Miniaturized Protein Microarray with Internal Calibration as Point-of-Care Device for Diagnosis of Neonatal Sepsis

被引:40
作者
Buchegger, Patricia [1 ]
Sauer, Ursula [1 ]
Toth-Szekely, Hedvig [1 ]
Preininger, Claudia [1 ]
机构
[1] AIT, Hlth & Environm Dept, A-3430 Tulln, Austria
关键词
protein microarray; streptavidin coated magnetic particles; miniaturization; point-of-care testing; neonatal sepsis; internal calibration; C-REACTIVE PROTEIN; PROCALCITONIN; IMMUNOASSAYS; CYTOKINES; MARKERS; CHIP;
D O I
10.3390/s120201494
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Neonatal sepsis is still a leading cause of death among newborns. Therefore a protein-microarray for point-of-care testing that simultaneously quantifies the sepsis associated serum proteins IL-6, IL-8, IL-10, TNF alpha, S-100, PCT, E-Selectin, CRP and Neopterin has been developed. The chip works with only a 4 mu L patient serum sample and hence minimizes excessive blood withdrawal from newborns. The 4 mu L patient samples are diluted with 36 mu L assay buffer and distributed to four slides for repetitive measurements. Streptavidin coated magnetic particles that act as distinct stirring detection components are added, not only to stir the sample, but also to detect antibody antigen binding events. We demonstrate that the test is complete within 2.5 h using a single step assay. S-100 conjugated to BSA is spotted in increasing concentrations to create an internal calibration. The presented low volume protein-chip fulfills the requirements of point-of-care testing for accurate and repeatable (CV < 14%) quantification of serum proteins for the diagnosis of neonatal sepsis.
引用
收藏
页码:1494 / 1508
页数:15
相关论文
共 28 条
[1]  
[Anonymous], 1980, ANAL CHEM, V52, P2242
[2]   Protein microarray for the analysis of human melanoma biomarkers [J].
Domnanich, Patrick ;
Sauer, Ursula ;
Pultar, Johanna ;
Preininger, Claudia .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 139 (01) :2-8
[3]   Increasing robustness and sensitivity of protein microarrays through microagitation and automation [J].
Hartmann, M ;
Toegl, A ;
Kirchner, R ;
Templin, MF ;
Joos, TO .
ANALYTICA CHIMICA ACTA, 2006, 564 (01) :66-73
[4]   Acceleration of incubation processes in DNA bio chips by magnetic particles [J].
Heer, Rudolf ;
Eggeling, Moritz ;
Schotter, Joerg ;
Noehammer, Christa ;
Pichler, Rudolf ;
Mansfeld, Markus ;
Brueckl, Hubert .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2007, 311 (01) :244-248
[5]   Neurological injury markers in children with septic shock [J].
Hsu, Angela A. ;
Fenton, Kimberly ;
Weinstein, Steven ;
Carpenter, Jessica ;
Dalton, Heidi ;
Bell, Michael J. .
PEDIATRIC CRITICAL CARE MEDICINE, 2008, 9 (03) :245-251
[6]   Point-of-Care Testing in Hospitals and Primary Care [J].
Junker, Ralf ;
Schlebusch, Harald ;
Luppa, Peter B. .
DEUTSCHES ARZTEBLATT INTERNATIONAL, 2010, 107 (33) :561-U14
[7]  
Konovalova E V, 2007, Mol Biol (Mosk), V41, P734
[8]   Optimal design of microarray immunoassays to compensate for kinetic limitations -: Theory and experiment [J].
Kusnezow, Wlad ;
Syagailo, Yana V. ;
Rueffer, Sven ;
Baudenstiel, Nina ;
Gauer, Christoph ;
Hoheisel, Joerg D. ;
Wild, David ;
Goychuk, Igor .
MOLECULAR & CELLULAR PROTEOMICS, 2006, 5 (09) :1681-1696
[9]  
Laborada G, 2003, AM J PERINAT, V20, P491, DOI 10.1055/s-2003-45382
[10]   Conventional and molecular methods for verification of results obtained with BacT/Alert nonvent blood culture bottles [J].
Marlowe, EM ;
Gibson, L ;
Hogan, J ;
Kaplan, S ;
Bruckner, DA .
JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (03) :1266-1269