Esophageal stent migration: Testing few hypothesis with a simplified mathematical model

被引:17
作者
Garbey, Marc [1 ]
Salmon, Remi [1 ]
Fikfak, Vid [2 ]
Clerc, Claude O. [3 ]
机构
[1] Houston Methodist Res Inst, Ctr Computat Surg, Houston, TX 77030 USA
[2] Houston Methodist, Dept Surg, Houston, TX USA
[3] Boston Sci Corp, Marlborough, MA USA
基金
美国国家科学基金会;
关键词
Esophageal stent; Stent migration; Computational model; Mass-spring model; Kelvin-Voigt material; Viscoelastic material; Esophageal stent design; LONGITUDINAL MUSCLE; MECHANICAL-PROPERTIES; PLACEMENT; DYSPHAGIA; ENDOPROSTHETICS; METHODOLOGY; PERISTALSIS; EXPERIENCES; STRICTURES; MOVEMENT;
D O I
10.1016/j.compbiomed.2016.10.024
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Esophageal stent placement has significantly improved the quality of life in patients with malignant as well as benign esophageal obstructing lesions. Despite its early success and rapid adoption, stent migration still occurs in as many as 30% of cases especially with fully covered stents. To date, few models of interaction between the stent and the esophageal wall have been published and these have only focused on the deployment of the stent or the static mechanical stress distribution of the stent material. To elucidate the mechanism behind esophageal stent migration we developed a simplified radially symmetric computational model of esophageal peristalsis and the stent. A thorough review of the literature on esophageal peristalsis was performed and pertinent data were implemented into the model. Similarly, mechanical properties of an existing esophageal stent were used for the stent model. A sensitivity analysis of the parameters of the model enabled identification of the key elements of stent design that influence the degree of stent migration including flares design, stent length as well as longitudinal and radial stiffness. A comparison of the model to the migration rate reported in clinical studies for various types of fully covered stents further verified our model, which can significantly contribute to the development of a more stable esophageal stent with lower rates of migration.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 47 条
[1]   Carotid artery stenting simulation: From patient-specific images to finite element analysis [J].
Auricchio, F. ;
Conti, M. ;
De Beule, M. ;
De Santis, G. ;
Verhegghe, B. .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (03) :281-289
[2]   Deployment of a self-expanding stent inside an artery: A finite element analysis [J].
Azaouzi, M. ;
Makradi, A. ;
Belouettar, S. .
MATERIALS & DESIGN, 2012, 41 :410-420
[3]  
Berg M. W., 2015, DIS ESOPHAGUS
[4]  
Chauviere A., 2010, CELL MECH SINGLE SCA, P221
[5]   MOVEMENT OF FELINE ESOPHAGUS ASSOCIATED WITH RESPIRATION AND PERISTALSIS - EVALUATION USING TANTALUM MARKERS [J].
DODDS, WJ ;
STEWART, ET ;
HODGES, D ;
ZBORALSKE, FF .
JOURNAL OF CLINICAL INVESTIGATION, 1973, 52 (01) :1-13
[6]   Efficacy and safety of a new fully covered self-expandable non-foreshortening metal esophageal stent [J].
Dua, Kulwinder S. ;
Latif, Sahibzada U. ;
Yang, Juliana F. ;
Fang, Tom C. ;
Khan, Abdul ;
Oh, Young .
GASTROINTESTINAL ENDOSCOPY, 2014, 80 (04) :577-585
[7]   SHORTENING OF THE ESOPHAGUS IN RESPONSE TO SWALLOWING [J].
EDMUNDOWICZ, SA ;
CLOUSE, RE .
AMERICAN JOURNAL OF PHYSIOLOGY, 1991, 260 (03) :G512-G516
[8]   Development of a prediction model of adverse events after stent placement for esophageal cancer [J].
Fuccio, Lorenzo ;
Scagliarini, Michele ;
Frazzoni, Leonardo ;
Battaglia, Giorgio .
GASTROINTESTINAL ENDOSCOPY, 2016, 83 (04) :746-752
[9]   Axial force measurement for esophageal function testing [J].
Gravesen, Flemming H. ;
Funch-Jensen, Peter ;
Gregersen, Hans ;
Drewes, Asbjorn Mohr .
WORLD JOURNAL OF GASTROENTEROLOGY, 2009, 15 (02) :139-143
[10]   Endoprosthetics for malignant esophageal disease [J].
Hirdes, Meike M. C. ;
Siersema, Peter .
TECHNIQUES IN GASTROINTESTINAL ENDOSCOPY, 2014, 16 (02) :64-70