Image Quality Comparison between Digital and Synthetic 2D Mammograms: A Qualitative and Quantitative Phantom Study

被引:2
作者
Barca, P. [1 ,2 ]
Lamastra, R. [1 ]
Caramella, D. [3 ]
Traino, A. C. [4 ]
Tucciariello, R. M. [1 ]
Fantacci, M. E. [1 ,2 ]
机构
[1] Univ Pisa, Dept Phys, Pisa, Italy
[2] Ist Nazl Fis Nucl, Pisa Sect, Pisa, Italy
[3] Univ Pisa, Dept Translat Res New Technol Med & Surg, Pisa, Italy
[4] Pisa Univ Hosp, Unit Med Phys, Azienda Osped Univ Pisana, Pisa, Italy
来源
BIOIMAGING: PROCEEDINGS OF THE 12TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES, VOL 2 | 2019年
关键词
Synthesized Mammography; Digital Mammography; Image Quality; Modulation Transfer Function; Noise Power Spectrum; Contrast to Noise Ratio; ACR Phantom; BREAST TOMOSYNTHESIS; DIAGNOSTIC-ACCURACY; DBT;
D O I
10.5220/0007472401290136
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The recent introduction of digital breast tomosynthesis (DBT) have lead to improvements in sensitivity and specificity of breast cancer detection, especially in cases of tumors developed in dense breasts. Since DBT provides tomographic slices of an entire tissue volume, it reduces the inherent tissue overlapping limitation of digital mammography (DM). In addition, DBT combined with DM has been proven to decrease recall and increase invasive cancer detection rates in breast cancer screening. However, the employment of DBT+DM implies a not negligible increment of patients absorbed dose. Therefore, Synthesized mammograms (SMs) generated from the DBT data have been recently introduced to eliminate the need of an additional DM. However, several studies showed differences between DM and SM images and some studies found contrasting results in terms of image quality when DM and SM images were compared. In our phantom study, we objectively compare image quality of SM and DM images in terms of noise, spatial resolution and contrast properties. Additionally, a qualitative analysis of the ACR mammographic phantom was performed in both modalities to assess the detectability of different features. SM images were characterized by different texture with respect to DM images, showing lower overall performances in terms of contrast-to-noise ratio and modulation transfer function. However, the goal of SM images is to provide a useful two-dimensional guide complementary to the DBT dataset and the performances in terms of high-contrast features detectability were satisfactory in comparison to those obtained in DM.
引用
收藏
页码:129 / 136
页数:8
相关论文
共 28 条
[1]   Comparison Between Digital and Synthetic 2D Mammograms in Breast Density Interpretation [J].
Alshafeiy, Taghreed I. ;
Wadih, Antoine ;
Nicholson, Brandi T. ;
Rochman, Carrie M. ;
Peppard, Heather R. ;
Patrie, James T. ;
Harvey, Jennifer A. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2017, 209 (01) :W36-W41
[2]  
[Anonymous], 2011, IAEA Human Health Series.
[3]   A comparative study of physical image quality in digital and synthetic mammography from commercially available mammography systems [J].
Baldelli, Paola ;
Bertolini, Marco ;
Contillo, Adriano ;
Della Gala, Giuseppe ;
Golinelli, Paola ;
Pagan, Laura ;
Rivetti, Stefano ;
Taibi, Angelo .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (16)
[4]  
CEP, 2006, FULL FIELD DIG MAMM
[5]   Synthesized Mammography: Clinical Evidence, Appearance, and Implementation [J].
Durand, Melissa A. .
DIAGNOSTICS, 2018, 8 (02)
[6]  
EFOMP, 2015, MAMM PROT QUAL CONTR
[7]  
EUREF, 2016, PROT QUAL CONTR PHYS
[8]  
Fluke Biomedical, 2005, MAMM ACCR PHANT OP M
[10]   Assessing task performance in FFDM, DBT, and synthetic mammography using uniform and anthropomorphic physical phantoms [J].
Ikejimba, Lynda C. ;
Glick, Stephen J. ;
Choudhury, Kingshuk Roy ;
Samei, Ehsan ;
Lo, Joseph Y. .
MEDICAL PHYSICS, 2016, 43 (10) :5593-5602