Comparison of Optical Low-Coherence Reflectometry and Swept-Source OCT-Based Biometry Devices in Dense Cataracts

被引:12
作者
Vasavada, Shail A. [1 ]
Patel, Priyanka [1 ]
Vaishnav, Vineet R. [1 ]
Ashena, Zahra [2 ]
Srivastava, Samaresh [1 ]
Vasavada, Vaishali [1 ]
Nanavaty, Mayank A. [2 ,3 ]
机构
[1] Iladevi Cataract & IOL Res Ctr, Ahmadabad, Gujarat, India
[2] Brighton & Sussex Univ Hosp NHS Trust, Sussex Eye Hosp, Eastern Rd, Brighton BN2 5BF, E Sussex, England
[3] Univ Sussex, Brighton & Sussex Med Sch, Brighton, E Sussex, England
关键词
CENTRAL CORNEAL THICKNESS; OCULAR BIOMETRY; AXIAL LENGTH; TOMOGRAPHY; REPEATABILITY; AGREEMENT; REPRODUCIBILITY; INTERFEROMETRY; POWER;
D O I
10.3928/1081597X-20200612-03
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To investigate agreement between biometric measurements obtained from the Lenstar LS 900 (Haag-Streit AG) based on optical low-coherence reflectometry (OLCR) and the Tomey OA-2000 (Tomey Corporation) based on swept-source optical coherence tomography (SS-OCT) in dense cataracts. METHODS: In this prospective observational study, axial. length, keratometry, anterior chamber depth (ACD), lens thickness, and central corneal thickness (CCT) were measured in 124 eyes of 76 patients. Intraocular lens prediction errors and absolute prediction errors were assessed based on postoperative manifest refraction. Analyses used were paired t test, Pearson correlation coefficient (r), and Bland-Altman plots. RESULTS: Failure in axial length measurements was noted in 28 eyes (22.58%) with OCLR compared to 2 eyes (1.6%) with SS-OCT. Although not statistically significant, axial length measurements were clinically higher with SS-OCT (23.78 +/- 1.76 mm) compared to OCLR (23.72 +/- 1.58 mm) (P = .81) (r = 0.23; P = .01). There was a significant trend toward myopic intraocular lens prediction with 55-OCT (-0.09 +/- 0.48) compared to OLCR (0.09 +/- 0.41, P < .01) (r = 0.25, P = .01). There was good agreement for keratometric and ACD values between the two devices. SS-OCT gave significantly higher values of tens thickness compared to OCLR (4.44 +/- 0.44 vs 4.18 +/- 0.48 mm, P < .01) (r = 0.39, P < .01). CCT measured with OCLR was significantly higher than SS-OCT (525.64 +/- 27.0 vs 513.21 +/- 29.24 mu m; P < .01) (r = 0.98, P < .01). CONCLUSIONS: One-fifth of the patients with dense cataracts failed axial length measurement on OCLR. SS-OCT gives clinically higher axial length measurements leading to more myopic intraocular lens prediction errors postoperatively. OCLR provides higher pachymetry and lower lens thickness values compared to the SS-OCT device.
引用
收藏
页码:557 / 564
页数:8
相关论文
共 29 条
  • [1] Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer
    Arriola-Villalobos, P.
    Almendral-Gomez, J.
    Garzon, N.
    Ruiz-Medrano, J.
    Fernandez-Perez, C.
    Martinez-de-la-Casa, J. M.
    Diaz-Valle, D.
    [J]. EYE, 2017, 31 (03) : 437 - 442
  • [2] Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter
    Bayhan, Hasan Ali
    Bayhan, Seray Aslan
    Can, Izzet
    [J]. INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2014, 7 (02) : 302 - 308
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] A new optical low coherence reflectometry device for ocular biometry in cataract patients
    Buckhurst, P. J.
    Wolffsohn, J. S.
    Shah, S.
    Naroo, S. A.
    Davies, L. N.
    Berrow, E. J.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2009, 93 (07) : 949 - 953
  • [5] Cho Young Joo, 2018, Korean J Ophthalmol, V32, P257, DOI 10.3341/kjo.2017.0091
  • [6] Evaluation of the Lenstar LS 900 non-contact biometer
    Cruysberg, L. P. J.
    Doors, M.
    Verbakel, F.
    Berendschot, T. T. J. M.
    De Brabander, J.
    Nuijts, R. M. M. A.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (01) : 106 - 110
  • [7] Emery J.M., 1979, Phacoemulsification and aspiration ofcataracts
  • [8] Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry
    Gao, Rongrong
    Chen, Hao
    Savini, Giacomo
    Miao, Yaxin
    Wang, Xiaorui
    Yang, Jing
    Zhao, Weiqi
    Wang, Qinmei
    Huang, Jinhai
    [J]. SCIENTIFIC REPORTS, 2017, 7
  • [9] Comparison of 3 biometry devices in cataract patients
    Goebels, Susanne
    Pattmoeller, Max
    Eppig, Timo
    Cayless, Alan
    Seitz, Berthold
    Langenbucher, Achim
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (11) : 2387 - 2393
  • [10] Reproducibility of a Long-Range Swept-Source Optical Coherence Tomography Ocular Biometry System and Comparison with Clinical Biometers
    Grulkowski, Ireneusz
    Liu, Jonathan J.
    Zhang, Jason Y.
    Potsaid, Benjamin
    Jayaraman, Vijaysekhar
    Cable, Alex E.
    Duker, Jay S.
    Fujimoto, James G.
    [J]. OPHTHALMOLOGY, 2013, 120 (11) : 2184 - 2190