Radiation transmission data for radionuclides and materials relevant to brachytherapy facility shielding

被引:25
作者
Papagiannis, P. [1 ]
Baltas, D. [2 ]
Granero, D. [3 ]
Perez-Calatayud, J. [3 ]
Gimeno, J. [4 ]
Ballester, F. [5 ,6 ]
Venselaar, J. L. M. [7 ]
机构
[1] Univ Athens, Sch Med, Med Phys Lab, GR-11527 Athens, Greece
[2] Klinikum Offenbach, Strahlenklin, Dept Med Phys & Engn, D-63069 Offenbach, Germany
[3] La Fe Univ Hosp, Dept Radiotherapy, E-46009 Valencia, Spain
[4] Fdn Inst Valenciano, FIVO, E-46009 Valencia, Spain
[5] Univ Valencia, Dept Atom Mol & Nucl Phys, E-46100 Burjassot, Spain
[6] CSIC, IFIC, E-46100 Burjassot, Spain
[7] Dr B Verbeeten Inst, Dept Med Phys, NL-5000 LA Tilburg, Netherlands
关键词
brachytherapy; radiation protection; shielding; HVL; TVL; Monte Carlo;
D O I
10.1118/1.2986153
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To address the limited availability of radiation shielding data for brachytherapy as well as some disparity in existing data, Monte Carlo simulation was used to generate radiation transmission data for Co-60, Cs-137, Au-198, Ir-192, Yb-169, Tm-170, Cs-131, I-125, and Pd-103 photons through concrete, stainless steel, lead, as well as lead glass and baryte concrete. Results accounting for the oblique incidence of radiation to the barrier, spectral variation with barrier thickness, and broad beam conditions in a realistic geometry are compared to corresponding data in the literature in terms of the half value layer (HVL) and tenth value layer (TVL) indices. It is also shown that radiation shielding calculations using HVL or TVL values could overestimate or underestimate the barrier thickness required to achieve a certain reduction in radiation transmission. This questions the use of HVL or TVL indices instead of the actual transmission data. Therefore, a three-parameter model is fitted to results of this work to facilitate accurate and simple radiation shielding calculations. (C) 2008 American Association of Physicists in Medicine. [DOI: 10.1118/1.2986153]
引用
收藏
页码:4898 / 4906
页数:9
相关论文
共 20 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]  
[Anonymous], 1997, UCRLLR50400 LAWR LIV
[3]  
[Anonymous], 2005, 151 NCRP
[4]   DIAGNOSTIC-X-RAY SHIELDING DESIGN BASED ON AN EMPIRICAL-MODEL OF PHOTON ATTENUATION [J].
ARCHER, BR ;
THORNBY, JI ;
BUSHONG, SC .
HEALTH PHYSICS, 1983, 44 (05) :507-517
[5]  
BOUTROUXJAFFE F, 1997, PRACTICAL MANUAL BRA, P3
[6]   ROOT - An object oriented data analysis framework [J].
Brun, R ;
Rademakers, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 389 (1-2) :81-86
[7]  
Delacroix D, 2002, RADIAT PROT DOSIM, V98, P9, DOI 10.1093/oxfordjournals.rpd.a006705
[8]   Broad-beam transmission data for new brachytherapy sources, Tm-170 and Yb-169 [J].
Granero, D ;
Pérez-Calatayud, J ;
Ballester, F ;
Bos, AJJ ;
Venselaar, J .
RADIATION PROTECTION DOSIMETRY, 2006, 118 (01) :11-15
[9]  
HENDRICKS S, 2002, 02408 LA CP LOS AL N
[10]  
HUBBELL JH, 2008, 2004 TABLES XRAY MAS