Selectivity enhancement for high-speed GC analysis of volatile organic compounds with portable instruments designed for vacuum-outlet and atmospheric-pressure inlet operation using air as the carrier gas

被引:28
作者
Whiting, J [1 ]
Sacks, R [1 ]
机构
[1] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
关键词
D O I
10.1021/ac010879g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A tandem ensemble of two 4.5-m-long x 0.25-mm-i.d. capillary columns with the first using a 0.50-mum film of nonpolar dimethyl polysiloxane and the second using a 0.25-mum film of polar trifluoropropylmethyl polysiloxane is operated with atmospheric pressure air as the carrier gas and an outlet pressure of 50.5 kPa established using a small vacuum pump. A thicker stationary-phase film is used in the first column to increase retention for very volatile compounds. This significantly increases the resolution of these compounds. The thicker film in the first (nonpolar) column decreases the polarity of the tandem column ensemble and, thus, changes its selectivity. A low-dead-volume valve, connected between the column junction point and a source of atmospheric pressure air, is used to obtain pulsed modulation of the carrier gas flow through the column ensemble. When the valve is open, the ensemble inlet pressure and the junction-point pressure are nearly the same, and carrier gas flow nearly stops in the first column, and flow in the second column increases. Enhanced resolution of a component pair that is separated by the first column but coelutes from the column ensemble can be obtained if the valve is opened for a few seconds after one of the components has crossed the junction and is in the second column, but the other component is still in the first column. A sequence of appropriately timed pulses is used to obtain enhanced resolution of several pairs of components that coelute from the column ensemble. These methods enabled the complete separation of an 18-component vapor mixture of common solvents in air in 3.5 min.
引用
收藏
页码:246 / 252
页数:7
相关论文
共 25 条
[1]  
ACKARD M, 1994, ANAL CHEM, V66, P3036
[2]   Programmable control of column selectivity for temperature-programmed GC [J].
Coutant, C ;
Sacks, R .
ANALYTICAL CHEMISTRY, 2000, 72 (21) :5450-5458
[3]   INCREASED SPEED OF ANALYSIS IN DIRECTLY COUPLED GAS CHROMATOGRAPHY-MASS SPECTROMETRY SYSTEMS - CAPILLARY COLUMNS AT SUB-ATMOSPHERIC OUTLET PRESSURES [J].
CRAMERS, CA ;
SCHERPENZEEL, GJ ;
LECLERCQ, PA .
JOURNAL OF CHROMATOGRAPHY, 1981, 203 (JAN) :207-216
[4]   GAS-CHROMATOGRAPHIC COLUMNS WITH ADJUSTABLE SEPARATION CHARACTERISTICS [J].
DEANS, DR ;
SCOTT, I .
ANALYTICAL CHEMISTRY, 1973, 45 (07) :1137-1141
[5]  
FRYE GC, 1994, EL SOC M MIAM BEACH, P1046
[6]   INFLUENCE OF INSTRUMENTAL CONTRIBUTIONS ON APPARENT COLUMN EFFICIENCY IN HIGH-SPEED GAS-CHROMATOGRAPHY [J].
GASPAR, G ;
ANNINO, R ;
VIDALMADJAR, C ;
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1978, 50 (11) :1512-1518
[7]   Column performance and stability for high-speed vacuum-outlet GC of volatile organic compounds using atmospheric pressure air as carrier gas [J].
Grall, AJ ;
Sacks, RD .
ANALYTICAL CHEMISTRY, 1999, 71 (22) :5199-5205
[8]   High speed analysis of complex indoor VOC mixtures by vacuum-outlet GC with air carrier gas and programmable retention [J].
Grall, AJ ;
Zellers, ET ;
Sacks, RD .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2001, 35 (01) :163-169
[9]   Pressure-tunable column selectivity for high speed vacuum-outlet GC [J].
Grall, AJ ;
Sacks, RD .
ANALYTICAL CHEMISTRY, 2000, 72 (11) :2507-2513
[10]   Review and summary of a silicon micromachined gas chromatography system [J].
Kolesar, ES ;
Reston, RR .
IEEE TRANSACTIONS ON COMPONENTS PACKAGING AND MANUFACTURING TECHNOLOGY PART B-ADVANCED PACKAGING, 1998, 21 (04) :324-328