Common Ground Method of Current Injection in Electrical Impedance Tomography

被引:0
作者
Bera, Tushar Kanti [1 ]
Nagaraju, J. [1 ]
机构
[1] Indian Inst Sci Bangalore, Dept Instrumentat & Appl Phys, Bangalore 560012, Karnataka, India
来源
GLOBAL TRENDS IN INFORMATION SYSTEMS AND SOFTWARE APPLICATIONS, PT 2 | 2012年 / 270卷
关键词
electrical impedance tomography; common ground current injection; surface electrodes; practical phantom; inhomogeneity; image reconstruction; resistivity image; EIDORS; RECONSTRUCTION ALGORITHMS; PHANTOM; ERRORS;
D O I
暂无
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Surface electrodes are essentially required to be switched for boundary data collection in electrical impedance tomography (Ell). Parallel digital data bits are required to operate the multiplexers used, generally, for electrode switching in ELT. More the electrodes in an EIT system more the digital data bits are needed. For a sixteen electrode system. 16 parallel digital data bits are required to operate the multiplexers in opposite or neighbouring current injection method. In this paper a common ground current injection is proposed for EIT and the resistivity imaging is studied. Common ground method needs only two analog multiplexers each of which need only 4 digital data bits and hence only 8 digital bits are required to switch the 16 surface electrodes. Results show that the USB based data acquisition system sequentially generate digital data required for multiplexers operating in common ground current injection method. The profile of the boundary data collected from practical phantom show that the multiplexers are operating in the required sequence in common ground current injection protocol. The voltage peaks obtained for all the inhomogeneity configurations are found at the accurate positions in the boundary data matrix which proved the sequential operation of multiplexers. Resistivity images reconstructed from the boundary data collected from the practical phantom with different configurations also show that the entire digital data generation module is functioning properly. Reconstructed images and their image parameters proved that the boundary data are successfully acquired by the DAQ system which in turn indicates a sequential and proper operation of multiplexers.
引用
收藏
页码:574 / 587
页数:14
相关论文
共 47 条
  • [1] [Anonymous], 17 WORLD C NOND TEST
  • [2] [Anonymous], 1990, A HILGER SERIES BIOM
  • [3] [Anonymous], P IEEE INT C SYST ME
  • [4] [Anonymous], 2009, 10 INT C BIOM APPL E
  • [5] Electrical impedance tomography of human brain function using reconstruction algorithms based on the finite element method
    Bagshaw, AP
    Liston, AD
    Bayford, RH
    Tizzard, A
    Gibson, AP
    Tidswell, AT
    Sparkes, MK
    Dehghani, H
    Binnie, CD
    Holder, DS
    [J]. NEUROIMAGE, 2003, 20 (02) : 752 - 764
  • [6] Bera Tushar Kanti, 2009, 2009 IEEE Instrumentation and Measurement Technology Conference (I2MTC), P511, DOI 10.1109/IMTC.2009.5168503
  • [7] Bera T K., 2011, Journal of Electrical Bioimpedance, V2, P48, DOI [10.5617/jeb.174, DOI 10.5617/JEB.174]
  • [8] Bera T.K., 2011, Journal_of_Electrical_Bioimpedance, V2, P33, DOI DOI 10.5617/JEB.158
  • [9] Bera TK., 2009, Sens. Transducers, V104, P33
  • [10] A chicken tissue phantom for studying an electrical impedance tomography (EIT) system suitable for clinical imaging
    Bera T.K.
    Nagaraju J.
    [J]. Sensing and Imaging: An International Journal, 2011, 12 (3-4) : 95 - 116