An in vivo evaluation of the Auto Apical Reverse function of the Root ZX II

被引:10
作者
Fadel, G. [1 ]
Piasecki, L. [2 ]
Westphalen, V. P. D. [1 ]
Silva Neto, U. X. [1 ]
Fariniuk, L. F. [1 ]
Carneiro, E. [1 ]
机构
[1] Pontificia Univ Catolica Parana, Dept Endodont, Curitiba, Parana, Brazil
[2] Univ Paranaense, Cascavel, Brazil
关键词
apical limit; electronic apex locator; root canal preparation; ROTARY CANAL INSTRUMENTATION; ELECTRONIC APEX LOCATORS; PROTAPER INSTRUMENTS; LOCATING HANDPIECE; ACCURACY; LENGTH; EXTENT; TEETH; OBTURATION; FORAMEN;
D O I
10.1111/j.1365-2591.2012.02054.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Fadel G, Piasecki L, Westphalen VPD, Silva Neto UX, Fariniuk LF, Carneiro E. An in vivo evaluation of the Auto Apical Reverse function of the Root ZX II. International Endodontic Journal,similar to 45, 950954, 2012. Abstract Aim To evaluate in vivo the accuracy of the Root ZX II (J. Morita) apex locator in controlling the apical extent of rotary instrumentation when using the Auto Apical Reverse (AAR) set at the levels 0.5, 1.0 and 1.5. Methodology Thirty single-rooted premolar teeth scheduled for extraction were divided into three groups (n = 10), according to the AAR setting 0.5, 1.0 and 1.5. The root canals were prepared using ProTaper (Dentsply Maillefer). After rotary instrumentation, the last file used (F3) was manually introduced into the extent of the root canal preparation and fixed before tooth extraction. The apical third of the root was dissected until exposure of the file. The distance from the file tip to the major apical foramen was obtained. Results Measurements within the range -1.0 to 0.0 mm were obtained in 30% of the teeth with AAR 0.5, 50% with AAR 1.0 and in 0% with AAR 1.5. The proportions test revealed a significant difference between the AAR settings 1.0 and 1.5 (P = 0.0188). Overinstrumentation occurred in 70% of the teeth with AAR 0.5 and in 40% with AAR 1.0. The measurements short of the acceptable range occurred in 10% of the teeth with setting AAR 1.0 and in 100% of the cases with AAR 1.5. A significant difference was found when comparing the percentage of teeth in which the file tip was short and beyond the established range between groups, except when comparing AAR 0.5 and AAR 1.0. Conclusion The AAR function of the Root ZX II was not an accurate method for controlling the apical extent of rotary instrumentation in vivo. The setting 0.5 presented overinstrumentation in most of the canals, the setting 1.5 was short in all cases, and the setting 1.0 provided an adequate working length in only 50% of the teeth.
引用
收藏
页码:950 / 954
页数:5
相关论文
共 22 条
[1]   Effect of canal preparation and residual root filling material on root impedance [J].
Al-bulushi, A. ;
Levinkind, M. ;
Flanagan, M. ;
Ng, Y. -L. ;
Gulabivala, K. .
INTERNATIONAL ENDODONTIC JOURNAL, 2008, 41 (10) :892-904
[2]  
American Association of Endodontists, 2003, GLOSS END TERMS, V7
[3]   Apical extent of rotary canal instrumentation with an apex-locating handpiece in vitro [J].
Campbell, D ;
Friedman, S ;
Nguyen, HQ ;
Kaufman, A ;
Keila, S .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1998, 85 (03) :319-324
[4]   Accuracy of Root Length Determination Using Tri Auto ZX and ProTaper Instruments: An In Vitro Study [J].
Carneiro, Everdan ;
Bramante, Clovis Monteiro ;
Picoli, Fabio ;
Letra, Ariadne ;
da Silva Neto, Ulisses Xavier ;
Menezes, Renato .
JOURNAL OF ENDODONTICS, 2006, 32 (02) :142-144
[5]   Ex vivo evaluation of the ability of the ROOT ZX II to locate the apical foramen and to control the apical extent of rotary canal instrumentation [J].
Felippe, W. T. ;
Felippe, M. C. S. ;
Carmona, J. Reyes ;
Crozoe, F. C. I. ;
Alvisi, B. B. .
INTERNATIONAL ENDODONTIC JOURNAL, 2008, 41 (06) :502-507
[6]   Electronic apex locators [J].
Gordon, MPJ ;
Chandler, NP .
INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (07) :425-437
[7]  
Green D, 1960, ORAL SURG ORAL MED O, V13, P725
[8]   In vivo determination of root canal length:: a preliminary report using the Tri Auto ZX apex-locating handpiece [J].
Grimberg, F ;
Banegas, G ;
Chiacchio, L ;
Zmener, O .
INTERNATIONAL ENDODONTIC JOURNAL, 2002, 35 (07) :590-593
[9]  
Grove CJ, 1930, J Am Dent Assoc., V17, P293, DOI [10.14219/jada.archive.1930.0049, DOI 10.14219/JADA.ARCHIVE.1930.0049]
[10]   Accuracy of electronic apex locators in comparison to actual length - an in vivo study [J].
Haffner, C ;
Folwaczny, M ;
Galler, K ;
Hickel, R .
JOURNAL OF DENTISTRY, 2005, 33 (08) :619-625