A comparison of dual kV energy integrating and energy discriminating photon counting detectors for dual energy x-ray imaging

被引:2
|
作者
Wang, Adam S. [1 ]
Pelc, Norbert J. [1 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
来源
MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING | 2012年 / 8313卷
关键词
Dual energy; photon counting; pulse pileup; spectral tailing; material decomposition; CRLB; PULSE-HEIGHT SPECTRA; DOUBLE SUM PULSES; GENERALIZED METHOD; PILEUP; STATISTICS; TOMOGRAPHY; DISTORTION; QUALITY; CT;
D O I
10.1117/12.912030
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Dual energy imaging enables material decomposition and requires attenuation measurements with at least two different energies. Today's clinically available implementations use two separate exposures, at a low and high tube voltage (dual kV). Photon counting x-ray detectors (PCXDs) are an alternative technology that takes advantage of an x-ray source's broad spectrum by counting the number of transmitted photons at each energy from a single exposure. The richness of the information contained in these measurements can depend heavily on the detector's energy response, itself dependent on count rate. We compare the material decomposition precision of dual kV with energy integrating detectors to that of realistic PCXDs. We model the three primary effects that degrade PCXD performance: count rate limitations, energy resolution, and spectrum tailing. For example, the high flux rates required for clinical imaging pose a serious challenge for PCXDs. The Cramer-Rao Lower Bound is used to predict the best possible material decomposition variance as a function of these detector imperfections. For dual kV and photon counting, we determined the optimal kV, mAs, and filtration for a broad range of imaging tasks, subject to dose and tube power constraints. We found that a well-optimized dual kV protocol performs on par with the estimated performance of today's PCXDs. For dual kV protocols, it is helpful to increase the energy separation between the spectra by increasing the kV separation and adding filtration. For PCXDs, the detector's count rate capabilities must be increased and the spectrum tailing reduced for photon counting to become a competitive technology at the high intensities of clinical imaging.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] Rock porosity quantification by dual-energy X-ray computed microtomography
    Teles, A. P.
    Lima, I.
    Topes, R. T.
    MICRON, 2016, 83 : 72 - 78
  • [42] Analysis of artifact with X-ray CT using energy band by photon counting CdTe detector
    Matsumoto, G.
    Imura, Y.
    Morii, H.
    Miyake, A.
    Aoki, T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3) : 292 - 294
  • [43] Energy response calibration of photon-counting detectors using x-ray fluorescence: a feasibility study
    Cho, H-M
    Ding, H.
    Ziemer, B. P.
    Molloi, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (23) : 7211 - 7227
  • [44] Vision 20/20: Single photon counting x-ray detectors in medical imaging
    Taguchi, Katsuyuki
    Iwanczyk, Jan S.
    MEDICAL PHYSICS, 2013, 40 (10) : 100901
  • [45] Empirical optimization of energy bin weights for compressing measurements with realistic photon counting x-ray detectors
    Yang, Yirong
    Wang, Sen
    Pal, Debashish
    Yin, Zhye
    Pelc, Norbert J.
    Wang, Adam S.
    MEDICAL PHYSICS, 2024, 51 (01) : 224 - 238
  • [46] Spectral Signatures of X-ray Scatter Using Energy-Resolving Photon-Counting Detectors
    Lewis, Cale E.
    Das, Mini
    SENSORS, 2019, 19 (22)
  • [47] Photon Counting Detectors for X-Ray Imaging With Emphasis on CT
    Ballabriga, R.
    Alozy, J.
    Bandi, F. N.
    Campbell, M.
    Egidos, N.
    Fernandez-Tenllado, J. M.
    Heijne, E. H. M.
    Kremastiotis, I.
    Llopart, X.
    Madsen, B. J.
    Pennicard, D.
    Sriskaran, V.
    Tlustos, L.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (04) : 422 - 440
  • [48] Robust Material Classification on Dual-Energy X-Ray Imaging Devices
    Cigla, Cevahir
    Kucukates, Busra
    Yalcin, Ozan
    Ak, Duygu Selin
    Isikoglu, Sukrucan Taylan
    ANOMALY DETECTION AND IMAGING WITH X-RAYS (ADIX) VII, 2022, 12104
  • [49] Spectroscopic X-ray imaging with photon counting pixel detectors
    Tlustos, L.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 623 (02) : 823 - 828
  • [50] The role of x-ray Swank factor in energy-resolving photon-counting imaging
    Tanguay, Jesse
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL PHYSICS, 2010, 37 (12) : 6205 - 6211