Pressure Pulse Drive: A Control Method for the Precise Bending of Hydraulic Active Catheters

被引:43
作者
Ikuta, Koji [1 ]
Matsuda, Yoshikatsu [1 ]
Yajima, Daisuke [1 ]
Ota, Yusuke [1 ]
机构
[1] Nagoya Univ, Nagoya, Aichi 4648603, Japan
关键词
Active catheter; hybrid microstereolithography; hydraulically driven system; minimally invasive surgery; SHAPE-MEMORY; ACTUATOR;
D O I
10.1109/TMECH.2011.2138711
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Our group has been developing a compact, active catheter that can be safely used in human blood vessels. The catheter is driven not by electronic mechanisms, but by a water pressure system with micro valves for multidegrees of freedom. The active catheter is composed of a soft, flexible, MRI-compatible material. In this paper, we propose what we call "pressure pulse drive," a new control method to precisely bend the active segment. We design a new set of optimized valves for this method and experimentally verify the precise bending motion and simultaneous drive of the two active catheter segments.
引用
收藏
页码:876 / 883
页数:8
相关论文
共 16 条
[1]   Mechanics Modeling of Tendon-Driven Continuum Manipulators [J].
Camarillo, David B. ;
Milne, Christopher F. ;
Carlson, Chfistopher R. ;
Zinn, Michael R. ;
Salisbury, J. Kenneth .
IEEE TRANSACTIONS ON ROBOTICS, 2008, 24 (06) :1262-1273
[2]   Design and Control of Concentric-Tube Robots [J].
Dupont, Pierre E. ;
Lock, Jesse ;
Itkowitz, Brandon ;
Butler, Evan .
IEEE TRANSACTIONS ON ROBOTICS, 2010, 26 (02) :209-225
[3]   A new approach to develop ionic polymer-metal composites (IPMC) actuator: Fabrication and control for active catheter systems [J].
Fang, Bo-Kai ;
Ju, Ming-Shaung ;
Lin, Chou-Ching K. .
SENSORS AND ACTUATORS A-PHYSICAL, 2007, 137 (02) :321-329
[4]   Research on the axis shape of an active catheter [J].
Fu, Yili ;
Li, Xianling ;
Wang, Shuguo ;
Liu, Hao ;
Liang, Zhaoguang .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2008, 4 (01) :69-76
[5]  
GUO S, 1996, J ROBOTICS SOC JAPAN, V14, P820
[6]  
Ikeuchi M, 2008, PROC IEEE MICR ELECT, P62
[7]   Micro concentrator with opto-sense micro reactor for biochemical IC chip family - 3D composite structure and experimental verification [J].
Ikuta, K ;
Maruo, S ;
Fujisawa, T ;
Yamada, A .
MEMS '99: TWELFTH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 1999, :376-381
[8]  
Ikuta K, 2002, LECT NOTES COMPUT SC, V2488, P182
[9]   Inversion-Based Feedforward Control of Polypyrrole Trilayer Bender Actuators [J].
John, Stephen W. ;
Alici, Gursel ;
Cook, Christopher D. .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2010, 15 (01) :149-156
[10]   Fabrication and characterization of laser-micromachined polypyrrole-based artificial muscle actuated catheters [J].
Lee, Kenneth K. C. ;
Munce, Nigel R. ;
Shoa, Tina ;
Charron, Luc G. ;
Wright, Graham A. ;
Madden, John D. ;
Yang, Victor X. D. .
SENSORS AND ACTUATORS A-PHYSICAL, 2009, 153 (02) :230-236