Comparison of Visual versus Microscopic Methods to Detect Blood Splatter from an Intravascular Catheter with Engineered Sharps Injury Protection

被引:5
作者
Ansari, Aiysha [1 ]
Ramaiah, Padmaja [2 ,3 ]
Collazo, Lillian [2 ,4 ]
Salihu, Hamisu M. [1 ,5 ]
Haiduven, Donna [1 ,2 ,3 ]
机构
[1] Univ S Florida, Dept Environm & Occupat Hlth, Coll Publ Hlth, Tampa, FL 33612 USA
[2] Univ S Florida, Dept Global Hlth, Coll Publ Hlth, Tampa, FL 33612 USA
[3] Tampa Vet Adm Res Ctr Excellence, Tampa, FL USA
[4] James A Haley Vet Adm Pathol & Lab Med Serv, Dept Microbiol, Tampa, FL USA
[5] Univ S Florida, Dept Epidemiol & Biostat, Coll Publ Hlth, Tampa, FL 33612 USA
关键词
NEEDLESTICK INJURY; PHLEBOTOMY; INFECTION; EXPOSURE; EFFICACY; HIV;
D O I
10.1086/673462
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
OBJECTIVE. To determine whether retractable intravenous devices produced blood splatter and whether blood splatter frequency differed between visual and microscopy detection methods. METHODS. In this laboratory-based experiment, 105 venipunctures were performed in a simulated brachial vein containing mock venous blood. The retraction mechanism was activated in a testing chamber with precut fabric filters, placed at 3 different locations, to capture blood splatter. Differences in filter mass, visual inspection, and microscopic analysis for presence of blood on filters were the units of analysis. Descriptive statistics, paired Student t tests, and k statistics were used for data analysis. RESULTS. Blood splatter was detected visually and microscopically as follows: filter A, 70% and 71%, respectively; filter B, 12% and 9%, respectively; and filter C, 13% and 10%, respectively. A statistically significant difference was observed in the mean mass of filter A between before and after activation when confirmed by the naked eye (Pp. 014) and microscopically (Pp. 0092). Substantial agreement between methods was observed for filter A (kappa = 0.78 [95% confidence interval, 0.64-0.92]), filter B (kp0.73 [95% confidence interval, 0.51-0.95]), and filter C (kappa = 0.75 [95% confidence interval, 0.55-0.96]). However, blood was detected by microscopy and not by the naked eye in 7 instances (7%). CONCLUSIONS. Our findings demonstrate that splatter, which can potentially expose healthcare workers (HCWs) to bloodborne pathogens, is associated with the activation of intravascular catheters with retraction mechanisms. HCWs may not detect this splatter when it occurs and may not report a splash to mucous membranes or nonintact skin. The need to wear personal protective equipment when using such devices is reinforced.
引用
收藏
页码:1174 / 1180
页数:7
相关论文
共 19 条
[1]  
Anonymous, 2005, Morbidity and Mortality Weekly Report, V54, P1
[2]   Prevention of needle-stick injury - Efficacy of a safeguarded intravenous cannula [J].
Asai, T ;
Matsumoto, S ;
Matsumoto, H ;
Yamamoto, K ;
Shingu, K .
ANAESTHESIA, 1999, 54 (03) :258-261
[3]   Efficacy of catheter needles with safeguard mechanisms [J].
Asai, T ;
Hidaka, I ;
Kawashima, A ;
Miki, T ;
Inada, K ;
Kawachi, S .
ANAESTHESIA, 2002, 57 (06) :572-577
[4]  
American Medical Association, 2001, MMWR Recomm Rep, V50, P1
[5]  
Foley M., 2005, Needlestick safety and prevention: Independent study module
[6]  
Ford Joanna, 2011, Nurs Stand, V26, P42
[7]  
Ford Joanna L, 2008, Nurs Times, V104, P42
[8]   A pilot study to measure the compressive and tensile forces required to use retractable intramuscular safety syringes [J].
Haiduven, Donna ;
Applegarth, Shawn ;
DiSalvo, Heidi ;
Mangipudy, Sitha ;
Konopack, Jason ;
Fisher, June .
AMERICAN JOURNAL OF INFECTION CONTROL, 2006, 34 (10) :661-668
[9]   An experimental method for detecting blood splatter from retractable phlebotomy and intravascular devices [J].
Haiduven, Donna ;
Applegarth, Shawn ;
Shroff, Miloni .
AMERICAN JOURNAL OF INFECTION CONTROL, 2009, 37 (02) :127-130
[10]   Contribution of a Winged Phlebotomy Device Design to Blood Splatter [J].
Haiduven, Donna J. ;
McGuire-Wolfe, Christine ;
Applegarth, Shawn P. .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2012, 33 (11) :1069-1076