weeks of home use and use during experimentally induced glucose excursions [J].
Boscari, F. ;
Galasso, S. ;
Facchinetti, A. ;
Marescotti, M. C. ;
Vallone, V. ;
Amato, A. M. L. ;
Avogaro, A. ;
Bruttomesso, D. .
NUTRITION METABOLISM AND CARDIOVASCULAR DISEASES, 2018, 28 (02) :180-186
[7]   Real-Time Continuous Glucose Monitoring Significantly Reduces Severe Hypoglycemia in Hypoglycemia-Unaware Patients With Type 1 Diabetes [J].
Choudhary, Pratik ;
Ramasamy, Sharmin ;
Green, Louisa ;
Gallen, Geraldine ;
Pender, Siobhan ;
Brackenridge, Anna ;
Amiel, Stephanie A. ;
Pickup, John C. .
DIABETES CARE, 2013, 36 (12) :4160-4162
[8]   Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus [J].
Fokkert, M. J. ;
van Dijk, P. R. ;
Edens, M. A. ;
Abbes, S. ;
de Jong, D. ;
Slingerland, R. J. ;
Bilo, H. J. G. .
BMJ OPEN DIABETES RESEARCH & CARE, 2017, 5 (01)
[9]   Continuous glucose monitoring - Roadmap for 21st century diabetes therapy [J].
Klonoff, DC .
DIABETES CARE, 2005, 28 (05) :1231-1239
[10]  
McIntosh Brendan, 2010, Open Med, V4, pe102